充血性心力衰竭患者心肌血流量定量:快速旋转机架SPECT与CZT SPECT头对头比较

Myocardial blood flow quantitation in patients with congestive heart failure: head-to-head comparison between rapid-rotating gantry SPECT and CZT SPECT.
PubMed 文献 · J Nucl Cardiol · 2019-02-19

本研究纳入20例充血性心力衰竭(CHF)和20例非CHF患者,连续两天分别用快速旋转机架(RRG)SPECT和CZT SPECT进行动态SPECT成像,比较静息心肌血流量(MBF)定量。两组中两种相机测得的静息MBF均无显著差异,线性回归和Bland-Altman分析显示系统差异小、相关性好,表明经物理校正后两种SPECT系统可提供可比的MBF定量。

要点

  • CHF与非CHF患者中,RRG-SPECT与CZT-SPECT测得的静息MBF均无显著差异。
  • CHF患者两种系统MBF线性回归r=0.97,Bland-Altman系统差异Δ=0.03 mL·min-1·g-1(95%CI -0.11至0.20)。
  • 非CHF患者线性回归r=0.89,系统差异Δ=0.02 mL·min-1·g-1(95%CI -0.14至0.19)。

核心结论

  • 经物理校正及其他图像校正后,无论使用RRG-SPECT还是CZT-SPECT,在CHF和非CHF患者中均可获得可比的MBF定量。

研究设计与样本

  • 纳入20例CHF和20例非CHF患者,连续两天分别行RRG-SPECT和CZT-SPECT动态SPECT成像,弹丸注射20 mCi 99mTc-MIBI。
  • 动态图像经全物理校正、心室溢出和部分容积效应校正,采用单组织动力学模型,静息MBF由K1转换并经MIBI提取分数校正和心率血压乘积调整。

主要发现

  • 两组中RRG-SPECT与CZT-SPECT测得的静息MBF均无显著差异(CHF组P=0.06,非CHF组P=0.2)。
  • CHF患者线性回归y(RRG)=1.0412x(CZT),r=0.97;Bland-Altman系统差异Δ=0.03 mL·min-1·g-1(95%CI -0.11至0.20)。
  • 非CHF患者线性回归y(RRG)=1.025x(CZT),r=0.89;Bland-Altman系统差异Δ=0.02 mL·min-1·g-1(95%CI -0.14至0.19)。

关键数值与推荐

  • CHF患者两种系统静息MBF比较P=0.06,非CHF患者P=0.2。
  • CHF患者线性回归r=0.97,非CHF患者r=0.89。
  • CHF患者Bland-Altman系统差异Δ=0.03 mL·min-1·g-1(95%CI -0.11至0.20),非CHF患者Δ=0.02 mL·min-1·g-1(95%CI -0.14至0.19)。

临床怎么用

  • 在CHF和非CHF患者中,RRG-SPECT与CZT-SPECT经物理校正后测得的静息MBF具有可比性,支持两种系统在临床MBF定量中互换使用。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

PURPOSE: Recently, the feasibility of myocardial blood flow (MBF) quantitation using rapid-rotating gantry (RRG) and cadmium-zinc-telluride (CZT) SPECT cameras has been demonstrated. We compared MBF quantitation using these two camera systems. METHODS: Twenty patients with congestive heart failure (CHF) and 20 patients without CHF (non-CHF) were included. On two consecutive days, dynamic SPECT imaging was performed after a bolus injection of 20 mCi of 99mTc-Sestamibi (MIBI) with RRG-SPECT and list-mode acquisition with CZT-SPECT. All dynamic SPECT images were reconstructed with full physical corrections, corrections for ventricular spillover and partial volume effect, using one-tissue kinetic modeling. Resting MBF converted from K1 was then corrected for MIBI extraction fraction and adjusted for rate-pressure product. RESULTS: In both patient groups, there was no significant difference between resting MBF values measured with RRG-SPECT and CZT-SPECT systems (P = 0.06, P = 0.2 respectively). For CHF patients, linear regression (LR) was y(RRG) = 1.0412x (CZT) (r = 0.97) with a small systemic difference (Δ = 0.03 mL·min-1·g-1, 95% CI - 0.11 to 0.20) by Bland-Altman analysis. For non-CHF patients, LR was y(RRG) = 1.025x (CZT) (r = 0.89) with also small systemic difference (ΔT= 0.02 mL·min-1·g-1, 95% CI - 0.14 to 0.19) in BA analysis. CONCLUSION: Physical corrections along with other image corrections can provide comparable MBF quantitations in both CHF and non-CHF patients, regardless of the type of SPECT systems used.

来源:J Nucl Cardiol,查看原文。版权归原作者所有。