新型弹性呼吸运动校正算法在腹部胸部PET/CT定量与图像质量中的评估

Evaluation of a novel elastic respiratory motion correction algorithm on quantification and image quality in abdomino-thoracic PET/CT.
PubMed 文献 · J Nucl Med · 2018-08-16

该研究在体模和患者中评估了新开发的弹性运动去模糊(EMDB)技术,并与相位门控(PBG)和HDChest(HDC)及静态全身(SWB)图像比较。体模中EMDB的对比噪声比(RCNR)最高(0.68),患者中仅EMDB使肺/肝肿瘤CNR增加(18/13%),图像噪声增加最少,主观图像质量评分仅次于SWB。

要点

  • EMDB使用全部采集数据,体模中RCNR为0.68,高于HDC(0.54)、PBGmax(0.41)和PBGave(0.31)。
  • 患者中仅EMDB使肺/肝肿瘤CNR增加(18/13%),HDC、PBGave和PBGmax均降低。
  • EMDB的肺/肝噪声增加最少(3/15%),主观图像质量评分3.7±1.0,低于SWB的4.0±0.8但高于其他方法。

核心结论

  • EMDB算法图像噪声增加最少,改善了病灶CNR,并具有最高的总体图像质量评分。

研究设计与样本

  • 体模研究模拟胸部呼吸运动,含5个内径10-28 mm的肺病灶球体,在0、1、2、3 cm运动幅度下扫描,进行SWB、PBG、HDC和EMDB四种重建。
  • 患者研究纳入46例肺或肝肿瘤直径<3 cm的患者,测量46个肺病灶和19个肝病灶的SUVmax、SUVpeak和CNR。
  • 由四名放射科医生对患者图像主观质量进行5分制评分。

主要发现

  • 体模中EMDB的RCNR为0.68,高于HDC的0.54、PBGmax的0.41和PBGave的0.31。
  • 患者中与SWB相比,SUVmax/SUVpeak分别增加:EMDB 19.3/11.1%,HDC 21.6/13.9%,PBGave 22.8/12.8%,PBGmax 45.6/26.8%。
  • 肺/肝噪声增加:EMDB 3/15%,HDC 35/56%,PBGave 100/170%,PBGmax 146/219%。
  • 肺/肝肿瘤CNR变化:EMDB增加18/13%,HDC降低14/23%,PBGave降低39/63%,PBGmax降低18/46%。
  • 平均图像质量评分:SWB 4.0±0.8 > EMDB 3.7±1.0 > HDC 3.1±1.0 > PBG 1.5±0.7。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

Our aim is to evaluate in phantom and patient studies a recently developed elastic motion debluring (EMDB) technique which makes use of all the acquired PET data and compare its performance to other conventional techniques such as phase based gating (PBG) and HDChest (HDC) both of which use fractions of the acquired data. Comparisons were made with respect to static whole-body (SWB) images with no motion correction. Methods: A phantom simulating respiratory motion of the thorax with lung lesions (5 spheres with ID=10- 28 mm) was scanned with 0, 1, 2, and 3 cm motion. Four reconstructions were performed: SWB, PBG, HDC, and EMDB. For PBG, the average (PBGave) and maximum bin (PBGmax) were used. To compare the reconstructions, the ratios of SUVmax (RSmax), SUVpeak (RSpeak), and CNR (RCNR) were calculated with respect to SWB. Additionally, 46 patients with lung or liver tumors < 3 cm diameter were also studied. Measurements of SUVmax, SUVpeak, and contrast-to-noise ratio (CNR) were made for 46 lung and 19 liver lesions. To evaluate image noise, the SUV standard deviation was measured in healthy lung and liver tissue and in the phantom background. Finally, subjective image quality of patient exams was scored on a 5 point scale by four radiologists. Results: In the phantom, EMDB increased SUVmax/SUVpeak over SWB but to a lesser extent than the other reconstruction methodologies. The RCNR for EMDB however was higher than all other reconstructions (0.68 EMDB > 0.54 HDC > 0.41 PBGmax > 0.31 PBGave). Similar results were seen in patient studies. The SUVmax/SUVpeak were higher by 19.3/11.1% EMDB, 21.6/13.9% HDC, 22.8/12.8% PBGave, and 45.6/26.8% PBGmax compared to SWB. Lung/liver noise increased EMDB (3/15%), HDC (35/56%), PBGave (100/170%), and PBGmax (146/219%). CNR increased in lung/liver tumors only for EMDB (18/13%), and decreased for HDC (-14/-23%), PBGave (-39/-63%), and PBGmax (-18/-46%). Average radiologist scores of image quality were SWB (4.0 ± 0.8) > EMDB (3.7 ± 1.0) > HDC (3.1 ± 1.0) > PBG (1.5 ± 0.7). Conclusion: The EMDB algorithm had the least increase in image noise, improved lesion CNR, and had the highest overall image quality score.