动态99mTc-甲氧基异丁基异腈CZT-SPECT评估心肌血流储备的首次验证:在疑似冠心病患者中与15O-水PET及血流储备分数的头对头比较——WATERDAY研究
研究前瞻性纳入30例疑似稳定冠心病患者,比较动态99mTc-甲氧基异丁基异腈CZT-SPECT与15O-水PET估算心肌血流和血流储备的可行性,并以血流储备分数(FFR≤0.8)为参考。CZT-SPECT与PET的MFR在各区域相似,检测MFR受损的敏感度83.3%、特异度95.8%、准确率93.3%。
要点
- CZT-SPECT估算的负荷和静息心肌血流量在整体、LAD和LCx区域高于PET,RCA区域无此差异。
- CZT-SPECT与PET估算的MFR在整体及各血管区域相似。
- CZT-SPECT检测MFR受损的敏感度83.3%、特异度95.8%、准确率93.3%。
核心结论
- 动态99mTc-甲氧基异丁基异腈CZT-SPECT技术可行,所测MFR与15O-水PET相似,对稳定冠心病患者MFR受损和FFR异常有较高诊断价值。
研究设计与样本
- 前瞻性纳入30例疑似稳定冠心病患者,行主要冠状动脉(LAD、LCx、RCA)FFR测量,并以动态静息/负荷99mTc-甲氧基异丁基异腈CZT-SPECT和15O-水PET计算MBF与MFR。
主要发现
- CZT-SPECT在整体、LAD和LCx区域的负荷及静息MBF高于PET,RCA区域无此差异。
- CZT-SPECT与PET的MFR在整体及各血管区域相似。
- CZT-SPECT检测缺血的敏感度83.3%、特异度95.8%、准确率93.3%、阳性预测值100%、阴性预测值85.7%。
- CZT-SPECT检测血流动力学显著狭窄(FFR≤0.8)的敏感度58.3%、特异度84.6%、准确率81.1%、阳性预测值36.8%、阴性预测值93%。
关键数值与推荐
- 狭窄≥50%视为阻塞性,FFR≤0.8视为异常。
- PET和CZT-SPECT判断缺血分别以MFR低于2和低于2.1为标准。
临床怎么用
- 原文未据此给出推荐。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
PURPOSE: We assessed the feasibility of myocardial blood flow (MBF) and flow reserve (MFR) estimation using dynamic SPECT with a novel CZT camera in patients with stable CAD, in comparison with 15O-water PET and fractional flow reserve (FFR). METHODS: Thirty patients were prospectively included and underwent FFR measurements in the main coronary arteries (LAD, LCx, RCA). A stenosis ≥50% was considered obstructive and a FFR abnormal if ≤0.8. All patients underwent a dynamic rest/stress 99mTc-sestamibi CZT-SPECT and 15O-water PET for MBF and MFR calculation. Net retention kinetic modeling was applied to SPECT data to estimate global uptake values, and MBF was derived using Leppo correction. Ischemia by PET and CZT-SPECT was considered present if MFR was lower than 2 and 2.1, respectively. RESULTS: CZT-SPECT yielded higher stress and rest MBF compared to PET for global and LAD and LCx territories, but not in RCA territory. MFR was similar in global and each vessel territory for both modalities. The sensitivity, specificity, accuracy, positive and negative predictive value of CZT-SPECT were, respectively, 83.3, 95.8, 93.3, 100 and 85.7% for the detection of ischemia and 58.3, 84.6, 81.1, 36.8 and 93% for the detection of hemodynamically significant stenosis (FFR ≤ 0.8). CONCLUSIONS: Dynamic 99mTc-sestamibi CZT-SPECT was technically feasible and provided similar MFR compared to 15O-water PET and high diagnostic value for detecting impaired MFR and abnormal FFR in patients with stable CAD.
引用
- Agostini D, Roule V, Nganoa C, et al. First validation of myocardial flow reserve assessed by dynamic 99mTc-sestamibi CZT-SPECT camera: head to head comparison with 15O-water PET and fractional flow reserve in patients with suspected coronary artery disease. The WATERDAY study. Eur J Nucl Med Mol Imaging. 2018. PMID: 29497801. doi:10.1007/s00259-018-3958-7