本研究通过模体和患者数据,首次比较新型全身CZT相机(DNM 670CZT)与专用心脏CZT相机(DNM 530c)在心肌灌注SPECT中的性能。结果显示两者空间分辨率相同,但DNM 670CZT计数灵敏度低两倍,导致锐度指数和对比噪声比更低;壁厚和心室容积相似,灌注模式定性和定量评估一致性良好。

要点

  • DNM 670CZT与DNM 530c空间分辨率相同。
  • DNM 670CZT计数灵敏度较DNM 530c低两倍。
  • 两相机灌注模式评估一致性良好,壁厚和心室容积相似。

核心结论

  • 全身DNM 670CZT相机可用于心肌灌注SPECT显像。

研究设计与样本

  • 模体研究使用99mTc线性源评估空间分辨率,心脏模体评估计数灵敏度、锐度指数、对比噪声比、壁厚、非均匀性指数、灌注评分和心室容积,并研究矩阵大小和采集时间的影响。
  • 患者研究使用QPS/QGS软件定量评估节段灌注、运动和厚度评分的一致性。

主要发现

  • 两相机空间分辨率相同。
  • DNM 670CZT计数灵敏度较DNM 530c低两倍,导致锐度指数和对比噪声比更低。
  • 壁厚和心肌容积相似。
  • 模体和患者中两相机灌注模式视觉和定量评估一致性良好。

临床怎么用

  • 新型全身CZT相机可用于心肌灌注SPECT显像,但需注意其计数灵敏度较低可能影响图像锐度和对比噪声比。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

BACKGROUND: Dedicated cardiac Cadmium-zinc-telluride (CZT) cameras show superior performances compared with Anger systems, particularly in terms of spatial resolution and count sensitivity. This study evaluated the performances of a new polyvalent whole body CZT camera (DNM 670CZT) compared with a cardiac dedicated CZT camera (DNM 530c) for myocardial perfusion SPECT. METHODS: The spatial resolution was evaluated with three linear sources filled with 99mTc. We used a cardiac phantom to evaluate count sensitivity, sharpness index, contrast-to-noise ratio, wall thickness, non-uniformity index, perfusion scores and ventricle volumes for both cameras. The impact of matrix size, and acquisition time was investigated. Concordance between the two cameras was evaluated in patients using QPS/QGS software for quantitative segmental perfusion, motion and thickness scores. RESULTS: The spatial resolution was identical with the two cameras. Count sensitivity of the DNM 670CZT was twofold lower compared with the DNM 530c, leading to lower sharpness index and contrast-to-noise ratio. The wall thickness and the myocardial volumes were similar. Visual and quantitative assessments of the perfusion patterns have shown a good concordance of the two cameras on phantoms and in patients. CONCLUSION: This study demonstrated the feasibility of myocardial perfusion SPECT imaging using the new whole-body DNM 670CZT camera.

来源:J Nucl Cardiol,查看原文。版权归原作者所有。

引用

  1. Morelle M, Bellevre D, Hossein-Foucher C, et al. First comparison of performances between the new whole-body cadmium-zinc-telluride SPECT-CT camera and a dedicated cardiac CZT camera for myocardial perfusion imaging: Analysis of phantom and patients. J Nucl Cardiol. 2019. PMID: 30963419. doi:10.1007/s12350-019-01702-2

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。