文献速递心脏核医学

动态对比增强心脏磁共振与15O-水PET同时定量心肌血流的一致性评估

Evaluation of quantitative CMR perfusion imaging by comparison with simultaneous 15O-water-PET.
PubMed 文献 · J Nucl Cardiol · 2019-07-16

本研究在集成PET-MR扫描仪上,以同时进行的15O-水PET为参考,评估动态对比增强MRI定量心肌血流(MBF)的准确性。MRI与PET的全局和区域MBF相关性均较强,但一致性仅中等,提示MRI定量MBF尚未达到临床常规应用水平。

要点

  • MRI与PET的全局和区域MBF相关性较强(r=0.86和r=0.75)。
  • 全局和区域MBF偏差可忽略,但一致性界限较宽,高MBF值变异更大。
  • MRI定量MBF与PET一致性仅中等,尚不推荐临床常规使用。

核心结论

  • 同时测量的DCE MRI与15O-水PET的MBF相关性较强,但一致性仅中等,MRI定量MBF尚未准备好进入临床。

研究设计与样本

  • 前瞻性研究,15例患者在集成PET-MR扫描仪上完成静息和腺苷负荷的同时DCE MRI与15O-水PET扫描。
  • 采用相关分析和Bland-Altman分析评估MRI与PET基于全局和区域MBF的相关性和一致性。

主要发现

  • 静息和负荷全局平均MBF:MRI为0.97±0.27和3.19±0.70 mL/g/min,PET为1.02±0.28和3.13±1.16 mL/g/min(P=0.66和P=0.81)。
  • 全局和区域MRI与PET基于MBF的相关性较强(r=0.86和r=0.75)。
  • 全局和区域MBF比较的偏差可忽略(均为0.01和0.00 mL/min/g),但一致性界限较宽,高MBF值变异更大。

关键数值与推荐

  • 静息全局MBF:MRI 0.97±0.27 mL/g/min vs PET 1.02±0.28 mL/g/min,P=0.66。
  • 负荷全局MBF:MRI 3.19±0.70 mL/g/min vs PET 3.13±1.16 mL/g/min,P=0.81。
  • 全局和区域MBF相关性:r=0.86和r=0.75。
  • 全局和区域MBF偏差:0.01和0.00 mL/min/g。

临床怎么用

  • 目前MRI定量MBF与15O-水PET一致性仅中等,尚不宜替代PET作为临床常规定量方法。
  • 高MBF值区域变异较大,判读MRI定量结果时需谨慎。

局限性

  • 原文未报告其他局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

BACKGROUND: We assessed the quantitative accuracy of cardiac perfusion measurements using dynamic contrast-enhanced MRI with simultaneous 15O-water PET as reference with a fully integrated PET-MR scanner. METHODS: 15 patients underwent simultaneous DCE MRI and 15O-water PET scans at rest and adenosine-stress on an integrated PET-MR scanner. Correlation and agreement between MRI- and PET-based global and regional MBF values were assessed using correlation and Bland-Altman analysis. RESULTS: Three subjects were excluded due to technical problems. Global mean (± SD) MBF values at rest and stress were 0.97 ± 0.27 and 3.19 ± 0.70 mL/g/min for MRI and 1.02 ± 0.28 and 3.13 ± 1.16 mL/g/min for PET (P = 0.66 and P = 0.81). The correlations between global and regional MRI- and PET-based MBF values were strong (r = 0.86 and r = 0.75). The biases were negligible for both global and regional MBF comparisons (0.01 and 0.00 mL/min/g for both), but the limits of agreement were wide for both global and regional MBF, with larger variability for high MBF-values. CONCLUSION: The correlation between simultaneous MBF measurements with DCE MRI and 15O-water PET measured in an integrated PET-MRI was strong but the agreement was only moderate indicating that MRI-based quantitative MBF measurements is not ready for clinical introduction.

来源:J Nucl Cardiol,查看原文。版权归原作者所有。

引用

  1. Kero T, Johansson E, Engström M, et al. Evaluation of quantitative CMR perfusion imaging by comparison with simultaneous 15O-water-PET. J Nucl Cardiol. 2019. PMID: 31313066. doi:10.1007/s12350-019-01810-z

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。