特定摄取大小指数在I-123-FP-CIT SPECT半定量分析中的诊断效能:多中心统一设置与典型临床单相机设置的比较

Diagnostic performance of the specific uptake size index for semi-quantitative analysis of I-123-FP-CIT SPECT: harmonized multi-center research setting versus typical clinical single-camera setting.
PubMed 文献 · EJNMMI Res · 2019-05-07

本研究比较了纹状体FP-CIT摄取的特定摄取大小指数(SUSI)与特定结合比值(SBR)在多中心/多相机(MULTI)设置和单中心/单相机(MONO)设置中的诊断效能。MULTI设置纳入帕金森病(PD)患者438例和健康对照207例,MONO设置纳入122例常规临床患者。结果显示,壳核SUSI在两种设置中均提供最佳诊断效能,SUSI在多中心设置中略优于SBR,在单中心设置中略差于SBR,但差异均无统计学意义。

要点

  • 壳核(最小半球)SUSI在两种设置中AUC最高,优于全纹状体SUSI。
  • SUSI分析前需将SPECT图像进行立体定向归一化。
  • 参考区非可置换计数的估计误差可导致SUSI和SBR误分类,尤其边缘病例。

核心结论

  • SUSI在多中心/多相机设置中尤其有用,但分析前需进行立体定向归一化,壳核SUSI诊断效能优于全纹状体。

研究设计与样本

  • MULTI设置纳入帕金森进展标记计划中PD患者438例和健康对照207例;MONO设置纳入122例常规临床患者,根据随访分为神经退行性(84例)和非神经退行性(38例)。

主要发现

  • 在两种设置中,壳核(最小半球)AUC最高,与半定量方法无关。
  • MNI空间壳核SUSI的AUC为0.969,患者空间为0.961(p=0.129)。
  • MULTI设置中SUSI的AUC为0.993,SBR为0.991(p=0.207);MONO设置中SUSI为0.969,SBR为0.976(p=0.259)。
  • 参考区体素强度变异性在误分类病例中大于正确分类病例(MULTI设置:SUSI p=0.007,SBR p=0.012)。

关键数值与推荐

  • 壳核(最小半球)SUSI在MNI空间和患者空间的AUC分别为0.969和0.961。
  • MULTI设置中SUSI和SBR的AUC分别为0.993和0.991;MONO设置中分别为0.969和0.976。

临床怎么用

  • 在多中心/多相机环境中可优先考虑使用SUSI进行半定量分析。
  • SUSI分析前应将SPECT图像立体定向归一化至MNI空间。
  • 应视觉检查参考区FP-CIT摄取,尤其关注高变异性病例以识别可能的误分类。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

INTRODUCTION: The specific uptake size index (SUSI) of striatal FP-CIT uptake is independent of spatial resolution in the SPECT image, in contrast to the specific binding ratio (SBR). This suggests that the SUSI is particularly appropriate for multi-site/multi-camera settings in which camera-specific effects increase inter-subject variability of spatial resolution. However, the SUSI is sensitive to inter-subject variability of striatum size. Furthermore, it might be more sensitive to errors of the estimate of non-displaceable FP-CIT binding. This study compared SUSI and SBR in the multi-site/multi-camera (MULTI) setting of a prospective multi-center study and in a mono-site/mono-camera (MONO) setting representative of clinical routine. METHODS: The MULTI setting included patients with Parkinson's disease (PD, n = 438) and healthy controls (n = 207) from the Parkinson Progression Marker Initiative. The MONO setting included 122 patients from routine clinical patient care in whom FP-CIT SPECT had been performed with the same double-head SPECT system according to the same acquisition and reconstruction protocol. Patients were categorized as "neurodegenerative" (n = 84) or "non-neurodegenerative" (n = 38) based on follow-up data. FP-CIT SPECTs were stereotactically normalized to MNI space. SUSI and SBR were computed for caudate, putamen, and whole striatum using unilateral ROIs predefined in MNI space. SUSI analysis was repeated in native patient space in the MONO setting. The area (AUC) under the ROC curve for identification of PD/"neurodegenerative" cases was used as performance measure. RESULTS: In both settings, the highest AUC was achieved by the putamen (minimum over both hemispheres), independent of the semi-quantitative method (SUSI or SBR). The putaminal SUSI provided slightly better performance with ROI analysis in MNI space compared to patient space (AUC = 0.969 vs. 0.961, p = 0.129). The SUSI (computed in MNI space) performed slightly better than the SBR in the MULTI setting (AUC = 0.993 vs. 0.991, p = 0.207) and slightly worse in the MONO setting (AUC = 0.969 vs. AUC = 0.976, p = 0.259). There was a trend toward larger AUC difference between SUSI and SBR in the MULTI setting compared to the MONO setting (p = 0.073). Variability of voxel intensity in the reference region was larger in misclassified cases compared to correctly classified cases for both SUSI and SBR (MULTI setting: p = 0.007 and p = 0.012, respectively). CONCLUSIONS: The SUSI is particularly useful in MULTI settings. SPECT images should be stereotactically normalized prior to SUSI analysis. The putaminal SUSI provides better diagnostic performance than the SUSI of the whole striatum. Errors of the estimate of non-displaceable count density in the reference region can cause misclassification by both SUSI and SBR, particularly in borderline cases. These cases might be identified by visual checking FP-CIT uptake in the reference region for particularly high variability.

来源:EJNMMI Res,查看原文。版权归原作者所有。

引用

  1. Buchert R, Lange C, Spehl TS, et al. Diagnostic performance of the specific uptake size index for semi-quantitative analysis of I-123-FP-CIT SPECT: harmonized multi-center research setting versus typical clinical single-camera setting. EJNMMI Res. 2019. PMID: 31065816. doi:10.1186/s13550-019-0506-9