文献速递肿瘤 PET/CT

临床68Ga-PSMA PET/MR中BSREM重建的定量性能与最佳正则化参数

Quantitative performance and optimal regularization parameter in block sequential regularized expectation maximization reconstructions in clinical 68Ga-PSMA PET/MR.
PubMed 文献 · EJNMMI Res · 2018-07-27

本研究回顾性重建25例患者盆腔区域2分钟发射数据,比较TOF BSREM与TOF OSEM在68Ga-PSMA PET/MR中的表现。共识别40个病灶,BSREM在β值400–550时可能在高病灶SUVmax与低背景噪声间取得最佳折中。

要点

  • BSREM重建后背景噪声和病灶SUVmax均随β值升高而降低。
  • β值低于400时背景噪声高于参考OSEM(p<0.01),实用性较低。
  • β值高于600时低活性病灶SUVmax低于参考(p<0.05),小低摄取病灶可能不可见。

核心结论

  • 本研究表明,TOF BSREM联合68Ga-PSMA示踪剂重建的PET图像较TOF OSEM背景噪声更低、病灶SUVmax更高。
  • 本研究提示β值在400至550之间可能是高SUVmax与低背景噪声之间的最佳折中。

研究设计与样本

  • 回顾性重建25例接受68Ga-PSMA PET/MR患者盆腔区域2分钟发射数据。
  • 参考OSEM重建采用28个子集和2次迭代;BSREM重建采用15个β值,范围150至1200。
  • 在病灶和均匀背景中勾画感兴趣区(ROI),计算背景SUVmean(平均值)和SUVstd(标准差),以及病灶SUVmax(5个最热体素的平均值)。
  • 采用Wilcoxon配对符号秩检验分析差异。

主要发现

  • 盆腔区域共识别40个病灶。
  • 背景噪声(SUVstd)和病灶SUVmax随β值升高而降低。
  • β值低于400的重建较参考OSEM重建背景噪声更高(p<0.01),因此实用性较低。
  • β值高于600的重建中低活性病灶SUVmax低于参考(p<0.05),这些重建可能因背景噪声较低而在视觉上更吸引人,但较低的SUVmax可能使小的低摄取病灶不可见。

关键数值与推荐

  • 参考OSEM重建参数为28个子集、2次迭代。
  • BSREM重建采用15个β值,范围150至1200。
  • β值低于400时背景噪声高于参考OSEM(p<0.01)。
  • β值高于600时低活性病灶SUVmax低于参考(p<0.05)。
  • β值在400至550之间可能是高SUVmax与低背景噪声之间的最佳折中。

临床怎么用

  • 原文未据此给出推荐。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

BACKGROUND: In contrast to ordered subset expectation maximization (OSEM), block sequential regularized expectation maximization (BSREM) positron emission tomography (PET) reconstruction algorithms can run until full convergence while controlling image quality and noise. Recent studies with BSREM and 18F-FDG PET reported higher signal-to-noise ratios and higher standardized uptake values (SUV). In this study, we investigate the optimal regularization parameter (β) for clinical 68Ga-PSMA PET/MR reconstructions in the pelvic region applying time-of-flight (TOF) BSREM in comparison to TOF OSEM. Two-minute emission data from the pelvic region of 25 patients who underwent 68Ga-PSMA PET/MR were retrospectively reconstructed. Reference OSEM reconstructions had 28 subsets and 2 iterations. BSREM reconstructions were performed with 15 β values between 150 and 1200. Regions of interest (ROIs) were drawn around lesions and in uniform background. Background SUVmean (average) and SUVstd (standard deviation), and lesion SUVmax (average of 5 hottest voxels) were calculated. Differences were analyzed using the Wilcoxon matched pairs signed-rank test. RESULTS: A total of 40 lesions were identified in the pelvic region. Background noise (SUVstd) and lesions SUVmax decreased with increasing β. Image reconstructions with β values lower than 400 have higher (p < 0.01) background noise, compared to the reference OSEM reconstructions, and are therefore less useful. Lesions with low activity on images reconstructed with β values higher than 600 have a lower (p < 0.05) SUVmax compared to the reference. These reconstructions are likely visually appealing due to the lower background noise, but the lower SUVmax could possibly render small low-uptake lesions invisible. CONCLUSIONS: In our study, we showed that PET images reconstructed with TOF BSREM in combination with the 68Ga-PSMA tracer result in lower background noise and higher SUVmax values in lesions compared to TOF OSEM. Our study indicates that a β value between 400 and 550 might be the optimal compromise between high SUVmax and low background noise.

来源:EJNMMI Res,查看原文。版权归原作者所有。

引用

  1. Ter Voert EEGW, Muehlematter UJ, Delso G, et al. Quantitative performance and optimal regularization parameter in block sequential regularized expectation maximization reconstructions in clinical 68Ga-PSMA PET/MR. EJNMMI Res. 2018. PMID: 30054750. doi:10.1186/s13550-018-0414-4

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。