高分辨率非人灵长类PET/CT系统性能评估
本研究依据NEMA NU 4-2008标准评估LFER 150 PET/CT系统性能,并在一只食蟹猴中采用18F-FE-PE2I进行活体脑PET成像。体模实验显示径向分辨率优于2.2 mm(中心60 mm直径区域)和3.2 mm(140 mm直径区域),NECR峰值在鼠样体模为430 kcps(115 MBq)、猴样体模为78 kcps(139 MBq)。活体结合电位与高分辨率研究断层仪既往报告相似,表明该系统适用于非人灵长类脑成像。
要点
- LFER 150在中心60 mm和140 mm直径区域的径向分辨率分别优于2.2 mm和3.2 mm
- NECR峰值鼠样体模430 kcps(115 MBq),猴样体模78 kcps(139 MBq)
- 18F-FE-PE2I活体结合电位与高分辨率研究断层仪既往报告相似
核心结论
- 体模实验和1例代表性非人灵长类PET测量证实LFER 150是适用于非人灵长类脑成像的高分辨率PET/CT系统。
研究设计与样本
- 采用NEMA NU 4-2008标准方案评估LFER 150 PET/CT系统,并进行1只食蟹猴(4.5 kg,氯胺酮/赛拉嗪麻醉)的活体PET测量。
主要发现
- 径向半高全宽分辨率在中心60 mm直径区域小于2.2 mm,在140 mm直径区域小于3.2 mm。
- 400-600 keV和250-750 keV能窗的最大灵敏度分别为30.03 cps/kBq(3.3%)和49.11 cps/kBq(5.4%)。
- 图像质量体模均匀性为3.3%,空气和水溢出比均为0.1。
- NECR曲线峰值在鼠样体模为430 kcps(115 MBq),在猴样体模为78 kcps(139 MBq)。
- Derenzo体模中直径1 mm的棒可肉眼分辨。
临床怎么用
- 该系统的高分辨率和灵敏度特征支持将其用于非人灵长类脑PET成像研究。
局限性
- 活体评估仅纳入1只非人灵长类,样本量有限。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
The LFER 150 PET/CT device (large-field-of-view extreme-resolution portable research imager) is a system for nonhuman primate (NHP) imaging. The objective of this study was to evaluate the performance of the system using the National Electrical Manufacturers Association NU 4-2008 standard protocol. As a preliminary in vivo evaluation of the system, a PET measurement in an NHP was also performed. Methods: Resolution, sensitivity, image quality, and noise-equivalent count rate (NECR) were measured. NECR measurement was performed with a ratlike phantom and a monkeylike phantom. A Derenzo phantom experiment was performed to test the resolution using 3-dimensional ordered-subset expectation maximization reconstruction. One cynomolgus monkey (4.5 kg, intravenous ketamine/xylazine anesthesia) was examined with the dopamine transporter radioligand 18F-FE-PE2I (94 MBq) to evaluate the in vivo performance of the system. List-mode PET data acquired for 93 min were reconstructed into 38 frames with the Tera-Tomo 3-dimensional engine. Binding potential for caudate nucleus, putamen, and substantia nigra was evaluated using the simplified reference tissue model. Results: Radial full-width half-maximum resolution using Fourier rebinning and a 2-dimensional filtered backprojection algorithm was less than 2.2 mm and less than 3.2 mm in the central 60-mm-diameter and 140-mm-diameter regions, respectively. Maximum sensitivity in the 400- to 600-keV and 250- to 750-keV energy windows was 30.03 cps/kBq (3.3%) and 49.11 cps/kBq (5.4%), respectively. The uniformity in the image-quality phantom was 3.3%, and the spillover ratio for air and water was 0.1. The peak of the NECR curve was 430 kcps (at 115 MBq) with the ratlike phantom and 78 kcps (at 139 MBq) with the monkeylike phantom. Rods of the Derenzo phantom with 1-mm diameter could be distinguished by eye. In the NHP experiment, binding potentials in the caudate, putamen, and substantia nigra (4.9, 4.9, and 1, respectively) were similar to those previously reported using the same radioligand and a high-resolution research tomograph. Conclusion: The results obtained from phantom experiments and 1 representative PET measurement in an NHP confirm that the LFER 150 is a high-resolution PET/CT system with suitable performance for brain imaging in NHPs.