迭代反卷积联合HYPR去噪对脑PET研究进行部分容积校正

Partial volume correction of brain PET studies using iterative deconvolution in combination with HYPR denoising.
PubMed 文献 · EJNMMI Res · 2017-04-21

本研究在Hoffman脑模体上优化了Van Cittert迭代反卷积法(IDM)联合高约束反投影(HYPR)去噪的参数与顺序,并在健康受试者的[11C]氟马西尼和健康受试者及阿尔茨海默病患者的[11C]PIB动态扫描中评估其对分布容积(VT)的影响。结果显示Hypr-IDM-Hypr使[11C]氟马西尼扫描的区域VT较未校正扫描增加5%至20%,而[11C]PIB扫描的VT变化为0%至10%增减,取决于感兴趣区或受试者类型。结论认为IDM提高了活性浓度的定量准确性,且联合HYPR可在不增加噪声的情况下校正部分容积效应。

要点

  • 在Hoffman脑模体上优化了Van Cittert IDM联合HYPR去噪的参数与实施顺序。
  • [11C]氟马西尼扫描中Hypr-IDM-Hypr使区域VT较未校正扫描增加5%至20%。
  • [11C]PIB扫描中VT较未校正扫描变化为0%至10%增减,取决于感兴趣区或受试者类型。

核心结论

  • IDM提高了实测活性浓度的定量准确性,Hypr-IDM-Hypr能够在不增加噪声的情况下校正部分容积效应。

研究设计与样本

  • 在Philips Ingenuity TF PET/CT上以动态脑协议扫描填充约25 kBq mL-1 [18F]FDG溶液的Hoffman脑模体30分钟,帧时长10至300秒,用于优化IDM参数及HYPR实施顺序。
  • 使用动态[11C]氟马西尼(5名健康受试者)和[11C]PIB(4名健康受试者和4名阿尔茨海默病患者)扫描,评估IDM联合或不联合HYPR对血浆输入法衍生分布容积(VT)的影响。

主要发现

  • [11C]氟马西尼扫描中,Hypr-IDM-Hypr使区域VT较未校正扫描增加5%至20%。
  • [11C]PIB扫描中,VT较未校正扫描变化为0%至10%增减,取决于感兴趣区或受试者类型(健康或患者)。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

BACKGROUND: Accurate quantification of PET studies depends on the spatial resolution of the PET data. The commonly limited PET resolution results in partial volume effects (PVE). Iterative deconvolution methods (IDM) have been proposed as a means to correct for PVE. IDM improves spatial resolution of PET studies without the need for structural information (e.g. MR scans). On the other hand, deconvolution also increases noise, which results in lower signal-to-noise ratios (SNR). The aim of this study was to implement IDM in combination with HighlY constrained back-PRojection (HYPR) denoising to mitigate poor SNR properties of conventional IDM. METHODS: An anthropomorphic Hoffman brain phantom was filled with an [18F]FDG solution of ~25 kBq mL-1 and scanned for 30 min on a Philips Ingenuity TF PET/CT scanner (Philips, Cleveland, USA) using a dynamic brain protocol with various frame durations ranging from 10 to 300 s. Van Cittert IDM was used for PVC of the scans. In addition, HYPR was used to improve SNR of the dynamic PET images, applying it both before and/or after IDM. The Hoffman phantom dataset was used to optimise IDM parameters (number of iterations, type of algorithm, with/without HYPR) and the order of HYPR implementation based on the best average agreement of measured and actual activity concentrations in the regions. Next, dynamic [11C]flumazenil (five healthy subjects) and [11C]PIB (four healthy subjects and four patients with Alzheimer's disease) scans were used to assess the impact of IDM with and without HYPR on plasma input-derived distribution volumes (V T) across various regions of the brain. RESULTS: In the case of [11C]flumazenil scans, Hypr-IDM-Hypr showed an increase of 5 to 20% in the regional V T whereas a 0 to 10% increase or decrease was seen in the case of [11C]PIB depending on the volume of interest or type of subject (healthy or patient). References for these comparisons were the V Ts from the PVE-uncorrected scans. CONCLUSIONS: IDM improved quantitative accuracy of measured activity concentrations. Moreover, the use of IDM in combination with HYPR (Hypr-IDM-Hypr) was able to correct for PVE without increasing noise.