联合PET/MRI心肌PET显像中衰减校正的评估
该研究在20例连续心力衰竭患者中采用双示踪剂方案(13N-NH3灌注与18F-FDG代谢)行PET/MR显像,评估基于MR的衰减校正(AC)图的伪影频率及其对PET定量准确性的影响。结果发现18例(90%)患者存在平均(7±4)mm的呼吸错配,10例(50%)有磁化率伪影,6例(30%)有组织反转伪影,伪影AC图重建的PET图像平均差异为14±10%,联合伪影效应导致3例(15%)患者出现假阳性发现。结论认为标准DIXON-AC图在用于心脏PET/MRI衰减校正前须仔细检查伪影与错配效应,以避免对灌注和代谢读数的误判。
要点
- 18例(90%)患者PET发射数据与AC图存在平均(7±4)mm呼吸错配。
- 10例(50%)有磁化率伪影,6例(30%)有组织反转伪影。
- 伪影AC图重建PET图像平均差异14±10%,3例(15%)出现假阳性。
核心结论
- 标准DIXON-AC图在用于心脏PET/MRI衰减校正前须仔细检查伪影与错配效应,以避免对灌注和代谢读数的误判。
研究设计与样本
- 在20例连续心力衰竭患者中行双示踪剂心肌活力研究,采用13N-NH3(813±86 MBq)评估灌注、18F-FDG(335±38 MBq)评估代谢,所有采集使用全集成PET/MR系统并以标准DIXON-AC图进行衰减校正。
主要发现
- 18例(90%)患者观察到PET发射数据与AC图之间平均(7±4)mm的呼吸错配。
- AC图肺体积在重测分析中显示实质性变化,比值为1.0±0.2,范围0.8-1.4。
- 10例(50%)患者观察到磁化率伪影,6例(30%)患者存在组织反转伪影。
- 使用伪影AC图重建的PET图像平均差异为14±10%。
- 联合伪影效应导致3例(15%)患者出现假阳性发现。
临床怎么用
- 作者认为,标准DIXON-AC图在用于心脏PET/MRI衰减校正前须仔细检查伪影与错配效应,以避免对灌注和代谢读数的误判。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
OBJECTIVE: To evaluate the frequency of artifacts in MR-based attenuation correction (AC) maps and their impact on the quantitative accuracy of PET-based flow and metabolism measurements in a cohort of consecutive heart failure patients undergoing combined PET/MR imaging. METHODS: Myocardial viability studies were performed in 20 patients following a dual-tracer protocol involving the assessment of myocardial perfusion (13N-NH3: 813 ± 86 MBq) and metabolism (18F-FDG: 335 ± 38 MBq). All acquisitions were performed using a fully-integrated PET/MR system, with standard DIXON-attenuation correction (DIXON-AC) mapping for each PET scan. All AC maps were examined for spatial misalignment with the emission data, total lung volume, susceptibility artifacts, and tissue inversion (TI). Misalignment and susceptibility artifacts were corrected using rigid co-registration and retrospective filling of the susceptibility-induced gaps, respectively. The effects of the AC artifacts were evaluated by relative difference measures and perceived changes in clinical interpretations. RESULTS: Average respiratory misalignment of (7 ± 4) mm of the PET-emission data and the AC maps was observed in 18 (90%) patients. Substantial changes in the lung volumes of the AC maps were observed in the test-retest analysis (ratio: 1.0 ± 0.2, range: 0.8-1.4). Susceptibility artifacts were observed in 10 (50%) patients, while six (30%) patients had TI artifacts. Average differences of 14 ± 10% were observed for PET images reconstructed with the artifactual AC maps. The combined artifact effects caused false-positive findings in three (15%) patients. CONCLUSION: Standard DIXON-AC maps must be examined carefully for artifacts and misalignment effects prior to AC correction of cardiac PET/MRI studies in order to avoid misinterpretation of biased perfusion and metabolism readings from the PET data.
引用
- Lassen ML, Rasul S, Beitzke D, et al. Assessment of attenuation correction for myocardial PET imaging using combined PET/MRI. J Nucl Cardiol. 2017. PMID: 29168158. doi:10.1007/s12350-017-1118-2
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。