模拟低剂量CZT-SPECT协议并额外计数校准后测定左心室射血分数
本研究在高灵敏度CZT-SPECT相机上模拟70%记录时间缩短,并通过额外计数校准防止EF高估,与平面放射性核素血管造影(2D-RNA)参考EF比较。结果显示校准后SPECT100和SPECT30的EF与2D-RNA呈强相关且均值接近,未校准则均值上移。左心室EF可在高灵敏度CZT相机上通过减少70%注射活度和额外计数校准测定。
要点
- CZT-SPECT缩短70%记录时间并额外计数校准后,EF与2D-RNA强相关且均值接近。
- 未进行计数校准时,SPECT100和SPECT30的EF均值均上移。
- 左心室EF可在高灵敏度CZT相机上通过减少70%注射活度和计数校准测定。
核心结论
- 左心室EF可在高灵敏度CZT相机上通过减少70%注射活度和额外计数校准测定,以增强与2D-RNA值的一致性。
研究设计与样本
- 在因心肌病(n=23)或化疗监测(n=50)转诊至2D-RNA的患者中,追加10分钟CZT-SPECT记录,使用850 MBq体内红细胞标记。
- 比较100%(SPECT100)或30%(SPECT30)投影时间及基于2D-RNA相应腔室体积计数校准的CZT-SPECT EF与2D-RNA EF。
主要发现
- 2D-RNA EF与校准后SPECT100 EF(y=0.89x+6.62;R2=0.87)和SPECT30 EF(y=0.87x+8.40;R2=0.85)呈强且等效关系。
- SPECT100平均EF(54%±15%)和SPECT30平均EF(53%±16%)接近2D-RNA(55%±15%)。
- 无计数校准时,SPECT100(60%±18%)和SPECT30(60%±18%)均值上移。
关键数值与推荐
- SPECT100校准后EF与2D-RNA的回归方程:y=0.89x+6.62,R2=0.87。
- SPECT30校准后EF与2D-RNA的回归方程:y=0.87x+8.40,R2=0.85。
- 2D-RNA平均EF为55%±15%,SPECT100校准后为54%±15%,SPECT30校准后为53%±16%。
- 无计数校准时SPECT100和SPECT30平均EF均为60%±18%。
临床怎么用
- 在临床中可采用高灵敏度CZT-SPECT相机,将注射活度减少70%并追加计数校准,以获得与2D-RNA一致的左心室EF。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
PURPOSE: To determine whether the left ventricular ejection fractions (EFs), measured on a high-sensitivity CZT single photon emission computed tomography (SPECT)-camera with a 70% reduction in recording times and a prevention of EF overestimation through an additional count-calibration, are concordant with reference EF from planar radionuclide angiography (2D-RNA). METHODS: An additional 10-minute CZT-SPECT recording was performed in patients referred to 2D-RNA for cardiomyopathy (n = 23) or chemotherapy monitoring (n = 50) with an in vivo red blood cell labeling with 850 MBq [Formula: see text]. The EF, obtained from CZT-SPECT with 100% (SPECT100) or 30% (SPECT30) projection times and with a SPECT-count calibration on the 2D-RNA counts of corresponding cavity volumes, were compared to EF from 2D-RNA. RESULTS: Strong and equivalent relationships were documented between the EF from 2D-RNA and the calibrated EF from SPECT100 (y = 0.89x + 6.62; R2 = 0.87) and SPECT30 (y = 0.87x + 8.40; R2 = 0.85), and the mean EF from SPECT100 (54% ± 15%) and SPECT30 (53% ± 16%) were close to that from 2D-RNA (55% ± 15%). However, upward shifts in these mean values were documented in the absence of count calibration for both SPECT100 (60% ± 18%) and SPECT30 (60% ± 18%). CONCLUSION: Left ventricular EF may be determined on a high-sensitivity CZT-camera, a 70% reduction in injected activities, and an additional count-calibration for further enhancing the concordance with 2D-RNA values.