文献速递骨显像与常规 SPECT

一种新的SPECT/CT重建算法:在非肿瘤性骨疾病临床常规中的可靠性与准确性

A new SPECT/CT reconstruction algorithm: reliability and accuracy in clinical routine for non-oncologic bone diseases.
PubMed 文献 · EJNMMI Res · 2018-02-12

本研究回顾性纳入213例因非肿瘤性骨疾病转诊至布雷斯特核医学科的患者,共207个SPECT/CT,由一名初级和一名资深核医学医师分别独立使用xSPECT Bone®(xB)/CT与Flash 3D®(F3D)/CT进行判读。结果显示xB与F3D的观察者间一致性无统计学显著差异(p=0.532),xB分类诊断一致性高(κ=0.89,95%CI 0.84-0.93),与F3D(κ=0.90,95%CI 0.86-0.94)无统计学显著差异;结论认为xB在临床常规中可靠,观察者间一致性高,诊断信心与F3D相似。

要点

  • xB与F3D的观察者间一致性无统计学显著差异(p=0.532)。
  • xB分类诊断一致性高(κ=0.89,95%CI 0.84-0.93),与F3D(κ=0.90,95%CI 0.86-0.94)无统计学显著差异。
  • 随访确认F3D诊断10例、xB诊断6例,5例无贡献。

核心结论

  • xB重建算法可靠,观察者间一致性高,诊断信心与F3D相似。

研究设计与样本

  • 回顾性纳入213例因非肿瘤性骨疾病转诊至布雷斯特核医学科的患者,共207个SPECT/CT。
  • 由一名初级和一名资深核医学医师独立判读,先使用xB/CT,三个月后使用F3D/CT。
  • 采用McNemar检验和未加权Kappa系数评估观察者间一致性和诊断信心,并通过>18个月临床和副临床随访再评估研究目标以进行验证。

主要发现

  • xB与F3D的观察者间一致性无统计学显著差异(p=0.532)。
  • xB分类诊断一致性高(κ=0.89,95%CI 0.84-0.93),与F3D(κ=0.90,95%CI 0.86-0.94)无统计学显著差异。
  • 观察到31例(14.9%)重建间诊断差异,其中21例(10.1%)被归类为主要差异。
  • 随访确认F3D诊断10例、xB诊断6例,5例无贡献。

关键数值与推荐

  • xB分类诊断一致性κ=0.89(95%CI 0.84-0.93),F3D κ=0.90(95%CI 0.86-0.94)。
  • 重建间诊断差异31例(14.9%),主要差异21例(10.1%)。

临床怎么用

  • 原文未据此给出推荐。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

BACKGROUND: xSPECT Bone® (xB) is a new reconstruction algorithm developed by Siemens® in bone hybrid imaging (SPECT/CT). A CT-based tissue segmentation is incorporated into SPECT reconstruction to provide SPECT images with bone anatomy appearance. The objectives of this study were to assess xB/CT reconstruction diagnostic reliability and accuracy in comparison with Flash 3D® (F3D)/CT in clinical routine. Two hundred thirteen consecutive patients referred to the Brest Nuclear Medicine Department for non-oncological bone diseases were evaluated retrospectively. Two hundred seven SPECT/CT were included. All SPECT/CT were independently interpreted by two nuclear medicine physicians (a junior and a senior expert) with xB/CT then with F3D/CT three months later. Inter-observer agreement (IOA) and diagnostic confidence were determined using McNemar test, and unweighted Kappa coefficient. The study objectives were then re-assessed for validation through > 18 months of clinical and paraclinical follow-up. RESULTS: No statistically significant differences between IOA xB and IOA F3D were found (p = 0.532). Agreement for xB after categorical classification of the diagnoses was high (κ xB = 0.89 [95% CI 0.84 -0.93]) but without statistically significant difference F3D (κ F3D = 0.90 [95% CI 0.86 - 0.94]). Thirty-one (14.9%) inter-reconstruction diagnostic discrepancies were observed of which 21 (10.1%) were classified as major. The follow-up confirmed the diagnosis of F3D in 10 cases, xB in 6 cases and was non-contributory in 5 cases. CONCLUSIONS: xB reconstruction algorithm was found reliable, providing high interobserver agreement and similar diagnostic confidence to F3D reconstruction in clinical routine.