文献速递肿瘤 PET/CT放射性药物与辐射防护

数字Biograph Vision PET/CT系统的性能特征

Performance Characteristics of the Digital Biograph Vision PET/CT System.
PubMed 文献 · J Nucl Med · 2019-01-10

本研究依据NEMA NU 2-2012标准评估了数字Biograph Vision PET/CT系统的性能,并补充了新NEMA NU 2-2018标准的额外测量。结果显示其空间分辨率、灵敏度和噪声等效计数率等指标优于前代模拟系统Biograph mCT Flow,且符合当前及未来的EARL性能标准。

要点

  • 数字Biograph Vision PET/CT在NEMA标准下空间分辨率达3.6 mm(横向)和3.5 mm(轴向)。
  • 系统灵敏度为16.4 kcps/MBq,峰值噪声等效计数率为306 kcps(32 kBq/mL)。
  • 该系统性能优于前代模拟系统Biograph mCT Flow,并满足EARL协调标准。

核心结论

  • Biograph Vision数字PET/CT系统性能优于模拟Biograph mCT Flow,且符合欧洲协调性能标准。

研究设计与样本

  • 依据NEMA NU 2-2012标准对Biograph Vision数字PET/CT系统进行性能评估,并补充新NEMA NU 2-2018标准的额外测量。
  • 系统采用硅光电倍增管探测器,3.2 mm硅酸镥晶体,轴向视野26.1 cm。
  • 测量内容包括空间分辨率、灵敏度、计数率性能、衰减与散射校正准确性、图像质量、配准精度和飞行时间性能。

主要发现

  • 横向和轴向空间分辨率分别为3.6 mm和3.5 mm(半高宽,距视野中心1 cm偏移处,使用22Na 0.25 mm点源测量)。
  • NEMA灵敏度为16.4 kcps/MBq,NEMA峰值噪声等效计数率为306 kcps(32 kBq/mL)。
  • 飞行时间分辨率随计数率增加至峰值噪声等效计数率时在210至215之间变化。
  • NEMA图像质量体模的整体图像对比度范围为77.2%至89.8%。

关键数值与推荐

  • 系统符合当前和未来的EARL性能标准。

临床怎么用

  • 该数字PET/CT系统性能优于前代模拟系统,可满足欧洲协调化性能标准,适用于临床PET成像。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

This study evaluated the performance of the Biograph Vision digital PET/CT system according to the NEMA NU 2-2012 standard (published by the National Electrical Manufacturers Association [NEMA]) to allow for a reliable, reproducible, and intersystem-comparable performance measurement. Methods: The new digital PET/CT system features silicon photomultiplier-based detectors with 3.2-mm lutetium oxyorthosilicate crystals and full coverage of the scintillator area. The PET components incorporate 8 rings of 38 detector blocks, and each block contains 4 × 2 mini blocks. Each mini block consists of a 5 × 5 lutetium oxyorthosilicate array of 3.2 × 3.2 × 20 mm crystals coupled to a silicon photomultiplier array of 16 × 16 mm, resulting in an axial field of view of 26.1 cm. In this study, PET/CT system performance was evaluated for conformation with the NEMA NU 2-2012 standard, with additional measurements described in the new NEMA NU 2-2018 standard. Spatial resolution, sensitivity, count-rate performance, accuracy of attenuation and scatter correction, image quality, coregistration accuracy, and time-of-flight performance were determined. Measurements were directly compared with results from its predecessor, the Biograph mCT Flow, using existing literature. Moreover, feasibility to comply with the European Association of Nuclear Medicine Research Ltd. (EARL) criteria was evaluated, and some illustrative patient PET images were obtained. Results: The Biograph Vision showed a transverse and axial spatial resolution of 3.6 and 3.5 mm, respectively, in full width at half maximum at a 1-cm offset from the center of the field of view (measured with a 22Na 0.25-mm point source), a NEMA sensitivity of 16.4 kcps/MBq, and a NEMA peak noise-equivalent count-rate of 306 kcps at 32 kBq/mL. Time-of-flight resolution varied from 210 to 215 as count-rate increased up to the peak noise-equivalent count-rate. The overall image contrast seen with the NEMA image quality phantom ranged from 77.2% to 89.8%. Furthermore, the system was able to comply with the current and future EARL performance criteria. Conclusion: The Biograph Vision outperforms the analog Biograph mCT Flow, and the system is able to meet European harmonizing performance standards.

来源:J Nucl Med,查看原文。版权归原作者所有。