文献速递心脏核医学肿瘤 PET/CT放射性药物与辐射防护

18F-FDG PET/CT显像对左心室辅助装置感染的定量分析:一项患者内回顾性研究

18F-FDG PET/CT-imaging of left ventricular assist device infection: a retrospective quantitative intrapatient analysis.
PubMed 文献 · J Nucl Cardiol · 2018-01-16

该回顾性研究纳入30例连续接受两次全身18F-FDG PET/CT的左心室辅助装置(LVAD)携带者,比较传动轴沿线五个感兴趣区的SUVmax变化。发生LVAD相关感染的患者平均ΔSUVmax为4.38±1.44,无感染者为0.03±0.43(P<0.05);ROC分析得出SUVmax阈值3.88,灵敏度和特异度均为100%。作者认为与基线PET/CT对比并采用特定SUVmax阈值可提高LVAD相关感染的诊断效能。

要点

  • LVAD相关感染患者传动轴沿线平均ΔSUVmax为4.38±1.44,无感染者为0.03±0.43,P<0.05。
  • ROC分析得出SUVmax阈值3.88,诊断灵敏度和特异度均为100%。
  • 感染患者传动轴沿线存在三种不同的摄取模式。

核心结论

  • 与基线检查对比并采用特定SUVmax阈值,可提高PET/CT对LVAD相关感染的诊断效能。

研究设计与样本

  • 回顾性研究,纳入2010年1月至2016年12月间连续接受两次全身18F-FDG PET/CT的LVAD携带者30例(平均年龄54±12岁,26例男性)。
  • PET/CT行定性(摄取模式)和半定量(SUVmax和ΔSUVmax)分析,沿LVAD传动轴测量并比较五个感兴趣区的SUVmax,计算SUVmax阈值。
  • 最终诊断依据临床检查、微生物学参数和分子影像。

主要发现

  • LVAD相关感染患者传动轴五个位置第一与第二次PET/CT之间的平均SUVmax差值(平均ΔSUVmax=4.38±1.44)显著高于无传动轴感染者(平均ΔSUVmax=0.03±0.43),P<0.05。
  • ROC分析得出SUVmax阈值3.88,灵敏度和特异度均为100%。
  • 传动轴感染患者存在三种不同的摄取模式。

关键数值与推荐

  • SUVmax阈值3.88,诊断灵敏度和特异度均为100%。
  • LVAD相关感染患者平均ΔSUVmax=4.38±1.44,无传动轴感染者平均ΔSUVmax=0.03±0.43,P<0.05。

临床怎么用

  • 原文未据此给出推荐。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

BACKGROUND: Despite the use of 18F-fluorodeoxyglucose (18F-FDG) positron emission tomography/computed tomography (PET/CT), diagnosis of a driveline infection in ventricular assist device (LVAD) recipients remains challenging. Our aim was to evaluate the potential of a baseline 18F-FDG PET/CT (prior to an infection) for the diagnosis of an LVAD-related infection. METHODS: We retrospectively selected all LVAD recipients who had undergone two consecutive whole-body 18F-FDG PET/CT examinations between January 2010 and December 2016. PET/CT was analyzed qualitatively (uptake pattern) and semi-quantitatively (SUVmax and ∆SUVmax). SUVmax was measured and compared in five distinctive volumes of interest along the LVAD driveline. An SUVmax threshold was calculated. Final diagnosis was made by clinical examination, microbiological parameters, and molecular imaging. RESULTS: Thirty patients were enrolled (mean age 54 ± 12 years; 26 male). Mean difference in SUVmax for all five positions between the first and the second PET/CT along the driveline was significantly higher in patients with an LVAD-related infection (mean ∆SUVmax = 4.38 ± 1.44) compared to those without a driveline infection (mean ∆SUVmax = 0.03 ± 0.43), P < 0.05. Applying ROC analysis, an SUVmax threshold of 3.88 resulted in a sensitivity and specificity of 100%, respectively. There were three distinctive uptake patterns in patients with a driveline infection. CONCLUSION: PET/CT diagnosis in the context of an LVAD-related infection can be improved by comparison to a baseline examination using a distinctive SUVmax threshold.

来源:J Nucl Cardiol,查看原文。版权归原作者所有。

引用

  1. Kanapinn P, Burchert W, Körperich H, et al. 18F-FDG PET/CT-imaging of left ventricular assist device infection: a retrospective quantitative intrapatient analysis. J Nucl Cardiol. 2018. PMID: 29340983. doi:10.1007/s12350-017-1161-z

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。