数据驱动投影域呼吸运动补偿用于心脏PET/CT与PET/MR的可行性研究
本研究探讨在图像重建前直接对PET列表模式数据应用数据驱动、基于投影的呼吸运动补偿(DPR-MoCo)技术的可行性。通过XCAT体模模拟、实验体模采集及7例双示踪剂PET/MR患者研究验证,并与无补偿、门控及基于图像的RTA-MoCo方法比较。结果表明DPR-MoCo可在不增加噪声的情况下改善心肌图像质量,且无需外部运动跟踪设备。
核心结论
- 基于投影的DPR-MoCo方法无需外部运动跟踪设备即可改善心肌PET图像质量。
研究设计与样本
- 研究采用XCAT体模模拟(Biograph mMR PET/MR)、实验体模采集(Biograph mCT PET/CT)及7例双示踪剂(18F-FDG/13N-NH3)PET/MR患者研究进行验证。
关键数值与推荐
- 评估指标包括心肌靶本比(TBRLV)和肝脏噪声(CoVliver)。
- 患者水平评估指标包括收缩末期容积(ESV)、舒张末期容积(EDV)和左心室射血分数(EF),并与心脏MR功能值比较。
临床怎么用
- DPR-MoCo可在重建前对PET列表模式数据直接进行呼吸运动补偿,无需外部运动跟踪设备。
- 该方法在保持噪声特性与无补偿图像相似的同时,可能改善心肌图像质量和功能定量。
局限性
- 患者样本量较小(7例),部分比较结果未达统计学显著性。
- 研究为可行性验证,尚需更大样本进一步确认临床价值。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
BACKGROUND: Respiratory patient motion causes blurring of the PET images that may impact accurate quantification of perfusion and infarction extents in PET myocardial viability studies. In this study, we investigate the feasibility of correcting for respiratory motion directly in the PET-listmode data prior to image reconstruction using a data-driven, projection-based, respiratory motion compensation (DPR-MoCo) technique. METHODS: The DPR-MoCo method was validated using simulations of a XCAT phantom (Biograph mMR PET/MR) as well as experimental phantom acquisitions (Biograph mCT PET/CT). Seven patient studies following a dual-tracer (18F-FDG/13N-NH3) imaging-protocol using a PET/MR-system were also evaluated. The performance of the DPR-MoCo method was compared against reconstructions of the acquired data (No-MoCo), a reference gate method (gated) and an image-based MoCo method using the standard reconstruction-transform-average (RTA-MoCo) approach. The target-to-background ratio (TBRLV) in the myocardium and the noise in the liver (CoVliver) were evaluated for all acquisitions. For all patients, the clinical effect of the DPR-MoCo was assessed based on the end-systolic (ESV), the end-diastolic volumes (EDV) and the left ventricular ejection fraction (EF) which were compared to functional values obtained from the cardiac MR. RESULTS: The DPR-MoCo and the No-MoCo images presented with similar noise-properties (CoV) (P = .12), while the RTA-MoCo and reference-gate images showed increased noise levels (P = .05). TBRLV values increased for the motion limited reconstructions when compared to the No-MoCo reconstructions (P > .05). DPR-MoCo results showed higher correlation with the functional values obtained from the cardiac MR than the No-MoCo results, though non-significant (P > .05). CONCLUSION: The projection-based DPR-MoCo method helps to improve PET image quality of the myocardium without the need for external devices for motion tracking.
引用
- Lassen ML, Beyer T, Berger A, et al. Data-driven, projection-based respiratory motion compensation of PET data for cardiac PET/CT and PET/MR imaging. J Nucl Cardiol. 2019. PMID: 30761482. doi:10.1007/s12350-019-01613-2