常规与CZT单光子发射计算机断层心肌灌注显像诊断冠心病的荟萃分析

Diagnostic performance of myocardial perfusion imaging with conventional and CZT single-photon emission computed tomography in detecting coronary artery disease: A meta-analysis.
PubMed 文献 · J Nucl Cardiol · 2019-05-14

该荟萃分析纳入40篇研究共7334例患者,比较常规SPECT与CZT-SPECT以侵入性冠状动脉造影≥50%狭窄为参考标准诊断冠心病的效能。C-SPECT汇总敏感度85%、特异度66%,CZT-SPECT为89%、69%,CZT-SPECT的AUC(0.89)略高于C-SPECT(0.83),两者诊断准确性差异有统计学意义(卡方11.28,P<.05)。

要点

  • C-SPECT与CZT-SPECT诊断造影证实的冠心病均有良好效能,CZT-SPECT准确性略高。
  • CZT-SPECT汇总敏感度89%、特异度69%,C-SPECT分别为85%、66%。
  • CZT-SPECT的AUC为0.89、诊断OR为17,均高于C-SPECT的0.83和11。

核心结论

  • 常规SPECT与CZT-SPECT诊断造影证实的冠心病均有良好效能,CZT-SPECT准确性略高。

研究设计与样本

  • 荟萃分析检索2000年1月至2018年2月发表的研究,纳入以侵入性冠状动脉造影≥50%直径狭窄为参考标准评估C-SPECT或CZT-SPECT诊断冠心病的研究。
  • 共纳入40篇合格文章(25篇C-SPECT、15篇CZT-SPECT),合计7334例患者(C-SPECT研究4997例,CZT-SPECT研究2337例)。

主要发现

  • C-SPECT汇总敏感度85%、特异度66%,CZT-SPECT汇总敏感度89%、特异度69%。
  • CZT-SPECT的AUC为0.89,高于C-SPECT的0.83;汇总诊断OR分别为17和11。
  • 两种检查准确性略有差异(卡方11.28,P<.05)。
  • 荟萃回归分析未发现所考虑的人口学和临床变量与C-SPECT及CZT-SPECT研究的敏感度和特异度存在显著关联。

关键数值与推荐

  • 冠心病定义为侵入性冠状动脉造影显示至少50%直径狭窄。
  • C-SPECT汇总敏感度85%、特异度66%,CZT-SPECT汇总敏感度89%、特异度69%。
  • CZT-SPECT的AUC为0.89、诊断OR为17;C-SPECT的AUC为0.83、诊断OR为11。
  • 两种检查准确性差异的卡方值为11.28,P<.05。

临床怎么用

  • 结果支持在临床常规中使用新型伽马相机,并考虑其采集时间缩短和辐射暴露降低的优势。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

BACKGROUND: We performed a meta-analysis to compare the diagnostic performance of conventional SPECT (C-SPECT) and cadmium-zinc-telluride (CZT)-SPECT systems in detecting angiographically proven coronary artery disease (CAD). METHODS: Studies published between January 2000 and February 2018 were identified by database search. We included studies assessing C-SPECT or CZT-SPECT as a diagnostic test to evaluate patients for the presence of CAD, defined as at least 50% diameter stenosis on invasive coronary angiography. A study was eligible regardless of whether patients were referred for suspected or known CAD. RESULTS: We identified 40 eligible articles (25 C-SPECT and 15 CZT-SPECT studies) including 7334 patients (4997 in C-SPECT and 2337 in CZT-SPECT studies). The pooled sensitivity and specificity were 85% and 66% for C-SPECT and 89% and 69% for CZT-SPECT imaging studies. The area under the curve was slightly higher for CZT-SPECT (0.89) compared to C-SPECT (0.83); accordingly, the summary diagnostic OR was 17 for CZT-SPECT and 11 for C-SPECT. The accuracy of the two tests slightly differs between C-SPECT and CZT-SPECT (chi-square 11.28, P < .05). At meta-regression analysis, no significant association between both sensitivity and specificity and demographical and clinical variables considered was found for C-SPECT and CZT-SPECT studies. CONCLUSIONS: C-SPECT and CZT-SPECT have good diagnostic performance in detecting angiographic proven CAD, with a slightly higher accuracy for CZT-SPECT. This result supports the use of the novel gamma cameras in clinical routine practices also considering the improvements in acquisition time and radiation exposure reduction.

来源:J Nucl Cardiol,查看原文。版权归原作者所有。

引用

  1. Cantoni V, Green R, Acampa W, et al. Diagnostic performance of myocardial perfusion imaging with conventional and CZT single-photon emission computed tomography in detecting coronary artery disease: A meta-analysis. J Nucl Cardiol. 2019. PMID: 31089962. doi:10.1007/s12350-019-01747-3