CT金属伪影减少技术改善PET/CT图像质量
本研究通过体模和患者扫描,评估CT迭代金属伪影减少(iMAR)算法对PET图像的影响。体模中18F-FDG浓度冷伪影导致30%缺失,经iMAR处理后被恢复;患者解读置信度平均提升28%±20%,病灶SUVmean绝对值平均差异为3.5%±3.3%。结论认为iMAR可提高PET/CT解读置信度并影响SUV。
要点
- 体模中金属伪影导致PET图像18F-FDG浓度缺失30%,iMAR处理后可恢复。
- 患者中使用iMAR后解读置信度增加或保持不变,平均改善28%±20%。
- 病灶SUVmean绝对值平均差异为3.5%±3.3%,SUV变化取决于金属伪影类型。
核心结论
- iMAR算法提高PET/CT扫描的解读置信度并影响SUV。
- iMAR的附加价值取决于PET/CT检查的适应证、假体的位置和大小/类型以及病变的位置和范围。
研究设计与样本
- 体模研究:2种髋关节假体置于18F-FDG水溶液中。
- 患者研究:14例金属植入物患者行18F-FDG PET/CT,7例髋关节假体患者行68Ga-PSMA PET/CT,由2名经验丰富的核医学医师评分。
- 所有患者病灶均位于金属植入物视野内,标准低剂量CT经iMAR算法处理,PET数据分别使用标准CT和iMAR处理CT提供的衰减校正重建。
主要发现
- 体模扫描中PET图像可见冷伪影,18F-FDG浓度缺失30%,经iMAR处理后被恢复。
- iMAR算法对大多数患者有用;使用iMAR时解读置信度增加或保持不变,平均改善28%±20%(置信度评分范围0%-100%)。
- SUV的升高或降低取决于金属伪影类型;病灶SUVmean绝对值的平均差异为3.5%±3.3%。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
In recent years, different metal artifact reduction methods have been developed for CT. These methods have only recently been introduced for PET/CT even though they could be beneficial for interpretation, segmentation, and quantification of the PET/CT images. In this study, phantom and patient scans were analyzed visually and quantitatively to measure the effect on PET images of iterative metal artifact reduction (iMAR) of CT data. Methods: The phantom consisted of 2 types of hip prostheses in a solution of 18F-FDG and water. 18F-FDG PET/CT scans of 14 patients with metal implants (either dental implants, hip prostheses, shoulder prostheses, or pedicle screws) and 68Ga-labeled prostate-specific membrane antigen (68Ga-PSMA) PET/CT scans of 7 patients with hip prostheses were scored by 2 experienced nuclear medicine physicians to analyze clinical relevance. For all patients, a lesion was located in the field of view of the metal implant. Phantom and patients were scanned in a PET/CT scanner. The standard low-dose CT scans were processed with the iMAR algorithm. The PET data were reconstructed using attenuation correction provided by both standard CT and iMAR-processed CT. Results: For the phantom scans, cold artifacts were visible on the PET image. There was a 30% deficit in 18F-FDG concentration, which was restored by iMAR processing, indicating that metal artifacts on CT images induce quantification errors in PET data. The iMAR algorithm was useful for most patients. When iMAR was used, the confidence in interpretation increased or stayed the same, with an average improvement of 28% ± 20% (scored on a scale of 0%-100% confidence). The SUV increase or decrease depended on the type of metal artifact. The mean difference in absolute values of SUVmean of the lesions was 3.5% ± 3.3%. Conclusion: The iMAR algorithm increases the confidence of the interpretation of the PET/CT scan and influences the SUV. The added value of iMAR depends on the indication for the PET/CT scan, location and size/type of the prosthesis, and location and extent of the disease.
引用
- van der Vos CS, Arens AIJ, Hamill JJ, et al. Metal Artifact Reduction of CT Scans to Improve PET/CT. J Nucl Med. 2017. PMID: 28490470. doi:10.2967/jnumed.117.191171
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。