文献速递心脏核医学

心脏SPECT/CT定量测定正常123I-间碘苄胍心纵隔比值(HMR)及其与平面HMR的相关性

Quantification and Determination of Normal 123I-Meta Iodobenzylguanidine Heart-to-Mediastinum Ratio (HMR) from Cardiac SPECT/CT and Correlation with Planar HMR.
PubMed 文献 · J Nucl Med · 2017-09-15

研究纳入21名健康志愿者,注射123I-mIBG后用传统SPECT相机采集平面图像、专用心脏多针孔SPECT/CT采集断层图像,建立半自动定量HMR方法并确定正常下限。结果:SPECT HMR采用上纵隔本底时与平面HMR相关性最高(R=0.91),SPECT HMR正常下限为5.5,平面HMR正常下限为1.6,两者呈线性相关,可由SPECT HMR经简单线性回归方程估算平面HMR。

要点

  • 健康志愿者中SPECT HMR(上纵隔本底)与平面HMR相关性最高,R=0.91。
  • SPECT HMR正常下限为5.5,平面HMR正常下限为1.6。
  • 平面HMR可由SPECT HMR经简单线性回归方程估算。

核心结论

  • 研究再次确认了既往平面HMR阈值,并确定了SPECT HMR的正常下限;平面HMR可由SPECT HMR经简单线性回归方程估算,使新型心脏专用SPECT相机可用于123I-mIBG显像。

研究设计与样本

  • 研究纳入21名健康志愿者,注射123I-mIBG后使用2种不同相机进行显像。
  • 平面图像使用配备平行孔准直器的传统SPECT相机采集,混合SPECT/CT图像使用配备19个针孔准直器并与64层CT接口的专用心脏SPECT系统采集。
  • 平面HMR按标准指南(手工传统法)及椭圆感兴趣区(Elip-ROI)和区域生长(RG-ROI)技术计算;SPECT HMR采用新方法,结合多种心脏与纵隔分割方案,其中心脏上下界由CT和左心室ROI确定,平均计数用Elip-ROI和RG-ROI技术计算。
  • 纵隔ROI平均计数从3个不同区域的固定体积计算:上纵隔(UM)、下纵隔(LM)和对侧肺(CL)。
  • HMR由2名观察者处理,重复性用组内相关系数和Bland-Altman分析评估。

主要发现

  • 平面HMR用RG-ROI法计算时,观察者内和观察者间一致性最高,优于Elip-ROI法和手工传统法。
  • 基于UM、LM和CL本底区域计算的SPECT HMR均显示极好的观察者内和观察者间一致性。
  • SPECT HMR采用UM本底时与平面HMR相关性最高(R=0.91),高于LM(R=0.74)和CL(R=0.73)。
  • SPECT HMR(UM本底)的正常下限为5.5,平面HMR的正常下限为1.6。
  • SPECT来源的HMR与平面来源的HMR正常值呈线性相关。

关键数值与推荐

  • SPECT HMR采用上纵隔(UM)本底时与平面HMR相关性最高,R=0.91。
  • SPECT HMR(UM本底)正常下限为5.5,平面HMR正常下限为1.6。
  • 平面HMR可由SPECT HMR经简单线性回归方程估算。

临床怎么用

  • 原文未据此给出推荐。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

Assessment of cardiac 123I-meta iodobenzylguanidine (123I-mIBG) uptake relies on the heart-to-mediastinum ratio (HMR) derived from planar images. We have developed novel semiautomated quantitative methodologies for assessing HMR from SPECT images using a dedicated cardiac multipinhole SPECT/CT system and determined the lower limit of normal (LLN) SPECT-derived HMR and the correlation to planar-derived HMR. Methods: Twenty-one healthy volunteers were injected with 123I-mIBG and imaged using 2 different cameras. Planar images were acquired using a conventional SPECT camera equipped with parallel hole collimators, and hybrid SPECT/CT images were acquired using a dedicated cardiac SPECT system with 19 pinhole collimators interfaced with 64-slice CT. Planar HMR was calculated as per standard guidelines (manual traditional method) and elliptic region-of-interest (Elip-ROI) and region-growing (RG-ROI) techniques. SPECT HMR was quantified using a new method that incorporates various cardiac and mediastinal segmentation schemes in which upper and lower limits of the heart were determined from CT and the left ventricular ROI, and mean counts were calculated using Elip-ROI and RG-ROI techniques. Mean counts in mediastinal ROI were computed from a fixed volume in 3 different regions: upper mediastinum (UM), lower mediastinum (LM), and contralateral lung (CL). HMRs were processed by 2 observers, and reproducibility was assessed by intraclass correlation coefficient and Bland-Altman analysis. Results: Planar HMR calculated using the RG-ROI method showed highest intra- and interobserver levels of agreement compared with Elip-ROI and manual traditional methods. SPECT HMR calculated on the basis of UM, LM, and CL background regions showed excellent intra- and interobserver agreement. SPECT HMR with UM resulted in highest correlation (R = 0.91) with planar HMR compared with that with LM (R = 0.74) and CL (R = 0.73). The LLN of SPECT HMR with UM and that of planar HMR was calculated as 5.5 and 1.6, respectively. The normal values of SPECT-derived HMR and planar-derived HMR were correlated linearly. Conclusion: We reconfirmed the previous planar HMR threshold and determined SPECT LLN HMR for SPECT. Planar HMR can be estimated from SPECT HMR via a simple linear regression equation, allowing use of the new cardiac-dedicated SPECT camera for 123I-mIBG imaging.

来源:J Nucl Med,查看原文。版权归原作者所有。

引用

  1. Alvi R, Miller EJ, Zonouz TH, et al. Quantification and Determination of Normal 123I-Meta Iodobenzylguanidine Heart-to-Mediastinum Ratio (HMR) from Cardiac SPECT/CT and Correlation with Planar HMR. J Nucl Med. 2017. PMID: 28916622. doi:10.2967/jnumed.117.197152

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。