依据NEMA NU2-2012标准与日本肿瘤FDG PET/CT数据采集指南第2.0版对Celesteion大孔径PET/CT扫描仪的初步评估
本研究依据NEMA NU-2 2012标准和日本肿瘤FDG PET/CT数据采集指南第2.0版,评估Celesteion大孔径飞行时间PET/CT扫描仪的空间分辨率、灵敏度、计数率特性、散射分数、能量分辨率、TOF时间分辨率和图像质量。结果显示该系统灵敏度和噪声等效计数率较低,但空间分辨率与时间分辨率良好;按日本指南,每床位2分钟采集时质量噪声比(QNR)仍满足推荐值,作者认为每床位2分钟采集可提供可接受的图像质量。
要点
- 空间分辨率为4.5–5 mm半高全宽(FWHM),灵敏度3.8 cps/kBq,散射分数37.3%。
- 峰值噪声等效计数率为70 kcps,对应体模放射性浓度29.6 kBq mL-1。
- 系统能量分辨率12.4%,TOF时间分辨率411 ps(FWHM)。
核心结论
- Celesteion大孔径PET/CT系统与其他PET/CT扫描仪相比灵敏度与NEC较低,但空间分辨率和时间分辨率良好。
- 即使采集2分钟,质量噪声比(QNR)也满足日本指南推荐值。
- 作者认为Celesteion以每床位2分钟采集(日本最常见扫描方案)可提供可接受的图像质量。
研究设计与样本
- 本研究为对Celesteion大孔径飞行时间PET/CT扫描仪的初步评估,依据NEMA NU-2 2012标准及日本肿瘤FDG PET/CT数据采集指南第2.0版进行。
- 使用18F溶液和NEMA NU-2 2012标准所述的IEC体模进行体模实验。
- 按日本指南计算检测10 mm热球(靶本比4:1)所需最短扫描时间、体模噪声等效计数(NECphantom)、%背景变异性(N10mm)、%对比度(QH,10mm)和恢复系数(RC)。
主要发现
- 测得空间分辨率为4.5–5 mm半高全宽(FWHM)。
- 灵敏度和散射分数分别为3.8 cps/kBq和37.3%。
- 峰值噪声等效计数率为70 kcps,对应体模中放射性浓度29.6 kBq mL-1。
- 系统能量分辨率为12.4%,TOF时间分辨率为411 ps(FWHM)。
- 直径10 mm球体的恢复系数(RC)为0.49,超过最低推荐值。
关键数值与推荐
- 日本指南推荐每床位最短扫描时间:视觉评分2分钟、NECphantom 7分钟、N10mm 6分钟、QH,10mm与N10mm比值(QNR)2分钟。
- 直径10 mm球体的RC为0.49,超过最低推荐值。
- QNR即使采集2分钟也满足日本指南推荐值。
临床怎么用
- 作者认为Celesteion以每床位2分钟采集(日本最常见扫描方案)可提供可接受的图像质量。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
BACKGROUND: The goal of this study was to evaluate the performance of the Celesteion positron emission tomography/computed tomography (PET/CT) scanner, which is characterized by a large-bore and time-of-flight (TOF) function, in accordance with the NEMA NU-2 2012 standard and version 2.0 of the Japanese guideline for oncology fluorodeoxyglucose PET/CT data acquisition protocol. Spatial resolution, sensitivity, count rate characteristic, scatter fraction, energy resolution, TOF timing resolution, and image quality were evaluated according to the NEMA NU-2 2012 standard. Phantom experiments were performed using 18F-solution and an IEC body phantom of the type described in the NEMA NU-2 2012 standard. The minimum scanning time required for the detection of a 10-mm hot sphere with a 4:1 target-to-background ratio, the phantom noise equivalent count (NECphantom), % background variability (N 10mm), % contrast (Q H,10mm), and recovery coefficient (RC) were calculated according to the Japanese guideline. RESULTS: The measured spatial resolution ranged from 4.5- to 5-mm full width at half maximum (FWHM). The sensitivity and scatter fraction were 3.8 cps/kBq and 37.3%, respectively. The peak noise-equivalent count rate was 70 kcps in the presence of 29.6 kBq mL-1 in the phantom. The system energy resolution was 12.4% and the TOF timing resolution was 411 ps at FWHM. Minimum scanning times of 2, 7, 6, and 2 min per bed position, respectively, are recommended for visual score, noise-equivalent count (NEC)phantom, N 10mm, and the Q H,10mm to N 10mm ratio (QNR) by the Japanese guideline. The RC of a 10-mm-diameter sphere was 0.49, which exceeded the minimum recommended value. CONCLUSIONS: The Celesteion large-bore PET/CT system had low sensitivity and NEC, but good spatial and time resolution when compared to other PET/CT scanners. The QNR met the recommended values of the Japanese guideline even at 2 min. The Celesteion is therefore thought to provide acceptable image quality with 2 min/bed position acquisition, which is the most common scan protocol in Japan.
引用
- Kaneta T, Ogawa M, Motomura N, et al. Initial evaluation of the Celesteion large-bore PET/CT scanner in accordance with the NEMA NU2-2012 standard and the Japanese guideline for oncology FDG PET/CT data acquisition protocol version 2.0. EJNMMI Res. 2017. PMID: 29022216. doi:10.1186/s13550-017-0331-y
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。