优化SPECT心肌血流量测量:衰减、运动和血液结合校正与PET的比较
本研究在同一批患者中比较了99mTc-替曲膦SPECT与PET测量的心肌血流量(MBF)和心肌血流储备(MFR)。31例患者在SPECT检查1个月内接受82Rb或13N-氨PET静息-负荷MBF检查。结果显示,红细胞结合校正和运动校正提高了SPECT MBF的准确度和精密度,而衰减校正对整体MBF和异常MFR检测无显著改善。固态SPECT测得的整体MBF和MFR与PET的差异分别为2%±32%和2%±28%。
要点
- 红细胞结合校正和运动校正提高了SPECT整体MBF的准确度和精密度
- 衰减校正对整体MBF和异常MFR检测无显著改善
- 固态SPECT测得的整体MBF和MFR与PET的差异分别为2%±32%和2%±28%
核心结论
- 固态SPECT可提供整体MBF和MFR测量,与PET的差异分别为2%±32%(MBF)和2%±28%(MFR)
研究设计与样本
- 31例患者在SPECT检查1个月内接受82Rb或13N-氨PET静息-负荷MBF检查,并在同一批患者中比较99mTc-替曲膦SPECT与PET的MBF测量
主要发现
- 红细胞结合校正和运动校正均提高了整体SPECT MBF相对于PET MBF的准确度和精密度,平均差异为0.06±0.37 mL/min/g
- 衰减校正对整体MBF和异常MFR检测无显著改善
- 在运动校正和红细胞结合校正下,整体SPECT MFR检测PET MFR小于2.0的异常时,受试者工作特征曲线下面积为0.90(加衰减校正)至0.95(不加衰减校正)
- 区域分析结果相似,受试者工作特征曲线下面积为0.84(加衰减校正)至0.87(不加衰减校正)
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
Myocardial blood flow (MBF) and myocardial flow reserve (MFR) measured with PET have clinical value. SPECT cameras with solid-state detectors can obtain dynamic images for measurement of MBF and MFR. In this study, SPECT measurements of MBF made using 99mTc-tetrofosmin were compared with PET in the same patients. Methods: Thirty-one patients underwent PET MBF rest-stress studies performed with 82Rb or 13N-ammonia within 1 mo of their SPECT study. Dynamic rest-stress measurements were made using a SPECT camera. Kinetic parameters were calculated using a 1-tissue-compartment model and converted to MBF and MFR. Processing with and without corrections for attenuation (+AC and -AC), patient body motion (+MC and -MC), and binding of the tracer to red blood cells (+BB and -BB) was evaluated. Results: Both +BB and +MC improved the accuracy and precision of global SPECT MBF compared with PET MBF, resulting in an average difference of 0.06 ± 0.37 mL/min/g. Global MBF and detection of abnormal MFR were not significantly improved with +AC. Global SPECT MFR with +MC and +BB had an area under the receiver-operating curve of 0.90 (+AC) to 0.95 (-AC) for detecting abnormal PET MFR less than 2.0. Regional analysis produced similar results with an area under the receiver-operating curve of 0.84 (+AC) to 0.87 (-AC). Conclusion: Solid-state SPECT provides global MBF and MFR measurements that differ from PET by 2% ± 32% (MBF) and 2% ± 28% (MFR).
引用
- Wells RG, Marvin B, Poirier M, et al. Optimization of SPECT Measurement of Myocardial Blood Flow with Corrections for Attenuation, Motion, and Blood Binding Compared with PET. J Nucl Med. 2017. PMID: 28611245. doi:10.2967/jnumed.117.191049
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。