使用心脏聚焦准直器评估左心室射血分数:IQ-SPECT、平面平衡法放射性核素血管造影与心脏磁共振的比较
本研究纳入60例患者行IQ-SPECT门控血池SPECT(GBPS)与平面显像,其中11例同时行心脏磁共振(cMRI)。GBPS左心室射血分数(LVEF)与平面测量相关,MHI软件法相关性最佳(r=0.83),QBS法r=0.70;cMRI与平面LVEF相关系数0.65,与QBS法GBPS为0.69,与MHI法GBPS为0.52。结论认为IQ-SPECT GBPS计算的LVEF与平面测量相关,MHI法相关性最佳,变异不依赖左心室舒张末期容积(LVEDV)。
要点
- IQ-SPECT GBPS计算的LVEF与平面平衡法放射性核素血管造影测量相关。
- MHI软件法相关性最佳(r=0.83),QBS法为r=0.70。
- GBPS与平面LVEF的变异不依赖LVEDV。
核心结论
- IQ-SPECT GBPS计算的LVEF与平面测量相关,MHI法相关性最佳,变异不依赖LVEDV。
研究设计与样本
- 60例患者接受IQ-SPECT GBPS与平面显像,其中11例同时接受cMRI与GBPS。
- GBPS的LVEF、LVEDV和LVESV使用QBS与MHI两种经验证软件计算。
- 比较不同模态获得的LVEF、LVEDV和LVESV。
主要发现
- 平面法平均LVEF为48±11%,平均LVEDV为177±59 mL(范围63至342 mL),平均LVESV为96±46 mL(范围16至234 mL)。
- GBPS LVEF在QBS法为40±12%,与平面LVEF相关系数r=0.70;MHI法为44±12%,r=0.83。
- cMRI与平面LVEF相关系数为0.65,与QBS法GBPS为0.69,与MHI法GBPS为0.52。
- GBPS与平面LVEF的变异不依赖LVEDV。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
BACKGROUND: IQ-SPECT has been shown to significantly reduce acquisition time and administered dose while preserving image quality in myocardial perfusion imaging. Whether IQ-SPECT provides accurate left ventricular ejection fractions (LVEF) with gated blood pool SPECT (GBPS) remains unknown. METHODS: Sixty patients underwent IQ-SPECT GBPS and planar imaging. Among those patients, 11 underwent both cMRI and GBPS. GBPS LVEF, LVEDV, and LVESV were calculated using 2 validated software; QBS (Cedars-Sinai Medical Center, Los Angeles, USA) and MHI (Montreal Heart Institute, Montreal, Canada). LVEF, LVEDV, and LVESV obtained with the different modalities were compared. RESULTS: Average planar LVEF was 48 ± 11% (mean ± SD), average LVEDV was 177 ± 59 mL (range 63 to 342 mL), and average LVESV was 96 ± 46 mL (range 16 to 234 mL). GBPS LVEF and their correlation coefficient with planar LVEF were 40 ± 12% (r = 0.70) and 44 ± 12% (r = 0.83) with QBS and MHI, respectively. Correlation coefficient between cMRI and planar LVEF was 0.65 and were 0.69 and 0.52 between cMRI and GBPS using QBS and MHI, respectively. CONCLUSIONS: LVEF calculated with GBPS using IQ-SPECT correlates with planar measurements. Correlation is best using the MHI method and variation is independent of LVEDV.
引用
- Pelletier-Galarneau M, Finnerty V, Tan S, et al. Assessment of left ventricular ejection fraction with cardiofocal collimators: Comparison between IQ-SPECT, planar equilibrium radionuclide angiography, and cardiac magnetic resonance. J Nucl Cardiol. 2018. PMID: 29520572. doi:10.1007/s12350-018-1251-6
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。