基于CT的衰减校正对专用心脏固态CZT SPECT/CT心肌灌注显像自动灌注评分的影响
研究在多重针孔CZT SPECT/CT上获取心脏CZT专用增强数据库及其对基于CT的衰减校正(AC)灌注评分的影响。三组独立患者分别比较CZT与常规碘化钠(NaI)设备的评分、建立参考CZT数据库、以及比较AC与非AC(NAC)评分。结果显示CZT专用数据库评分与常规SPECT一致,AC显著降低男性SSS和SRS,女性无显著差异。
要点
- CZT专用数据库评分与常规NaI默认数据库评分呈显著1:1线性相关但不一致。
- AC显著降低男性平均SSS(4.29±6.30 vs 5.37±7.26,P<0.001)和SRS(2.37±4.72 vs 3.13±5.85,P<0.001)。
- 女性AC与NAC评分无显著差异(SSS 2.28±3.42 vs 2.28±3.08,p NS;SRS 0.46±1.51 vs 0.61±1.86,p NS)。
核心结论
- 专为固态CZT心脏SPECT设计的数据库能提供与常规SPECT已验证评分一致的灌注评分准确定量。
- AC与NAC的CZT评分存在显著差异,尤其在男性中。
研究设计与样本
- 研究在多重针孔CZT SPECT/CT上获取数据,纳入三个独立患者组:组1(n=80)同时接受CZT和常规碘化钠(NaI)设备MPI;组2(n=80)冠心病可能性低且MPI正常,提供参考CZT数据库;组3(n=152)比较CZT上的AC与NAC评分。
主要发现
- 组1中,CZT评分参照自定义数据库与NaI评分参照默认数据库呈显著1:1线性相关,但CZT评分参照默认数据库时两者不一致。
- 男性中AC较NAC显著降低平均SSS(4.29±6.30 vs 5.37±7.26,P<0.001)和SRS(2.37±4.72 vs 3.13±5.85,P<0.001)。
- 女性中AC与NAC的SSS(2.28±3.42 vs 2.28±3.08,p NS)和SRS(0.46±1.51 vs 0.61±1.86,p NS)无显著差异。
关键数值与推荐
- 组1 CZT与NaI评分呈显著1:1线性相关(CZT参照自定义数据库 vs NaI参照默认数据库)。
- 男性AC vs NAC:SSS 4.29±6.30 vs 5.37±7.26(P<0.001),SRS 2.37±4.72 vs 3.13±5.85(P<0.001)。
- 女性AC vs NAC:SSS 2.28±3.42 vs 2.28±3.08(p NS),SRS 0.46±1.51 vs 0.61±1.86(p NS)。
临床怎么用
- 原文未据此给出推荐。
- 专为固态CZT心脏SPECT设计的数据库可提供与常规SPECT一致的灌注评分定量。
- AC与NAC的CZT评分存在显著差异,尤其在男性中。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
BACKGROUND: Data regarding cardiac cadmium-zinc-telluride (CZT)-specific augmented databases and their impact on CT-based attenuation correction (AC) perfusion scores in myocardial perfusion imaging (MPI) were obtained on a multiple-pinhole CZT SPECT/CT. METHODS AND RESULTS: Summed stress (SSS) and rest scores (SRS) were measured using automated software in three independent patient groups: group 1 (n = 80) underwent MPI on both CZT and conventional sodium iodide (NaI) devices, group 2 (n = 80) with low coronary artery disease likelihood and normal MPI provided reference CZT databases; and group 3 (n = 152) served to compare AC and non-AC (NAC) scores on CZT. Group 1 CZT and NaI scores gave a significant 1:1 linear correlation for CZT scores referenced to the custom database vs NaI scores referenced to the default database, but these were not concordant when CZT scores were referenced to the default database. AC significantly decreased average SSS and SRS in men vs NAC, 4.29 ± 6.30 vs 5.37 ± 7.26 (P < 0.001) and 2.37 ± 4.72 vs 3.13 ± 5.85 (P < 0.001), but not in women, 2.28 ± 3.42 vs 2.28 ± 3.08 (p NS) and 0.46 ± 1.51 vs 0.61 ± 1.86, (p NS), respectively. CONCLUSIONS: Specifically designed databases for solid-state CZT cardiac SPECT provide accurate quantitation of perfusion scores concordant with those previously validated for conventional SPECT. AC and NAC CZT scores differed significantly, especially in men.