文献速递心脏核医学

人体测量与循环变量对CZT SPECT门控核素血管造影计数率的影响

The influence of anthropometric and basic circulatory variables on count rate in cadmium-zinc-telluride SPECT gated radionuclide angiography.
PubMed 文献 · J Nucl Cardiol · 2018-08-17

研究纳入2015年8月至2016年11月共1065例常规评估左心室射血分数的患者,在GE Discovery 530c专用心脏CZT SPECT相机上静脉注射550 MBq 99mTc标记人血清白蛋白后进行核素血管造影,记录心率、收缩压、舒张压、年龄、性别、身高、体重与计数率。结果显示计数率范围为1.2至8.9 counts per second,所有测试变量均与计数率显著相关,体重是解释计数率变异的主要贡献因素(R2=0.68),由体重、身高、性别和年龄组成的模型可解释计数率变异的75%。作者认为身高、体重、性别和年龄对CZT核素血管造影计数率有显著影响,可能后续用于个体化规划示踪剂剂量。

要点

  • 1065例患者中计数率范围为1.2至8.9 counts per second。
  • 体重是解释计数率变异的主要贡献因素,R2=0.68。
  • 体重、身高、性别和年龄模型可解释计数率变异的75%。

核心结论

  • 患者身高、体重、性别和年龄对CZT核素血管造影计数率有显著影响,可能后续用于个体化规划示踪剂剂量。

研究设计与样本

  • 2015年8月至2016年11月共登记1065例符合条件、转诊进行左心室射血分数常规评估的患者。
  • 记录心率、收缩压、舒张压、年龄、性别、身高、体重和计数率,所有核素血管造影均在专用心脏CZT SPECT相机(GE Discovery 530c)上完成。
  • 每例患者静脉给予550 MBq 99mTc标记人血清白蛋白。

主要发现

  • 计数率范围为1.2至8.9 counts per second。
  • 所有测试变量均与计数率显著相关。
  • 初步分析中体重是解释计数率变异的主要贡献因素,R2=0.68。
  • 由体重、身高、性别和年龄组成的模型解释计数率变异的75%。

关键数值与推荐

  • 每例患者静脉给予550 MBq 99mTc标记人血清白蛋白。
  • 体重单独解释计数率变异的R2=0.68。
  • 体重、身高、性别和年龄模型解释计数率变异的75%。

临床怎么用

  • 原文结论指出患者身高、体重、性别和年龄对CZT核素血管造影计数率有显著影响,并可能后续用于个体化规划示踪剂剂量。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

AIM: To determine the influence of weight, height, gender, age, heart rate, and blood pressure on count rate in cadmium-zinc-telluride single-photon emission computed tomography (CZT SPECT) gated radionuclide angiography. METHOD: A total of 1,065 eligible patients referred for routine assessment of left ventricular ejection fraction were registered from August 2015 to November 2016. Data were recorded on heart rate, systolic and diastolic blood pressure, age, gender, height, weight, and count rate. All radionuclide angiographies were performed on a dedicated cardiac CZT SPECT camera, GE Discovery 530c. A dose of 550 MBq 99mTc-labeled human serum albumin was administered intravenously to each patient. RESULTS: Count rate varied from 1.2 to 8.9 counts per second. All test variables were significantly associated with count rate. From the preliminary analysis, weight appeared as the main contributing factor for explaining the variations in count rate with and R2 of 0.68. A model consisting of weight, height, gender, and age explains 75% of the variance in count rate. CONCLUSION: Patient height, weight, gender, and age have significant impact on count rate when performing CZT radionuclide angiography and may subsequently be used for individualized planning of tracer dosage.

来源:J Nucl Cardiol,查看原文。版权归原作者所有。

引用

  1. Rydberg J, Andersen J, Haarmark C, et al. The influence of anthropometric and basic circulatory variables on count rate in cadmium-zinc-telluride SPECT gated radionuclide angiography. J Nucl Cardiol. 2018. PMID: 30120749. doi:10.1007/s12350-018-1402-9

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。