J-Hart多中心研究:使用通用软件对PET SUV进行协调
J-Hart研究纳入9家机构的12台PET扫描仪,使用NEMA体模和通用软件进行SUV协调。三种扫描仪无需额外高斯滤波,其余九种需加5至9毫米半高全宽的高斯滤波。协调后SUVmax重现性(P=0.0055)和重复性(P<0.0001)均显著改善。
要点
- 12台PET扫描仪中,3台无需额外高斯滤波,9台需加5至9毫米半高全宽的高斯滤波。
- 协调后SUVmax重现性变异系数由9.45%±4.69%降至6.05%±3.61%(P=0.0055)。
- 协调后SUVmax重复性变异系数由6.59%±1.29%降至4.88%±1.64%(P<0.0001)。
主要发现
- 12台PET扫描仪中,3台无需额外高斯滤波进行协调,其余9台需额外高斯滤波,半高全宽值为5至9毫米。
- 协调前六个球体的SUVmax重现性变异系数为9.45%±4.69%(范围3.83-15.3%),协调后为6.05%±3.61%(范围2.30-10.7%),协调显著改善重现性(P=0.0055)。
- 协调前九台扫描仪的SUVmax重复性变异系数为6.59%±1.29%(范围5.00-8.98%),协调后为4.88%±1.64%(范围2.65-6.72%),协调显著改善重复性(P<0.0001)。
关键数值与推荐
- 每个扫描仪需确定高斯滤波的半高全宽参数以实现协调,参数范围为5至9毫米(针对需额外滤波的九台扫描仪)。
- SUVmax协调目标为日本核医学会提出的NEMA体模参考范围。
临床怎么用
- 原文未据此给出推荐。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
BACKGROUND: Recent developments in hardware and software for PET technologies have resulted in wide variations in basic performance. Multicentre studies require a standard imaging protocol and SUV harmonization to reduce inter- and intra-scanner variability in the SUV. The Japanese standardised uptake value (SUV) Harmonization Technology (J-Hart) study aimed to determine the applicability of vendor-neutral software on the SUV derived from positron emission tomography (PET) images. The effects of SUV harmonization were evaluated based on the reproducibility of several scanners and the repeatability of an individual scanner. Images were acquired from 12 PET scanners at nine institutions. PET images were acquired over a period of 30 min from a National Electrical Manufacturers Association (NEMA) International Electrotechnical Commission (IEC) body phantom containing six spheres of different diameters and an 18F solution with a background activity of 2.65 kBq/mL and a sphere-to-background ratio of 4. The images were reconstructed to determine parameters for harmonization and to evaluate reproducibility. PET images with 2-min acquisition × 15 contiguous frames were reconstructed to evaluate repeatability. Various Gaussian filters (GFs) with full-width at half maximum (FWHM) values ranging from 1 to 15 mm in 1-mm increments were also applied using vendor-neutral software. The SUVmax of spheres was compared with the reference range proposed by the Japanese Society of Nuclear Medicine (JSNM) and the digital reference object (DRO) of the NEMA phantom. The coefficient of variation (CV) of the SUVmax determined using 12 PET scanners (CVrepro) was measured to evaluate reproducibility. The CV of the SUVmax determined from 15 frames (CVrepeat) per PET scanner was measured to determine repeatability. RESULTS: Three PET scanners did not require an additional GF for harmonization, whereas the other nine required additional FWHM values of GF ranging from 5 to 9 mm. The pre- and post-harmonization CVrepro of six spheres were (means ± SD) 9.45% ± 4.69% (range, 3.83-15.3%) and 6.05% ± 3.61% (range, 2.30-10.7%), respectively. Harmonization significantly improved reproducibility of PET SUVmax (P = 0.0055). The pre- and post-harmonization CVrepeat of nine scanners were (means ± SD) 6.59% ± 1.29% (range, 5.00-8.98%) and 4.88% ± 1.64% (range, 2.65-6.72%), respectively. Harmonization also significantly improved the repeatability of PET SUVmax (P < 0.0001). CONCLUSIONS: Harmonizing SUV using vendor-neutral software produced SUVmax for 12 scanners that fell within the JSNM reference range of a NEMA body phantom and improved SUVmax reproducibility and repeatability.
引用
- Tsutsui Y, Daisaki H, Akamatsu G, et al. Multicentre analysis of PET SUV using vendor-neutral software: the Japanese Harmonization Technology (J-Hart) study. EJNMMI Res. 2018. PMID: 30128776. doi:10.1186/s13550-018-0438-9