脑PET/MRI中窦/边缘校正的零回波时间衰减校正评估
研究纳入12例患者,在PET/CT后行全身18F-FDG PET/MRI,比较窦/边缘校正的ZTAC(ZTACSEC)与未校正ZTAC(ZTACUC)及CT图谱衰减校正(ATAC)相对CT衰减校正(CTAC)的PET定量准确性。ZTACSEC总体定量偏差0.2%±2.4%(P=0.23),与CTAC相似;小脑改善明显(0.6%±2.7%)。
要点
- ZTACSEC总体PET定量偏差0.2%±2.4%,与CTAC无显著差异(P=0.23)。
- ZTACUC总体偏差5.6%±3.5%(P<0.01),ATAC为-0.9%±5.0%(P=0.03)。
- 小脑ZTACSEC偏差0.6%±2.7%,优于ZTACUC(8.1%±3.5%)和ATAC(-4.1%±4.3%)。
核心结论
- ZTACSEC可在脑PET/MRI中提供与CTAC相当的准确PET定量,尤其在小脑。
研究设计与样本
- 12例患者先行PET/CT,后行全身18F-FDG PET/MRI,仅使用包含头部的床位数据。
- 对PET数据分别应用ZTACUC、ZTACSEC、ATAC和参考CTAC进行衰减校正。
- 在20个1-mL脑组织感兴趣区测量PET平均摄取值,用配对t检验比较,误差直方图展示体素水平差异分布。
主要发现
- ZTACSEC总体PET定量偏差0.2%±2.4%(P=0.23),与CTAC相似。
- ZTACUC总体偏差5.6%±3.5%(P<0.01),ATAC总体偏差-0.9%±5.0%(P=0.03)。
- 小脑ZTACSEC偏差0.6%±2.7%(P<0.01),ZTACUC偏差8.1%±3.5%(P<0.01),ATAC偏差-4.1%±4.3%(P<0.01)。
- 体素水平误差直方图显示ZTACSEC较ZTACUC和ATAC大幅降低误差幅度和变异。
关键数值与推荐
- ZTACUC伪CT转换:ZTE<0.2为空气(-1,000 HU),ZTE>0.75为软组织(42 HU),0.2≤ZTE≤0.75为骨(-2,000×[ZTE-1]+42 HU)。
- ZTACSEC在二值掩膜内将高估值重新缩放至42 HU以下。
- ATAC将变形至MR同相位的图谱分割为空气、内空气、软组织和连续骨。
临床怎么用
- 原文未据此给出临床推荐。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
In brain PET/MRI, the major challenge of zero-echo-time (ZTE)-based attenuation correction (ZTAC) is the misclassification of air/tissue/bone mixtures or their boundaries. Our study aimed to evaluate a sinus/edge-corrected (SEC) ZTAC (ZTACSEC), relative to an uncorrected (UC) ZTAC (ZTACUC) and a CT atlas-based attenuation correction (ATAC). Methods: Whole-body 18F-FDG PET/MRI scans were obtained for 12 patients after PET/CT scans. Only data acquired at a bed station that included the head were used for this study. Using PET data from PET/MRI, we applied ZTACUC, ZTACSEC, ATAC, and reference CT-based attenuation correction (CTAC) to PET attenuation correction. For ZTACUC, the bias-corrected and normalized ZTE was converted to pseudo-CT with air (-1,000 HU for ZTE < 0.2), soft-tissue (42 HU for ZTE > 0.75), and bone (-2,000 × [ZTE - 1] + 42 HU for 0.2 ≤ ZTE ≤ 0.75). Afterward, in the pseudo-CT, sinus/edges were automatically estimated as a binary mask through morphologic processing and edge detection. In the binary mask, the overestimated values were rescaled below 42 HU for ZTACSEC For ATAC, the atlas deformed to MR in-phase was segmented to air, inner air, soft tissue, and continuous bone. For the quantitative evaluation, PET mean uptake values were measured in twenty 1-mL volumes of interest distributed throughout brain tissues. The PET uptake was compared using a paired t test. An error histogram was used to show the distribution of voxel-based PET uptake differences. Results: Compared with CTAC, ZTACSEC achieved the overall PET quantification accuracy (0.2% ± 2.4%, P = 0.23) similar to CTAC, in comparison with ZTACUC (5.6% ± 3.5%, P < 0.01) and ATAC (-0.9% ± 5.0%, P = 0.03). Specifically, a substantial improvement with ZTACSEC (0.6% ± 2.7%, P < 0.01) was found in the cerebellum, in comparison with ZTACUC (8.1% ± 3.5%, P < 0.01) and ATAC (-4.1% ± 4.3%, P < 0.01). The histogram of voxel-based uptake differences demonstrated that ZTACSEC reduced the magnitude and variation of errors substantially, compared with ZTACUC and ATAC. Conclusion: ZTACSEC can provide an accurate PET quantification in brain PET/MRI, comparable to the accuracy achieved by CTAC, particularly in the cerebellum.