SiPM PET-CT中BSREM重建采集时间与惩罚因子的影响
噪声随β和AT增加而降低,SUVmax变化大而SUVpeak较稳定;AT为6时β 600图像质量最佳、β 500病灶检出最佳,AT为4因噪声出现假阳性摄取。建议AT 6、β 500-600,不推荐AT<6。
要点
- 建议全身18F-FDG PET-CT采用AT 6(4 MBq/kg且1.5 min/bed)联合BSREM β 500-600。
- AT为4时图像质量差且噪声导致假阳性摄取,不推荐AT<6。
- SUVmax随β和AT变化较大,SUVpeak相对稳定。
核心结论
- 对SiPM PET-CT全身18F-FDG显像,建议AT 6(4 MBq/kg、1.5 min/bed)联合BSREM β 500-600以保证图像质量、病灶检出率和高患者流通量。
- 不推荐AT<6,因噪声导致假阳性摄取的风险增加。
研究设计与样本
- 数据采用BSREM重建,β值100-700,AT 4-16 MBq/kg×min/bed(采集时间1、1.5、2、3、4 min/bed)。
- 计算噪声水平、病灶SUVmax和SUVpeak,并由四名核医学医师对1.0和1.5 min/bed采集时间评估图像质量和病灶可检出性。
主要发现
- 噪声水平随β值和AT增加而降低。
- 病灶SUVmax在不同β值和AT间变化较大,而SUVpeak更稳定。
- AT为6(本研究中为1.5 min/bed)时,β 600获得最佳图像质量,β 500获得最佳病灶检出性。
- AT为4产生质量差的图像,并因噪声出现假阳性摄取。
关键数值与推荐
- AT为6(4 MBq/kg且1.5 min/bed)时,最佳图像质量对应β 600,最佳病灶检出性对应β 500。
- AT为4因噪声导致假阳性摄取和图像质量差,不推荐AT<6。
临床怎么用
- 在SiPM PET-CT全身18F-FDG肿瘤显像中,可采用AT 6(4 MBq/kg、1.5 min/bed)联合BSREM β 500-600进行重建。
- 避免使用AT<6,以降低噪声相关假阳性摄取风险。
局限性
- 图像质量和病灶可检出性仅对1.0和1.5 min/bed采集时间进行评估。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
BACKGROUND: Block-sequential regularized expectation maximization (BSREM), commercially Q. Clear (GE Healthcare, Milwaukee, WI, USA), is a reconstruction algorithm that allows for a fully convergent iterative reconstruction leading to higher image contrast compared to conventional reconstruction algorithms, while also limiting noise. The noise penalization factor β controls the trade-off between noise level and resolution and can be adjusted by the user. The aim was to evaluate the influence of different β values for different activity time products (ATs = administered activity × acquisition time) in whole-body 18F-fluorodeoxyglucose (FDG) positron emission tomography with computed tomography (PET-CT) regarding quantitative data, interpretation, and quality assessment of the images. Twenty-five patients with known or suspected malignancies, referred for clinical 18F-FDG PET-CT examinations acquired on a silicon photomultiplier PET-CT scanner, were included. The data were reconstructed using BSREM with β values of 100-700 and ATs of 4-16 MBq/kg × min/bed (acquisition times of 1, 1.5, 2, 3, and 4 min/bed). Noise level, lesion SUVmax, and lesion SUVpeak were calculated. Image quality and lesion detectability were assessed by four nuclear medicine physicians for acquisition times of 1.0 and 1.5 min/bed position. RESULTS: The noise level decreased with increasing β values and ATs. Lesion SUVmax varied considerably between different β values and ATs, whereas SUVpeak was more stable. For an AT of 6 (in our case 1.5 min/bed), the best image quality was obtained with a β of 600 and the best lesion detectability with a β of 500. AT of 4 generated poor-quality images and false positive uptakes due to noise. CONCLUSIONS: For oncologic whole-body 18F-FDG examinations on a SiPM-based PET-CT, we propose using an AT of 6 (i.e., 4 MBq/kg and 1.5 min/bed) reconstructed with BSREM using a β value of 500-600 in order to ensure image quality and lesion detection rate as well as a high patient throughput. We do not recommend using AT < 6 since the risk of false positive uptakes due to noise increases.
引用
- Trägårdh E, Minarik D, Almquist H, et al. Impact of acquisition time and penalizing factor in a block-sequential regularized expectation maximization reconstruction algorithm on a Si-photomultiplier-based PET-CT system for 18F-FDG. EJNMMI Res. 2019. PMID: 31342214. doi:10.1186/s13550-019-0535-4
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。