文献速递肿瘤 PET/CT放射性药物与辐射防护

基于下一代硅光电倍增管的飞行时间PET/CT系统的性能研究

Studies of a Next-Generation Silicon-Photomultiplier-Based Time-of-Flight PET/CT System.
PubMed 文献 · J Nucl Med · 2017-04-27

本文报告了基于硅光电倍增管的飞行时间PET/CT系统Discovery MI的系统性能研究,在斯坦福大学和乌普萨拉大学的两台扫描仪上独立测试并取平均值,并与其它商用PET/CT和PET/MR系统及不同重建算法进行比较。结果显示该系统具有375.4 ps FWHM符合时间分辨率和13.7 cps/kBq中心灵敏度等优异性能,Q.Clear重建较OSEM显著提升图像质量和对比噪声性能。

要点

  • Discovery MI的平均符合时间分辨率为375.4 ps FWHM,视野中心平均灵敏度为13.7 cps/kBq。
  • 平均峰值噪声等效计数率为193.4 kcps(对应21.9 kBq/mL),散射分数为40.6%。
  • Q.Clear重建较有序子集期望最大化显著提升图像质量和对比噪声性能。

核心结论

  • Discovery MI系统取得了优异性能,包括375 ps FWHM符合时间分辨率和14 cps/kBq灵敏度。
  • 与其它重建算法及多模态硅光电倍增管和非硅光电倍增管PET探测器系统设计的比较表明,该下一代系统可显著提升性能。

研究设计与样本

  • 本研究为系统性能研究,采用美国电气制造商协会NU-2 2012标准表征空间分辨率、灵敏度、噪声等效计数率、散射分数、计数率准确性和图像质量。
  • 测量了能量分辨率和符合时间分辨率,测试在斯坦福大学和乌普萨拉大学安装的两台Discovery MI扫描仪上独立进行,结果取平均值。
  • 在Discovery MI、Discovery 690 PET/CT和SIGNA PET/MR系统之间进行了背靠背患者扫描。
  • 临床图像分别使用有序子集期望最大化和Q.Clear(带点扩散函数建模的块序贯正则化期望最大化)重建并进行定性检查。

主要发现

  • 滤波反投影重建的径向/切向/轴向空间分辨率半高全宽平均值在距系统中心1 cm处分别为4.10/4.19/4.48 mm,10 cm处分别为5.47/4.49/6.01 mm,20 cm处分别为7.53/4.90/6.10 mm。
  • 视野中心平均灵敏度为13.7 cps/kBq,平均峰值噪声等效计数率为193.4 kcps(对应21.9 kBq/mL),散射分数为40.6%。
  • 图像质量体模的平均对比恢复系数在10、13、17、22、28和37 mm直径球体上分别为53.7、64.0、73.1、82.7、86.8和90.7。
  • 平均光电峰能量分辨率为9.40% FWHM,平均符合时间分辨率为375.4 ps FWHM。
  • PET/CT系统间临床图像比较显示Discovery MI图像质量高;Discovery MI与SIGNA比较显示相似的空间分辨率和整体成像性能;Q.Clear较有序子集期望最大化显著增强图像质量和对比噪声性能。

关键数值与推荐

  • 滤波反投影重建下距系统中心1 cm处径向/切向/轴向空间分辨率半高全宽平均值分别为4.10/4.19/4.48 mm。
  • 滤波反投影重建下距系统中心10 cm处径向/切向/轴向空间分辨率半高全宽平均值分别为5.47/4.49/6.01 mm。
  • 滤波反投影重建下距系统中心20 cm处径向/切向/轴向空间分辨率半高全宽平均值分别为7.53/4.90/6.10 mm。
  • 视野中心平均灵敏度为13.7 cps/kBq,平均峰值噪声等效计数率为193.4 kcps(对应21.9 kBq/mL),散射分数为40.6%。
  • 平均光电峰能量分辨率为9.40% FWHM,平均符合时间分辨率为375.4 ps FWHM。

临床怎么用

  • 原文未据此给出推荐。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

This article presents system performance studies for the Discovery MI PET/CT system, a new time-of-flight system based on silicon photomultipliers. System performance and clinical imaging were compared between this next-generation system and other commercially available PET/CT and PET/MR systems, as well as between different reconstruction algorithms. Methods: Spatial resolution, sensitivity, noise-equivalent counting rate, scatter fraction, counting rate accuracy, and image quality were characterized with the National Electrical Manufacturers Association NU-2 2012 standards. Energy resolution and coincidence time resolution were measured. Tests were conducted independently on two Discovery MI scanners installed at Stanford University and Uppsala University, and the results were averaged. Back-to-back patient scans were also performed between the Discovery MI, Discovery 690 PET/CT, and SIGNA PET/MR systems. Clinical images were reconstructed using both ordered-subset expectation maximization and Q.Clear (block-sequential regularized expectation maximization with point-spread function modeling) and were examined qualitatively. Results: The averaged full widths at half maximum (FWHMs) of the radial/tangential/axial spatial resolution reconstructed with filtered backprojection at 1, 10, and 20 cm from the system center were, respectively, 4.10/4.19/4.48 mm, 5.47/4.49/6.01 mm, and 7.53/4.90/6.10 mm. The averaged sensitivity was 13.7 cps/kBq at the center of the field of view. The averaged peak noise-equivalent counting rate was 193.4 kcps at 21.9 kBq/mL, with a scatter fraction of 40.6%. The averaged contrast recovery coefficients for the image-quality phantom were 53.7, 64.0, 73.1, 82.7, 86.8, and 90.7 for the 10-, 13-, 17-, 22-, 28-, and 37-mm-diameter spheres, respectively. The average photopeak energy resolution was 9.40% FWHM, and the average coincidence time resolution was 375.4 ps FWHM. Clinical image comparisons between the PET/CT systems demonstrated the high quality of the Discovery MI. Comparisons between the Discovery MI and SIGNA showed a similar spatial resolution and overall imaging performance. Lastly, the results indicated significantly enhanced image quality and contrast-to-noise performance for Q.Clear, compared with ordered-subset expectation maximization. Conclusion: Excellent performance was achieved with the Discovery MI, including 375 ps FWHM coincidence time resolution and sensitivity of 14 cps/kBq. Comparisons between reconstruction algorithms and other multimodal silicon photomultiplier and non-silicon photomultiplier PET detector system designs indicated that performance can be substantially enhanced with this next-generation system.

来源:J Nucl Med,查看原文。版权归原作者所有。