文献速递肿瘤 PET/CT

数据驱动呼吸门控波形在临床PET成像中的评估

Evaluation of data-driven respiratory gating waveforms for clinical PET imaging.
PubMed 文献 · EJNMMI Res · 2019-01-03

本研究评估了一种基于主成分分析的数据驱动呼吸门控算法在常规FDG PET中的临床稳健性。结果显示R与视觉评分强相关,高R波形多位于胸腹部,数据驱动门控使病灶SUVmax平均提高11%。

要点

  • 数据驱动呼吸门控算法提供的信噪比指标R与视觉评分强相关(r=0.86)。
  • R≥15的波形中86%被评分为适合呼吸门控,且90%位于肝脏穹顶上下特定范围。
  • 数据驱动门控使R≥15床位中病灶SUVmax平均提高11%。

核心结论

  • 多数高R波形对应预期存在呼吸运动的部位,视觉评估认为适合呼吸门控,且使用后提高病灶SUVmax。

研究设计与样本

  • 两位读者对波形进行3分制视觉评分并取平均,同时计算信噪比指标R;对胸腹部明确病灶比较门控与非门控重建的SUVmax。

主要发现

  • R与视觉评分S呈强且显著相关(r=0.86)。
  • R≥15的波形中86%被评分为可接受用于呼吸门控。
  • 平均每例检查有1.2个床位R≥15。
  • R≥15的波形中90%的床位中心位于肝脏穹顶上方5.6 cm至下方27 cm范围内。
  • 在预期呼吸运动极小的区域,R倾向于<6且S倾向于0。

关键数值与推荐

  • 视觉评分采用3分制:0=无呼吸信号,1=有呼吸样信号但不确定,2=可接受的呼吸信号。
  • 信噪比指标R≥15的波形中86%被评分为可接受。

临床怎么用

  • 数据驱动呼吸门控可用于常规PET成像,尤其胸腹部床位,有助于提高病灶SUVmax。
  • 可利用R值辅助判断是否需要应用呼吸门控,R≥15提示可能存在有效呼吸信号。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

BACKGROUND: We aimed to evaluate the clinical robustness of a commercially developed data-driven respiratory gating algorithm based on principal component analysis, for use in routine PET imaging. METHODS: One hundred fifty-seven adult FDG PET examinations comprising a total of 1149 acquired bed positions were used for the assessment. These data are representative of FDG scans currently performed at our institution. Data were acquired for 4 min/bed position (3 min/bed for legs). The data-driven gating (DDG) algorithm was applied to each bed position, including those where minimal respiratory motion was expected. The algorithm provided a signal-to-noise measure of respiratory-like frequencies within the data, denoted as R. Qualitative evaluation was performed by visual examination of the waveforms, with each waveform scored on a 3-point scale by two readers and then averaged (score S of 0 = no respiratory signal, 1 = some respiratory-like signal but indeterminate, 2 = acceptable signal considered to be respiratory). Images were reconstructed using quiescent period gating and compared with non-gated images reconstructed with a matched number of coincidences. If present, the SUVmax of a well-defined lesion in the thorax or abdomen was measured and compared between the two reconstructions. RESULTS: There was a strong (r = 0.86) and significant correlation between R and scores S. Eighty-six percent of waveforms with R ≥ 15 were scored as acceptable for respiratory gating. On average, there were 1.2 bed positions per patient examination with R ≥ 15. Waveforms with high R and S were found to originate from bed positions corresponding to the thorax and abdomen: 90% of waveforms with R ≥ 15 had bed centres in the range 5.6 cm superior to 27 cm inferior from the dome of the liver. For regions where respiratory motion was expected to be minimal, R tended to be < 6 and S tended to be 0. The use of DDG significantly increased the SUVmax of focal lesions, by an average of 11% when considering lesions in bed positions with R ≥ 15. CONCLUSIONS: The majority of waveforms with high R corresponded to the part of the patient where respiratory motion was expected. The waveforms were deemed suitable for respiratory gating when assessed visually, and when used were found to increase SUVmax in focal lesions.

来源:EJNMMI Res,查看原文。版权归原作者所有。

引用

  1. Walker MD, Morgan AJ, Bradley KM, et al. Evaluation of data-driven respiratory gating waveforms for clinical PET imaging. EJNMMI Res. 2019. PMID: 30607651. doi:10.1186/s13550-018-0470-9

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。