文献速递肿瘤 PET/CT放射性药物与辐射防护

不同混合PET/CT系统间的SUV协调

SUV Harmonization Between Different Hybrid PET/CT Systems.
PubMed 文献 · Clin Nucl Med · 2018-01-01

研究在LSO晶体PSF+TOF与BGO晶体无PSF/TOF的两台PET/CT间测试EQ.filter算法协调SUV,体模及7例肿瘤患者39个高活性病灶验证显示SUVmax差异从高达149%降至10%以下,单次重建即可替代EANM/EARL要求的二次重建。

要点

  • PSF+TOF的LSO PET/CT对小病灶对比度提升,恢复系数高于3D-OSEM重建。
  • 两系统间SUVmax差异最高达149%,经EQ.filter优化后降至10%以下。
  • EQ.filter单次重建即可协调SUV,避免EANM/EARL要求的额外平滑滤波二次重建。

核心结论

  • 经EQ.filter优化的单次重建可最大化病灶空间可检测性,是协调不同PET/CT扫描仪间SUV的简便精确方案,可避免EANM/EARL要求的第二次平滑滤波重建。

研究设计与样本

  • 体模与在体患者数据分析,在Siemens Biograph mCT(LSO晶体,PSF和TOF算法)与GE Discovery STE(BGO晶体,无PSF和TOF)两台系统上采集分析。
  • 测试专用算法EQ.filter按EANM/EARL规范协调两扫描仪间SUV,使用NEMA IQ体模和Jaszczak圆柱体模(含小可填充球,病灶本底比8:1和4:1)。
  • 体模数据在7例肿瘤患者39个高活性病灶(3 mm至26 mm)上验证。

主要发现

  • PSF+TOF的LSO PET/CT系统主要优势为小活性病灶对比度增加。
  • 按NEMA标准测量的恢复系数超过3D-OSEM重建所得。
  • 两PET/CT系统间SUVmax差异最高达149%,经EQ.filter优化后降至10%以下。

关键数值与推荐

  • 体模病灶本底比为8:1和4:1。
  • 患者病灶范围为3 mm至26 mm,共39个高活性病灶。
  • 两系统SUVmax差异最高达149%,EQ.filter优化后低于10%。

临床怎么用

  • 原文结论认为EQ.filter优化的单次重建是协调不同PET/CT扫描仪间SUV的简便精确方案,可避免EANM/EARL要求的第二次平滑滤波重建。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

AIM: Current PET/CT lutetium oxyorthosilicate (LSO) or lutetium-yttrium oxyorthosilicate crystal scanners are equipped with sophisticated softwares including point-spread function (PSF) and time-of-flight (TOF) image reconstruction. These softwares are associated with increased SUVs compared with 3D-OSEM reconstructions associated with older BGO PET/CT scanners. The European Association of Nuclear Medicine (EANM) identified the problem of SUV harmonization since 2010 through the EANM Research Ltd European Association of Nuclear Medicine Research Ltd [EARL] FDG PET/CT accreditation program. This required processing 2 reconstructions, one optimized for maximum spatial lesion detection and one for harmonized quantitation. We investigated an alternative single reconstruction method for both qualitative and quantitative analysis optimized to maximize spatial lesion detectability, followed by an intrinsic postreconstruction algorithm for SUV harmonization. METHODS: Phantom and "in vivo" patient data analysis were acquired and analyzed on (a) a Siemens Biograph mCT system with LSO crystals and PSF and TOF algorithms, and on (b) a General Electric Discovery STE system with BGO crystals, without PSF and TOF. A dedicated algorithm (EQ.filter) was tested to harmonize SUV between the 2 scanners compared with EANM/EARL specifications. NEMA IQ phantom and a Jaszczak cylindrical phantom equipped with small fillable spheres (lesion to background ratios of 8:1 and 4:1) were used. Phantom data were validated on 7 oncologic patients with 39 hyperactive lesions ranging from 3 mm to 26 mm. RESULTS: The main benefit of PSF + TOF LSO PET/CT systems was increased contrast for small active lesions. Recovery coefficients measured according to NEMA standards exceeded those obtained by 3D-OSEM reconstruction. SUVmax discrepancies between the 2 PET/CT systems were as high as 149%, dropping to below 10% when optimized by EQ.filter. CONCLUSIONS: A single reconstruction optimized by EQ.filter for maximum spatial lesion detectability is an easy and precise solution to harmonize SUVs between different PET/CT scanners, avoiding a second reconstruction with an additional smoothing filter as requested by EANM/EARL.

来源:Clin Nucl Med,查看原文。版权归原作者所有。

引用

  1. Rubello D, Colletti PM. SUV Harmonization Between Different Hybrid PET/CT Systems. Clin Nucl Med. 2018. PMID: 30199381. doi:10.1097/RLU.0000000000002284

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。