uEXPLORER全身动态重建与参数成像
本研究在首台194厘米全身PET/CT(uEXPLORER)上开展首例人体动态扫描,对一名健康女性静脉注射256 MBq 18F-FDG后采集1小时动态数据,采用OSEM与核方法重建并用Patlak模型分析全身动力学,结果显示直接参数重建较间接法方差降低2至3倍。
要点
- uEXPLORER全身动态PET可实现更准确的示踪剂动力学分析。
- 直接Patlak参数重建较间接法方差降低2至3倍。
- 核方法间接Patlak图像质量接近直接法且计算更快。
核心结论
- uEXPLORER具备全身参数成像能力。
- 核正则化重建与直接参数重建均优于传统间接OSEM。
研究设计与样本
- 在194厘米全身PET/CT(uEXPLORER)上开展首例人体动态全身PET研究。
- 招募1名健康女性,静脉注射256 MBq 18F-FDG,采集1小时动态扫描。
主要发现
- OSEM重建图像质量良好、噪声低,1秒帧亦如此。
- 核方法进一步改善图像质量。
- 直接重建法较间接法方差降低2至3倍。
- 核方法间接Patlak图像质量与直接Patlak相似,计算时间更少、收敛更快。
关键数值与推荐
- 动力学分析采用Patlak图模型,输入函数取自降主动脉区域。
- 间接Patlak分析基于重建帧,直接重建基于列表模式数据,均取最后30分钟数据。
临床怎么用
- 全身动态PET可支持更准确的示踪剂动力学研究。
- 参数成像可优先考虑核正则化重建或直接参数重建以改善图像质量。
局限性
- 研究仅纳入1名健康女性受试者。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
The world's first 194-cm-long total-body PET/CT scanner (uEXPLORER) has been built by the EXPLORER Consortium to offer a transformative platform for human molecular imaging in clinical research and health care. Its total-body coverage and ultra-high sensitivity provide opportunities for more accurate tracer kinetic analysis in studies of physiology, biochemistry, and pharmacology. The objective of this study was to demonstrate the capability of total-body parametric imaging and to quantify the improvement in image quality and kinetic parameter estimation by direct and kernel reconstruction of the uEXPLORER data. Methods: We developed quantitative parametric image reconstruction methods for kinetic analysis and used them to analyze the first human dynamic total-body PET study. A healthy female subject was recruited, and a 1-h dynamic scan was acquired during and after an intravenous injection of 256 MBq of 18F-FDG. Dynamic data were reconstructed using a 3-dimensional time-of-flight list-mode ordered-subsets expectation maximization (OSEM) algorithm and a kernel-based algorithm with all quantitative corrections implemented in the forward model. The Patlak graphical model was used to analyze the 18F-FDG kinetics in the whole body. The input function was extracted from a region over the descending aorta. For comparison, indirect Patlak analysis from reconstructed frames and direct reconstruction of parametric images from the list-mode data were obtained for the last 30 min of data. Results: Images reconstructed by OSEM showed good quality with low noise, even for the 1-s frames. The image quality was further improved using the kernel method. Total-body Patlak parametric images were obtained using either indirect estimation or direct reconstruction. The direct reconstruction method improved the parametric image quality, having a better contrast-versus-noise tradeoff than the indirect method, with a 2- to 3-fold variance reduction. The kernel-based indirect Patlak method offered image quality similar to the direct Patlak method, with less computation time and faster convergence. Conclusion: This study demonstrated the capability of total-body parametric imaging using the uEXPLORER. Furthermore, the results showed the benefits of kernel-regularized reconstruction and direct parametric reconstruction. Both can achieve superior image quality for tracer kinetic studies compared with the conventional indirect OSEM for total-body imaging.
引用
- Zhang X, Xie Z, Berg E, et al. Total-Body Dynamic Reconstruction and Parametric Imaging on the uEXPLORER. J Nucl Med. 2019. PMID: 31302637. doi:10.2967/jnumed.119.230565
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。