177Lu SPECT/CT定量重建xSPECT Quant的噪声与分辨率特征

Characterization of Noise and Resolution for Quantitative 177Lu SPECT/CT with xSPECT Quant.
PubMed 文献 · J Nucl Med · 2018-07-05

该研究在填充177Lu的体模中评估xSPECT Quant与Flash3D两种重建算法的噪声行为、活度测定准确性、恢复系数和分辨率。xSPECT Quant的最优参数为48次迭代、1个子集,在Jaszczak圆柱体中定量准确性达1.2%。若重建参数选择得当,两种算法均可提供绝对定量SPECT图像,xSPECT Quant的标准化活度测定无需各中心单独校准。

要点

  • xSPECT Quant在48次迭代、1个子集时达到噪声与分辨率的最佳平衡,Jaszczak圆柱体定量准确性为1.2%。
  • 两种重建算法均以噪声增加为代价改善空间分辨率,部分容积效应仍存在。
  • xSPECT Quant的标准化活度测定无需各中心单独校准,适用于多中心分子放疗研究。

核心结论

  • 若重建参数选择得当,xSPECT Quant与Flash3D均可提供具有高图像质量(亚厘米分辨率、可接受的噪声增加)和高定量准确性的绝对定量SPECT图像,其中Flash3D需有良好校准的计数至活度浓度转换。
  • xSPECT Quant凭借标准化且可溯源的活度测定,无需各中心单独校准协议,可实现跨中心可比的可靠活度测定,尤其适用于多中心分子放疗研究。

研究设计与样本

  • 在系列填充177Lu的体模测量中,对xSPECT Quant与Flash3D的噪声行为、活度测定准确性、恢复系数和分辨率进行评估。
  • 噪声行为与活度测定准确性在大圆柱体中评估,恢复系数在热球体模中有和无背景下评估,分辨率在线源体模及热球-冷背景体模的匹配滤波分辨率分析中确定。

主要发现

  • 两种重建算法均以噪声增加为代价改善空间分辨率。
  • Flash3D收敛较慢,但恢复效率更高。
  • 两种重建虽均应用分辨率恢复方法,部分容积误差(物体中心活度高估和物体边缘计数外溢)仍具有相关性。
  • Flash3D仅总更新次数(迭代次数×子集数)起作用,而xSPECT Quant的迭代与子集的具体细分影响其所有特性(最优为1个子集)。
  • xSPECT Quant在48次迭代和1个子集时达到噪声增加与分辨率改善的最佳平衡,在Jaszczak圆柱体中定量准确性为1.2%(xSPECT Quant经剂量校准器交叉校准)。

关键数值与推荐

  • xSPECT Quant的最优重建参数为48次迭代和1个子集。
  • xSPECT Quant在Jaszczak圆柱体中的定量准确性为1.2%(经剂量校准器交叉校准)。

临床怎么用

  • xSPECT Quant的标准化且可溯源的活度测定无需各中心单独校准协议,可实现跨中心可比的可靠活度测定,尤其适用于多中心分子放疗研究。
  • Flash3D需有良好校准的计数至活度浓度转换,方可提供高定量准确性的绝对定量SPECT图像。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

Quantitative SPECT/CT imaging forms the basis for internal dosimetry in molecular radiotherapies. While the conversion from counts to activity is typically based on conversion factors individually measured by each site, a recently introduced commercially available reconstruction (xSPECT Quant) offers a standardized and traceable calibration of SPECT/CT systems. The aim of this work was to assess the characteristics of xSPECT Quant in combination with 177Lu as one of the most important radionuclides used in molecular radiotherapies and to compare it to a widely used ordered-subset expectation-maximization reconstruction (Flash3D). Methods: In a series of 177Lu-filled phantom measurements, several important features were investigated for xSPECT Quant and Flash3D. Noise behavior and accuracy of the activity determination were evaluated in a large cylinder. Recovery coefficients were assessed in a hot-sphere phantom with and without background. Additionally, the resolutions were determined in a line source phantom as well as in a matched-filter resolution analysis of the hot-sphere-cold-background phantom. Results: Both reconstruction algorithms improve the spatial resolution at the cost of noise build-up. Despite its slower convergence, Flash3D features a more efficient recovery. Although resolution recovery methods are applied within both reconstructions, partial-volume errors-namely activity overestimation in the object center and spill-out of counts from the object edges-remain of relevance. In contrast to Flash3D, for which only the total number of updates (iterations × subsets) plays a role, the exact subdivision into iterations and subsets affected all characteristics of xSPECT Quant (optimum, 1 subset). The optimal trade-off between noise build-up and resolution improvement was found for 48 iterations and 1 subset, resulting in a quantitative accuracy of 1.2% in the Jaszczak cylinder (xSPECT Quant cross-calibrated to the dose calibrator). Conclusion: If the reconstruction parameters are chosen with care, both examined reconstructions can provide absolute quantitative SPECT images with high image quality (subcentimeter resolution at an acceptable noise build-up) as well as high quantitative accuracy (given a well-calibrated Flash3D conversion from counts to activity concentration). With its standardized (and traceable) activity determination, xSPECT Quant dispenses with site-specific calibration protocols, enabling a reliable activity determination comparable across sites, which is especially useful for multicentric molecular radiotherapy studies.