数字与模拟PET/CT:目标病灶及参考区SUVmax的个体内比较
本研究纳入100例肿瘤患者,同日先后行数字与模拟PET/CT,比较目标病灶及参考区SUVmax。数字PET/CT的目标病灶和纵隔血池SUVmax显著高于模拟PET/CT,肝脏SUVmax无显著差异,提示随访中互换两种系统需谨慎。
要点
- 数字PET/CT的目标病灶SUVmax显著高于模拟PET/CT(9.97±6.14 vs 8.14±6.39,P=0.000)。
- 纵隔血池SUVmax数字显著高于模拟(2.54±0.74 vs 2.30±0.67,P=0.000),肝脏无显著差异。
- 随访中互换数字与模拟PET/CT系统时需考虑SUVmax差异。
核心结论
- 数字PET/CT的光子计数技术及延迟摄取增加效应会显著提高SUVmax,随访中互换使用两种系统前需考虑此差异。
主要发现
- 总样本中目标病灶SUVmax模拟为8.14±6.39,数字为9.97±6.14(P=0.000)。
- 总样本中肝脏SUVmax模拟为4.39±2.59,数字为4.46±3.18(P=0.477)。
- 总样本中纵隔血池SUVmax模拟为2.30±0.67,数字为2.54±0.74(P=0.000)。
- 先模拟组目标病灶SUVmax模拟6.64±4.71,数字9.48±5.60(P=0.000);先数字组目标病灶SUVmax数字10.63±6.88,模拟10.16±7.76(P=0.046)。
- 两种时间延迟间未发现显著差异。
临床怎么用
- 在随访研究中若需互换数字与模拟PET/CT系统,应认识到数字系统可能使目标病灶和纵隔血池SUVmax升高。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
PURPOSE: The purpose of this study was to assess whether digital photon counting technology in digital PET/CT influences the quantification of SUVmax in target lesions and regions of reference compared to analog PET/CT before an interchangeable use of either system in follow up studies. METHODS: From January to June of 2018, 100 oncological patients underwent successive PET/CT imaging with digital and analog systems in the same day. Fifty-eight patients underwent analog imaging first and digital imaging thereafter, and 42 patients the other way round. SUVmax was measured in reference regions (liver and mediastinal blood pool) and in the most metabolically active target lesion in each patient. According to the sequence order of PET/CT acquisition, two groups of SUVmax values were obtained, i.e. group 1: analog PET/CT performed first; group 2: digital PET/CT performed first. RESULTS: Mean SUVmax in the total sample (regardless of the order of PET/CT acquisition) in the target lesions with the analog PET/CT was 8.14 ± 6.39 and the digital 9.97 ± 6.14 (P = 0.000). Total mean SUVmax in the liver with the analog was 4.39 ± 2.59 and the digital 4.46 ± 3.18 (P = 0.477). Total mean SUVmax in the mediastinal blood pool with the analog was 2.30 ± 0.67 and the digital 2.54 ± 0.74 (P = 0.000). Group 1: mean SUVmax in the target lesions with the analog system was 6.64 ± 4.71 and the digital 9.48 ± 5.60 (P = 0.000). Mean liver SUVmax with the analog was 4.70 ± 2.90 and the digital 4.80 ± 3.72 (P = 0.088). Mediastinal blood pool SUVmax with the analog was 2.33 ± 0.66 and the digital 2.45 ± 0.73 (P = 0.041). Group 2: mean SUVmax in target lesions with the digital system was 10.63 ± 6.88 and the analog 10.16 ± 7.76 (P = 0.046). Mean liver SUVmax with the digital was 3.99 ± 2.20 and the analog 3.96 ± 2.04 (P = 0.218). Mediastinal blood pool SUVmax with the digital was 2.66 ± 0.75 and the analog 2.27 ± 0.68 (P = 0.000). No significant differences between both time delays were found. CONCLUSIONS: Improved photon counting technology in the digital PET/CT, and the effect of delayed increased uptake and retention significantly increases SUVmax values. This has to be taken into account before interchangeable use of either system in follow up studies.
引用
- Fuentes-Ocampo F, López-Mora DA, Flotats A, et al. Digital vs. analog PET/CT: intra-subject comparison of the SUVmax in target lesions and reference regions. Eur J Nucl Med Mol Imaging. 2019. PMID: 30617960. doi:10.1007/s00259-018-4256-0