单剂量静态-动态联合成像方案:动态SPECT与常规静态SPECT定量比较的初步可行性研究
本研究纳入15例患者,采用单剂量一日法静态-动态联合SPECT心肌灌注成像方案,其中11例有选择性冠状动脉造影(SCA)结果作参考。动态SPECT测得的心肌血流储备(CFR)诊断冠状动脉疾病(CAD)的准确度91%、敏感度80%、特异度100%,SCA异常患者平均CFR显著低于正常患者(1.94±0.67 vs 3.86±1.06,P<0.001)。
要点
- 动态SPECT测得的CFR诊断CAD准确度91%、敏感度80%、特异度100%。
- SCA异常者平均CFR显著低于正常者(1.94±0.67 vs 3.86±1.06,P<0.001)。
- 静态与动态SPECT视觉评估主观,可能不反映SCA所示冠脉病变的生理学。
核心结论
- 静态与动态SPECT的视觉图像判读具有主观性,可能不反映基于SCA的冠脉病变直接生理学测量。
- 动态SPECT测得的CFR完全客观,敏感度和特异度更好,且仅能由动态SPECT方法获得的数据计算。
研究设计与样本
- 初步可行性研究,纳入15例患者(11男4女,平均年龄71±9岁),采用单剂量一日法静态-动态联合SPECT(Infinia Hawkeye 4,GE Healthcare)成像方案,静息与负荷各给一剂99mTc-替曲膦。
- 15例中11例有SCA,其中8例在SPECT成像后6个月内、其余在24个月内完成,期间无新发症状或干预。
主要发现
- 常规静态SPECT正常10例、异常5例;动态SPECT正常12例、异常3例;动态SPECT测得的CFR正常9例、异常6例。
- 11例有SCA者中,常规SPECT正常5例,其中3例SCA证实CAD,总体准确度64%、敏感度40%、特异度83%。
- 动态SPECT图像分析得到相似的准确度、敏感度和特异度。
- CFR正常6例,均SCA证实CAD,总体准确度91%、敏感度80%、特异度100%。
- SCA异常者平均CFR显著低于正常者(1.94±0.67 vs 3.86±1.06,P<0.001)。
关键数值与推荐
- 常规静态SPECT诊断CAD:准确度64%、敏感度40%、特异度83%。
- 动态SPECT图像分析诊断CAD:准确度、敏感度、特异度与常规静态SPECT相似。
- 动态SPECT测得的CFR诊断CAD:准确度91%、敏感度80%、特异度100%。
- SCA异常者平均CFR 1.94±0.67 vs SCA正常者3.86±1.06,P<0.001。
临床怎么用
- 原文未据此给出推荐。
局限性
- 原文报告为小样本患者的初步可行性研究。
- 原文指出静态与动态SPECT的视觉图像判读具有主观性,可能不反映基于SCA的冠脉病变直接生理学测量。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
BACKGROUND: SPECT myocardial perfusion imaging (MPI) is a clinical mainstay that is typically performed with static imaging protocols and visually or semi-quantitatively assessed for perfusion defects based upon the relative intensity of myocardial regions. Dynamic cardiac SPECT presents a new imaging technique based on time-varying information of radiotracer distribution, which permits the evaluation of regional myocardial blood flow (MBF) and coronary flow reserve (CFR). In this work, a preliminary feasibility study was conducted in a small patient sample designed to implement a unique combined static-dynamic single-dose one-day visit imaging protocol to compare quantitative dynamic SPECT with static conventional SPECT for improving the diagnosis of coronary artery disease (CAD). METHODS: Fifteen patients (11 males, four females, mean age 71 ± 9 years) were enrolled for a combined dynamic and static SPECT (Infinia Hawkeye 4, GE Healthcare) imaging protocol with a single dose of 99mTc-tetrofosmin administered at rest and a single dose administered at stress in a one-day visit. Out of 15 patients, eleven had selective coronary angiography (SCA), 8 within 6 months and the rest within 24 months of SPECT imaging, without intervening symptoms or interventions. The extent and severity of perfusion defects in each myocardial region was graded visually. Dynamically acquired data were also used to estimate the MBF and CFR. Both visually graded images and estimated CFR were tested against SCA as a reference to evaluate the validity of the methods. RESULTS: Overall, conventional static SPECT was normal in ten patients and abnormal in five patients, dynamic SPECT was normal in 12 patients and abnormal in three patients, and CFR from dynamic SPECT was normal in nine patients and abnormal in six patients. Among those 11 patients with SCA, conventional SPECT was normal in 5, 3 with documented CAD on SCA with an overall accuracy of 64%, sensitivity of 40% and specificity of 83%. Dynamic SPECT image analysis also produced a similar accuracy, sensitivity, and specificity. CFR was normal in 6, each with CAD on SCA with an overall accuracy of 91%, sensitivity of 80%, and specificity of 100%. The mean CFR was significantly lower for SCA detected abnormal than for normal patients (3.86±1.06 vs 1.94±0. 0.67, P < 0.001). CONCLUSIONS: The visually assessed image findings in static and dynamic SPECT are subjective, and may not reflect direct physiologic measures of coronary lesion based on SCA. The CFR measured with dynamic SPECT is fully objective, with better sensitivity and specificity, available only with the data generated from the dynamic SPECT method.
引用
- Sciammarella M, Shrestha UM, Seo Y, et al. A combined static-dynamic single-dose imaging protocol to compare quantitative dynamic SPECT with static conventional SPECT. J Nucl Cardiol. 2017. PMID: 28776314. doi:10.1007/s12350-017-1016-7