G8高灵敏度台式临床前PET/CT断层扫描仪的性能评估

Performance Evaluation of G8, a High-Sensitivity Benchtop Preclinical PET/CT Tomograph.
PubMed 文献 · J Nucl Med · 2018-06-14

本研究评估了G8台式集成PET/CT扫描仪(用于小鼠高灵敏度高分辨率成像)的NU 4-2008性能及CT子系统基础成像性能。PET子系统在350-650 keV能量窗下峰值灵敏度为9.0%,晶体能量分辨率均值19.3%±3.7%,平均固有空间分辨率横向1.30 mm、轴向1.38 mm(FWHM),小鼠体模峰值噪声等效计数率为44 kcps(总活度2.9 MBq),散射分数11%。结论认为G8整体性能可为分子影像生物医学研究提供高质量图像。

要点

  • PET子系统在350-650 keV能量窗下峰值灵敏度为9.0%,位于视野中心附近。
  • 晶体能量分辨率范围15.0%至69.6%(FWHM),均值19.3%±3.7%。
  • 平均固有空间分辨率横向1.30 mm、轴向1.38 mm(FWHM),空气中点源重建图像平均0.81±0.11 mm(FWHM)。

核心结论

  • 整体性能表明G8可为基于分子影像的生物医学研究生成高质量图像。

研究设计与样本

  • 本研究表征G8台式集成PET/CT扫描仪的美国国家电气制造商协会NU 4-2008性能(适用处),并评估CT子系统基础成像性能。
  • PET子系统由4个平板探测器以盒状几何排列构成,每个平板含2个26×26像素化锗酸铋闪烁体阵列模块,单个晶体尺寸1.75×1.75×7.2 mm。
  • 晶体阵列通过锥形像素化玻璃光导耦合至多通道光电倍增管;锥束CT扫描仪由微焦点X射线源与互补金属氧化物半导体探测器组成,提供解剖信息。
  • 评估PET子系统的灵敏度、空间分辨率、能量分辨率、散射分数、计数率性能及体模与小鼠成像能力。
  • 评估CT子系统的噪声、剂量水平、对比度与分辨率。

主要发现

  • 在350-650 keV能量窗下,视野中心附近峰值灵敏度为9.0%。
  • 晶体能量分辨率范围为15.0%至69.6%(FWHM),均值19.3%±3.7%。
  • 平均固有空间分辨率横向1.30 mm、轴向1.38 mm(FWHM)。
  • 空气中点源的最大似然期望最大化重建图像平均为0.81±0.11 mm(FWHM)。
  • 小鼠体模峰值噪声等效计数率为44 kcps(总活度2.9 MBq),散射分数为11%;CT子系统调制传递函数在10%处为2.05 cycles/mm。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

G8 is a benchtop integrated PET/CT scanner dedicated to high-sensitivity and high-resolution imaging of mice. This work characterizes its National Electrical Manufacturers Association NU 4-2008 performance where applicable and also assesses the basic imaging performance of the CT subsystem. Methods: The PET subsystem in G8 consists of 4 flat-panel detectors arranged in a boxlike geometry. Each panel consists of 2 modules of a 26 × 26 pixelated bismuth germanate scintillator array with individual crystals measuring 1.75 × 1.75 × 7.2 mm. The crystal arrays are coupled to multichannel photomultiplier tubes via a tapered, pixelated glass lightguide. A cone-beam CT scanner consisting of a MicroFocus x-ray source and a complementary metal oxide semiconductor detector provides anatomic information. Sensitivity, spatial resolution, energy resolution, scatter fraction, count-rate performance, and the capability of performing phantom and mouse imaging were evaluated for the PET subsystem. Noise, dose level, contrast, and resolution were evaluated for the CT subsystem. Results: With an energy window of 350-650 keV, the peak sensitivity was 9.0% near the center of the field of view. The crystal energy resolution ranged from 15.0% to 69.6% in full width at half maximum (FWHM), with a mean of 19.3% ± 3.7%. The average intrinsic spatial resolution was 1.30 and 1.38 mm FWHM in the transverse and axial directions, respectively. The maximum-likelihood expectation maximization reconstructed image of a point source in air averaged 0.81 ± 0.11 mm FWHM. The peak noise-equivalent count rate for the mouse-sized phantom was 44 kcps for a total activity of 2.9 MBq (78 μCi), and the scatter fraction was 11%. For the CT subsystem, the value of the modulation transfer function at 10% was 2.05 cycles/mm. Conclusion: The overall performance demonstrates that the G8 can produce high-quality images for molecular imaging-based biomedical research.