文献速递肿瘤 PET/CT放射性药物与辐射防护

7 Tesla下用于高分辨率小动物PET/MRI的PET插入件的性能

Performance of a PET Insert for High-Resolution Small-Animal PET/MRI at 7 Tesla.
PubMed 文献 · J Nucl Med · 2017-09-14

该研究表征了一种紧凑型磁共振兼容PET插入件,用于同时临床前PET/MRI。插入件由16个探测器块环组成,使用双层偏移硅酸镥钇晶体阵列和硅光电倍增器阵列读出,空间分辨率在径向偏移0-15 mm范围内为1.17-1.86 mm半高全宽,峰值灵敏度2.2%,可分辨0.7 mm小特征。小鼠和大鼠脑18F-FDG PET/MR图像无模态间干扰,能清晰分辨脑内亚结构。

要点

  • 空间分辨率径向偏移0-15 mm范围内为1.17-1.86 mm半高全宽。
  • 300-800 keV能量窗下峰值灵敏度2.2%,小鼠体模噪声等效计数率在3.7 MBq时为11.1 kcps,14.5 MBq时峰值20.8 kcps。
  • 可分辨小至0.7 mm的特征,18F-FDG PET/MR脑图像无模态间干扰。

核心结论

  • 由于出色的空间分辨率和无模态间干扰,该PET插入件将成为临床前PET/MR的有用工具。

研究设计与样本

  • 研究表征了一种紧凑型磁共振兼容PET插入件,专为适配BioSpec 70/20和94/20系列小动物MRI系统所用BGA-12S梯度线圈的114 mm内径而设计,也可安装于任何适当MRI扫描仪或作为独立PET系统使用。
  • 插入件由16个探测器块环组成,每个探测器块由双层偏移的铈掺杂硅酸镥钇晶体阵列构成,具备作用深度能力,由硅光电倍增器阵列读出。
  • 闪烁晶体阵列底层为22×10晶体、顶层为21×9晶体,层厚分别为6 mm和4 mm,排列间距1.27 mm,可用视野长28 mm、宽约55 mm。

主要发现

  • 空间分辨率在径向偏移0-15 mm范围内为1.17至1.86 mm半高全宽。
  • 300-800 keV能量窗下峰值灵敏度为2.2%,小鼠体模噪声等效计数率在3.7 MBq时为11.1 kcps,在14.5 MBq时峰值达20.8 kcps。
  • 体模成像显示可分辨小至0.7 mm的特征。
  • 小鼠和大鼠脑18F-FDG PET/MR图像无模态间干扰迹象,能极好地分辨脑内亚结构。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

We characterize a compact MR-compatible PET insert for simultaneous preclinical PET/MRI. Although specifically designed with the strict size constraint to fit inside the 114-mm inner diameter of the BGA-12S gradient coil used in the BioSpec 70/20 and 94/20 series of small-animal MRI systems, the insert can easily be installed in any appropriate MRI scanner or used as a stand-alone PET system. Methods: The insert consists of a ring of 16 detector-blocks each made from depth-of-interaction-capable dual-layer-offset arrays of cerium-doped lutetium-yttrium oxyorthosilicate crystals read out by silicon photomultiplier arrays. Scintillator crystal arrays are made from 22 × 10 and 21 × 9 crystals in the bottom and top layers, respectively, with respective layer thicknesses of 6 and 4 mm, arranged with a 1.27-mm pitch, resulting in a useable field of view 28 mm long and about 55 mm wide. Results: Spatial resolution ranged from 1.17 to 1.86 mm full width at half maximum in the radial direction from a radial offset of 0-15 mm. With a 300- to 800-keV energy window, peak sensitivity was 2.2% and noise-equivalent count rate from a mouse-sized phantom at 3.7 MBq was 11.1 kcps and peaked at 20.8 kcps at 14.5 MBq. Phantom imaging showed that features as small as 0.7 mm could be resolved. 18F-FDG PET/MR images of mouse and rat brains showed no signs of intermodality interference and could excellently resolve substructures within the brain. Conclusion: Because of excellent spatial resolvability and lack of intermodality interference, this PET insert will serve as a useful tool for preclinical PET/MR.

来源:J Nucl Med,查看原文。版权归原作者所有。