InSPira HD与NeuroFocus脑专用SPECT系统对[123I]FP-CIT显像的头对头临床评估
本研究将新型脑专用SPECT扫描仪InSPira HD与经验证的NeuroFocus系统进行头对头比较,使用纹状体体模和14例接受[123I]FP-CIT注射的受试者进行扫描。纹状体比值在体模和患者数据中均显示强相关(R=0.99和R=0.92),14例中13例视觉评估一致,作者认为[123I]FP-CIT SPECT显像可在InSPira系统上充分完成。
要点
- 纹状体体模与患者数据的纹状体比值在InSPira与NeuroFocus间呈强相关。
- 14例[123I]FP-CIT扫描中13例视觉评估在两系统间一致。
- NeuroFocus患者数据比值略高,可能源于系统性能差异。
核心结论
- 作者认为[123I]FP-CIT SPECT显像可在新型InSPira系统上充分完成。
研究设计与样本
- 在InSPira与经验证的脑专用NeuroFocus系统上对纹状体体模(充填[123I])和14例受试者(注射[123I]FP-CIT后)进行头对头扫描比较。
主要发现
- 纹状体体模的纹状体比值两系统强相关(R=0.99,p<0.05)。
- 患者数据的纹状体比值两系统强相关(R=0.92,p<0.01)。
- 患者数据中NeuroFocus的比值略高于InSPira,可能源于系统性能差异。
- 14例[123I]FP-CIT扫描的视觉评估中13例两系统一致。
关键数值与推荐
- 纹状体比值由BRASS软件包自动计算。
临床怎么用
- 原文未据此给出推荐。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
The InSPira HD system, a novel brain-dedicated SPECT scanner, allows for imaging with a high spatial resolution. Here, we tested whether this scanner can be used to image the dopamine transporter adequately. Therefore, striatal phantom and patient data acquired on the InSPira were compared head-to-head with the well-validated brain-dedicated NeuroFocus system. A striatal phantom filled with [123I] and 14 subjects (after [123I]FP-CIT injection) were scanned on both systems. [123I]FP-CIT SPECT scans were visually assessed. Striatal binding ratios were calculated automatically using the software package BRASS. Striatal phantom and patient data showed strong correlations with respect to striatal ratios (R = 0.99 and R = 0.92; p < 0.05 and p < 0.01, respectively). Slightly higher ratios were found for the NeuroFocus patient data, probably due to differences in system performance. Visual assessment of [123I]FP-CIT scans showed agreement between systems in 13 of the 14 cases. We conclude that [123I]FP-CIT SPECT imaging can be performed adequately on the new InSPira system.
引用
- Adriaanse SM, de Wit TC, Stam M, et al. Clinical evaluation of [123I]FP-CIT SPECT scans on the novel brain-dedicated InSPira HD SPECT system: a head-to-head comparison. EJNMMI Res. 2018. PMID: 30136081. doi:10.1186/s13550-018-0436-y