多中心多厂商PET影像组学特征性能评估:3D打印模体插入物实验

Experimental Multicenter and Multivendor Evaluation of the Performance of PET Radiomic Features Using 3-Dimensionally Printed Phantom Inserts.
PubMed 文献 · J Nucl Med · 2019-08-16

本研究通过3D打印模拟非小细胞肺癌肿瘤形状与异质性的插入物,在多中心多厂商PET/CT系统上评估影像重建对影像组学特征一致性的影响。结果显示,EARL合规重建可协调多数影像组学特征,固定箱宽离散化及重采样至各向同性体素能增强这一优势。

要点

  • EARL合规重建可提高多中心环境下影像组学特征的一致性。
  • 固定箱宽离散化联合各向同性体素重采样进一步放大EARL合规重建的益处。
  • 统计等效重复扫描中高ICC特征比例最高,多中心场景下比例下降。

核心结论

  • EARL合规重建可协调广泛的影像组学特征。
  • 固定箱宽离散化与各向同性体素重采样能增强EARL合规重建的益处。

研究设计与样本

  • 基于真实非小细胞肺癌PET/CT肿瘤形状制作三个3D打印插入物,置于NEMA图像质量模体中进行多次扫描。
  • 扫描设置包括:列表模式扫描重建5个统计等效重复、同一扫描仪重复扫描4次、6台PET/CT系统扫描。
  • 所有图像采用EARL合规重建(不含或含点扩散函数,即EARL1或EARL2)及本地临床偏好重建。
  • 图像分析包括重采样至2-mm立方体素与否,离散化采用固定箱宽0.25或固定箱数64。
  • 计算每种扫描设置的组内相关系数(ICC),ICC高于0.75视为高。

主要发现

  • 统计等效重复中高ICC特征比例最大:固定箱数离散化为70%-91%,固定箱宽为90%-96%。
  • 同一系统重复扫描中高ICC特征比例下降,但多数特征仍高:固定箱数52%-63%,固定箱宽75%-85%。
  • 多中心设置中高ICC特征比例进一步下降;固定箱宽离散化时,EARL1为40%、EARL2为60%,临床偏好重建为21%。
  • 固定箱宽离散化并重采样至各向同性体素时,EARL合规重建的益处更明显。

关键数值与推荐

  • ICC高于0.75定义为高一致性。
  • EARL合规重建分EARL1(不含点扩散函数)和EARL2(含点扩散函数)。
  • 离散化参数:固定箱宽0.25或固定箱数64。
  • 重采样参数:2-mm立方体素。

临床怎么用

  • 多中心PET影像组学研究建议采用EARL合规重建以提高特征一致性。
  • 推荐联合固定箱宽离散化与各向同性体素重采样以增强协调效果。

局限性

  • 原文未明确报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

The sensitivity of radiomic features to several confounding factors, such as reconstruction settings, makes clinical use challenging. To investigate the impact of harmonized image reconstructions on feature consistency, a multicenter phantom study was performed using 3-dimensionally printed phantom inserts reflecting realistic tumor shapes and heterogeneity uptakes. Methods: Tumors extracted from real PET/CT scans of patients with non-small cell lung cancer served as model for three 3-dimensionally printed inserts. Different heterogeneity pattern were realized by printing separate compartments that could be filled with different activity solutions. The inserts were placed in the National Electrical Manufacturers Association image-quality phantom and scanned various times. First, a list-mode scan was acquired and 5 statistically equal replicates were reconstructed. Second, the phantom was scanned 4 times on the same scanner. Third, the phantom was scanned on 6 PET/CT systems. All images were reconstructed using EANM Research Ltd. (EARL)-compliant and locally clinically preferred reconstructions. EARL-compliant reconstructions were performed without (EARL1) or with (EARL2) point-spread function. Images were analyzed with and without resampling to 2-mm cubic voxels. Images were discretized with a fixed bin width (FBW) of 0.25 and a fixed bin number (FBN) of 64. The intraclass correlation coefficient (ICC) of each scan setup was calculated and compared across reconstruction settings. An ICC above 0.75 was regarded as high. Results: The percentage of features yielding a high ICC was largest for the statistically equal replicates (70%-91% for FBN; 90%-96% for FBW discretization). For scans acquired on the same system, the percentage decreased, but most features still resulted in a high ICC (FBN, 52%-63%; FBW, 75%-85%). The percentage of features yielding a high ICC decreased more in the multicenter setting. In this case, the percentage of features yielding a high ICC was larger for images reconstructed with EARL-compliant reconstructions: for example, 40% for EARL1 and 60% for EARL2 versus 21% for the clinically preferred setting for FBW discretization. When discretized with FBW and resampled to isotropic voxels, this benefit was more pronounced. Conclusion: EARL-compliant reconstructions harmonize a wide range of radiomic features. FBW discretization and a sampling to isotropic voxels enhances the benefits of EARL-compliant reconstructions.