文献速递心脏核医学

利用散射与光电峰窗数据分割进行衰减校正可提高心肌灌注SPECT对狭窄冠状动脉血管数量的诊断准确性

Improved diagnosis of the number of stenosed coronary artery vessels by segmentation with scatter and photo-peak window data for attenuation correction in myocardial perfusion SPECT.
PubMed 文献 · J Nucl Cardiol · 2017-09-13

本研究回顾性分析183例在3个月内接受99mTc-替曲膦负荷/静息SPECT及冠状动脉造影的患者,比较SSPAC衰减校正与非校正心肌灌注显像在诊断狭窄冠状动脉血管数量上的准确性。SSPAC在179/183例(98%)中成功确定衰减图,在单支及多支血管病变中诊断冠状动脉疾病的多项指标优于非校正法。

要点

  • SSPAC衰减校正成功确定衰减图的比例为179/183例(98%)。
  • 血管水平诊断中,SSPAC的敏感度、特异度、阳性预测值、阴性预测值分别为77%、89%、74%、90%,非校正法分别为51%、87%、62%、82%。
  • 在35例多支血管病变患者中,SSPAC的敏感度、特异度、阳性预测值、阴性预测值分别为78%、81%、93%、55%,非校正法分别为49%、81%、89%、34%。

核心结论

  • SSPAC校正的SPECT心肌灌注图像在检测狭窄冠状动脉血管数量方面准确性提高,即使在多支血管病变患者中也是如此。

研究设计与样本

  • 回顾性分析183例连续患者,均在3个月内接受99mTc-替曲膦负荷/静息SPECT检查及冠状动脉造影。

主要发现

  • SSPAC在179/183例(98%)中成功确定衰减图。
  • 血管水平诊断冠状动脉疾病:SSPAC与非校正法的敏感度分别为77%与51%,特异度分别为89%与87%,阳性预测值分别为74%与62%,阴性预测值分别为90%与82%(*P < .05)。
  • 35例多支血管病变患者中:SSPAC与非校正法的敏感度分别为78%与49%,特异度分别为81%与81%,阳性预测值分别为93%与89%,阴性预测值分别为55%与34%(*P < .05;AUC:0.82 vs 0.62,P < .05)。

关键数值与推荐

  • 使用SSPAC衰减校正与非校正方法,以 summed differential score 进行定量判读,检测冠状动脉疾病。
  • 血管水平诊断冠状动脉疾病:SSPAC敏感度77%、特异度89%、阳性预测值74%、阴性预测值90%;非校正法敏感度51%、特异度87%、阳性预测值62%、阴性预测值82%(*P < .05)。
  • 多支血管病变患者(35例):SSPAC敏感度78%、特异度81%、阳性预测值93%、阴性预测值55%;非校正法敏感度49%、特异度81%、阳性预测值89%、阴性预测值34%(*P < .05;AUC:0.82 vs 0.62,P < .05)。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

BACKGROUND: Attenuation correction using segmentation of scatter and photo-peak window data (SSPAC) enables an evaluation of the attenuation map in a patient-specific manner without additional radiation exposure. We compared the accuracy of SSPAC and non-corrected myocardial perfusion scintigraphy methods for diagnosing the number of stenosed coronary artery vessels. METHODS AND RESULTS: We retrospectively reviewed the data from 183 consecutive patients who underwent 99mTc-tetrofosmin stress/rest SPECT examination and a coronary angiography within 3 months. The MPS images were reconstructed with and without SSPAC attenuation correction. We examined the accuracy of the quantitative interpretation using summed differential score in the detection of coronary artery disease (CAD). The attenuation maps were successfully determined in 179 of 183 patients (98%). In terms of the vessel-based diagnostic ability, sensitivity, specificity, positive predictive and negative predictive values of the SSPAC and non-correction methods for diagnosing CAD in individual coronary territories were 77%*, 89%, 74%*, and 90%* vs 51%, 87%, 62%, and 82%, respectively (*P < .05). In 35 patients with multi-vessel CAD, those values were 78%*, 81%, 93%, and 55%* vs 49%, 81%, 89%, and 34%, respectively (*P < .05; AUC: 0.82 vs 0.62, P < .05). CONCLUSION: SSPAC-corrected SPECT myocardial perfusion images exhibit improved accuracy in the detection of the number of stenosed coronary artery vessels, even in patients with multi-vessel CAD.

来源:J Nucl Cardiol,查看原文。版权归原作者所有。

引用

  1. Yamauchi Y, Kanzaki Y, Hayashi M, et al. Improved diagnosis of the number of stenosed coronary artery vessels by segmentation with scatter and photo-peak window data for attenuation correction in myocardial perfusion SPECT. J Nucl Cardiol. 2017. PMID: 28905206. doi:10.1007/s12350-017-1058-x

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。