血池相与组织相患者运动对铷-82 PET 心肌血流量定量的影响

Blood pool and tissue phase patient motion effects on 82rubidium PET myocardial blood flow quantification.
PubMed 文献 · J Nucl Cardiol · 2018-03-23

本研究连续纳入 225 例接受动态负荷/静息氯化铷-82(82Rb)PET 心肌灌注显像的患者,由 3 名医师阅读者评估血池相与组织相运动的发生率、幅度、时间与方向,并以运动校正图像作为误差参照,分析孤立运动对整体与区域心肌血流量(MBF)及血流储备(MFR)估计的影响。结果发现轻度至中度运动(5–15 mm)在血池相最突出,负荷与静息研究分别占 63% 与 44%;血池相孤立运动所致整体 MBF 与 MFR 平均误差为 2%–5%,而下方方向的血池相孤立运动在下壁(RCA)供血区所致 MBF 与 MFR 平均误差达 29%–44%;组织相孤立运动所致血流误差在 1% 以内。作者认为需要针对这些运动进行运动校正以减少 MBF 与 MFR 误差。

要点

  • 轻度至中度运动(5–15 mm)在血池相最突出,负荷与静息研究分别占 63% 与 44%。
  • 血池相孤立运动所致整体 MBF 与 MFR 平均误差为 2%–5%。
  • 下方方向血池相孤立运动使 RCA 供血区 MBF 与 MFR 平均误差达 29%–44%。

核心结论

  • 患者运动在血池相最普遍,且下方方向运动时 MBF 与 MFR 误差增加最明显。
  • 作者认为需要针对这些运动进行运动校正以减少 MBF 与 MFR 误差。

研究设计与样本

  • 连续纳入 225 例接受动态负荷/静息氯化铷-82(82Rb)PET 心肌灌注显像的患者。
  • 动态图像系列在最初 2 分钟内以 5–10 秒帧时长重建用于血池相,以 10–80 秒帧时长重建用于组织相。
  • 由 3 名医师阅读者从动态系列评估运动位移,分析运动频率、幅度、时间与方向。
  • 以运动校正图像得出的血流估计作为误差参照,评估孤立于时间、方向与幅度的运动对整体与区域 MBF 与 MFR 估计的影响。

主要发现

  • 轻度至中度运动(5–15 mm)在血池相最突出,负荷研究占 63%、静息研究占 44%。
  • 该运动在负荷研究以室间隔与下方方向频率 75% 出现,在静息研究以室间隔方向频率 44% 出现。
  • 血池相孤立运动的图像整体 MBF 与 MFR 平均误差为 2%–5%。
  • 下方方向血池相孤立运动导致 RCA 供血区 MBF 与 MFR 平均误差为 29%–44%。
  • 组织相孤立运动所致血流误差在 1% 以内。

临床怎么用

  • 原文未据此给出推荐。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

BACKGROUND: Patient motion can lead to misalignment of left ventricular volumes of interest and subsequently inaccurate quantification of myocardial blood flow (MBF) and flow reserve (MFR) from dynamic PET myocardial perfusion images. We aimed to identify the prevalence of patient motion in both blood and tissue phases and analyze the effects of this motion on MBF and MFR estimates. METHODS: We selected 225 consecutive patients that underwent dynamic stress/rest rubidium-82 chloride (82Rb) PET imaging. Dynamic image series were iteratively reconstructed with 5- to 10-second frame durations over the first 2 minutes for the blood phase and 10 to 80 seconds for the tissue phase. Motion shifts were assessed by 3 physician readers from the dynamic series and analyzed for frequency, magnitude, time, and direction of motion. The effects of this motion isolated in time, direction, and magnitude on global and regional MBF and MFR estimates were evaluated. Flow estimates derived from the motion corrected images were used as the error references. RESULTS: Mild to moderate motion (5-15 mm) was most prominent in the blood phase in 63% and 44% of the stress and rest studies, respectively. This motion was observed with frequencies of 75% in the septal and inferior directions for stress and 44% in the septal direction for rest. Images with blood phase isolated motion had mean global MBF and MFR errors of 2%-5%. Isolating blood phase motion in the inferior direction resulted in mean MBF and MFR errors of 29%-44% in the RCA territory. Flow errors due to tissue phase isolated motion were within 1%. CONCLUSIONS: Patient motion was most prevalent in the blood phase and MBF and MFR errors increased most substantially with motion in the inferior direction. Motion correction focused on these motions is needed to reduce MBF and MFR errors.

来源:J Nucl Cardiol,查看原文。版权归原作者所有。

引用

  1. Lee BC, Moody JB, Poitrasson-Rivière A, et al. Blood pool and tissue phase patient motion effects on 82rubidium PET myocardial blood flow quantification. J Nucl Cardiol. 2018. PMID: 29572594. doi:10.1007/s12350-018-1256-1

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。