兼容磁共振的可移动PET系统序贯PET/MRI与PET/CT的比较
本研究前瞻性纳入17例疑似或已知恶性肿瘤患者,比较新型柔性PET(fxPET)与常规全身PET的图像质量、病灶检出率和定量值。fxPET图像质量可满足恶性肿瘤诊断,两种方法病灶检出率无显著差异,SUVmax和SUVpeak呈强正相关,fxPET测量值约为全身PET的1.1倍,fxPET与MR图像平均配准误差为5.5±3.4 mm。
要点
- fxPET图像质量可满足恶性肿瘤诊断,病灶检出率与全身PET无显著差异。
- fxPET与全身PET的SUVmax(ρ=0.88)和SUVpeak(ρ=0.81)呈强正相关。
- fxPET测量的SUVmax和SUVpeak约为全身PET的1.1倍。
核心结论
- 初步数据表明,在现有MRI系统附近运行fxPET扫描仪可获得视觉和定量上可接受的融合PET/MR图像,用于恶性病灶诊断。
研究设计与样本
- 前瞻性研究,纳入17例疑似或已知恶性肿瘤患者,静脉注射18F-FDG约1 h后行全身PET/CT,随后行fxPET和MRI。
- fxPET图像重建采用基于MRI的衰减校正。
- 视觉评估fxPET图像质量,比较两种成像方法检出病灶数量,比较病灶SUVmax和1 cm3球形体素内最大平均SUV(SUVpeak),并评估fxPET与MR图像配准误差。
主要发现
- fxPET图像质量对恶性肿瘤诊断可接受。
- fxPET与全身PET对恶性病灶的检出率无显著差异(P > 0.05),但fxPET系统未表现出优于全身PET系统的性能。
- fxPET与全身PET的SUVmax呈强正相关(ρ = 0.88),SUVpeak呈强正相关(ρ = 0.81)。
- fxPET测量的SUVmax和SUVpeak约为全身PET测量值的1.1倍。
- fxPET与MR图像的平均配准误差为5.5 ± 3.4 mm。
临床怎么用
- 原文未据此给出推荐。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
The current study tested a newly developed flexible PET (fxPET) scanner prototype. This fxPET system involves dual arc-shaped detectors based on silicon photomultipliers that are designed to fit existing MRI devices, allowing us to obtain fused PET and MR images by sequential PET and MR scanning. This prospective study sought to evaluate the image quality, lesion detection rate, and quantitative values of fxPET in comparison with conventional whole-body (WB) PET and to assess the accuracy of registration. Methods: Seventeen patients with suspected or known malignant tumors were analyzed. Approximately 1 h after intravenous injection of 18F-FDG, WB PET/CT was performed, followed by fxPET and MRI. For reconstruction of fxPET images, MRI-based attenuation correction was applied. The quality of fxPET images was visually assessed, and the number of detected lesions was compared between the 2 imaging methods. SUVmax and maximum average SUV within a 1 cm3 spheric volume (SUVpeak) of lesions were also compared. In addition, the magnitude of misregistration between fxPET and MR images was evaluated. Results: The image quality of fxPET was acceptable for diagnosis of malignant tumors. There was no significant difference in detectability of malignant lesions between fxPET and WB PET (P > 0.05). However, the fxPET system did not exhibit superior performance to the WB PET system. There were strong positive correlations between the 2 imaging modalities in SUVmax (ρ = 0.88) and SUVpeak (ρ = 0.81). SUVmax and SUVpeak measured with fxPET were approximately 1.1-fold greater than measured with WB PET. The average misregistration between fxPET and MR images was 5.5 ± 3.4 mm. Conclusion: Our preliminary data indicate that running an fxPET scanner near an existing MRI system provides visually and quantitatively acceptable fused PET/MR images for diagnosis of malignant lesions.