文献速递心脏核医学放射性药物与辐射防护

患者个体化SPECT成像方案标准化图像噪声

Patient-specific SPECT imaging protocols to standardize image noise.
PubMed 文献 · J Nucl Cardiol · 2019-03-04

该研究基于43例在常规和多针孔心脏SPECT相机上采集的图像计算图像噪声,建立噪声与放射性活度、扫描时间和体重的函数关系,并在58例测试集中验证。结果显示两台相机光子灵敏度相差3.6倍,但匹配噪声水平需16倍活度-扫描时间差异;按体重线性缩放活度可标准化噪声,但缩放因子取决于相机和重建参数。

要点

  • 光子计数相等不足以保证重建图像噪声相等。
  • 按体重线性缩放活度可标准化噪声,但缩放因子因相机和重建参数而异。
  • 常规相机图像噪声从45kg到128kg翻倍(12.8分钟),多针孔相机三倍(5分钟)。

核心结论

  • 按患者体重线性缩放放射性活度可标准化重建图像噪声,但缩放因子取决于相机型号和图像重建参数。

研究设计与样本

  • 研究基于43例患者分别在常规和多针孔心脏SPECT相机上采集的图像计算图像噪声,建立噪声与放射性活度、扫描时间和体重的函数关系,并在58例测试集中验证。

主要发现

  • 两台相机光子灵敏度相差3.6倍,但匹配噪声水平需16倍的活度-扫描时间差异。
  • 常规相机图像噪声从45kg到128kg翻倍(12.8分钟),多针孔相机三倍(5分钟),对应350 MBq(9.5 mCi)99mTc-tetrofosmin。
  • 平均体型患者的图像噪声分别为16.3%和6.1%。

临床怎么用

  • 临床制定SPECT采集方案时需考虑相机和重建算法对噪声的影响,不能仅以光子计数相等作为标准化依据。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

BACKGROUND: In addition to acquired photon counts, image noise depends on the image reconstruction algorithm. This work develops patient-specific activity or acquisition time protocols to standardize the average noise in a reconstructed image for different patients, cameras, and reconstruction algorithms. METHODS: Image noise was calculated for images from 43 patients acquired on both a conventional and a multiple-pinhole cardiac SPECT camera. Functions were found to relate image noise to radiotracer activity, scan time, and body mass and were validated by normalizing the image noise in a test set of 58 patients. RESULTS: There was a 3.6-fold difference in photon sensitivity between the two cameras but a 16-fold difference in activity-scan time was necessary to match the noise levels. Image noise doubled from 45 to 128 kg for the conventional camera (12.8 minutes) and tripled for the multiple-pinhole camera (5 minutes) for 350 MBq (9.5 mCi) 99mTc-tetrofosmin. It was 16.3% and 6.1% respectively for an average sized patient. CONCLUSIONS: A linear scaling of activity with respect to the patient weight normalizes image noise but the scaling factors depend on the choice of camera and image reconstruction parameters. Therefore, equivalent numbers of acquired photon counts are not sufficient to guarantee equivalent image noise.

来源:J Nucl Cardiol,查看原文。版权归原作者所有。