金属植入物相关磁化率伪影对同步TOF PET/MR中邻近病灶评估的影响
研究纳入27例无植入物但存在18F-FDG高摄取病灶邻近常见植入部位的患者,通过在人造信号空洞的衰减校正图中模拟金属植入物,比较TOF与非TOF重建PET图像对邻近病灶检测的影响。结果显示TOF信息可显著减少股骨头、胸骨和脊柱的模拟植入物伪影并改善图像质量,且TOF重建的SUVmax平均百分比误差较非TOF降低28.5%(P<0.01)。
要点
- TOF信息显著减少股骨头、胸骨和脊柱的模拟植入物伪影(P分别为0.01、0.01、0.03)。
- TOF信息改善上述部位图像质量(P分别为0.02、0.01、0.01)。
- TOF使SUVmax平均百分比误差降低28.5%(P<0.01),对较小、中度18F-FDG高摄取病灶影响更明显。
核心结论
- 定性上,股骨头、胸骨和脊柱的伪影显著减少,这些部位的图像质量也有显著改善。
- 模拟的金属植入物相关磁化率伪影对植入部位附近小的、中度18F-FDG高摄取病灶有显著影响,若无TOF信息这些病灶可能被忽略。
- 对于较大的、高度18F-FDG高摄取病灶,金属植入物的影响有限。
- 胸骨伪影中发现了最大的显著定量差异。
- 受伪影影响的病灶与距植入物距离之间仅存在弱负相关。
研究设计与样本
- 纳入27例无植入物但经确认存在18F-FDG高摄取病灶且邻近常见植入部位的患者,所有患者均接受临床指征的全身18F-FDG PET/MR扫描。
- 重建基线非TOF和TOF PET图像,随后在衰减校正图中引入人造信号空洞以模拟标准解剖区域的金属植入物并重复重建。
- 对所有重建图像进行定性和定量评估,并与基线图像比较,共评估51个病灶。
主要发现
- 在非TOF和TOF重建中,分别有40个和50个病灶的可检测性未改变,分别有9个和1个病灶的可检测性改变,2个非TOF病灶在引入金属伪影后不再可见。
- TOF信息的纳入显著减少了股骨头、胸骨和脊柱的模拟植入物所致伪影(P分别为0.01、0.01、0.03),并改善了这些部位的图像质量(P分别为0.02、0.01、0.01)。
- TOF重建的平均百分比误差为-3.5%,非TOF重建为-4.8%,即TOF信息的纳入使SUVmax的百分比误差降低28.5%(P<0.01)。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
Metalic implants may affect attenuation correction (AC) in PET/MR imaging. The purpose of this study was to evaluate the effect of susceptibility artifacts related to metallic implants on adjacent metabolically active lesions in clinical simultaneous PET/MR scanning for both time-of-flight (TOF) and non-TOF reconstructed PET images. Methods: We included 27 patients without implants but with confirmed 18F-FDG-avid lesions adjacent to common implant locations. In all patients, a clinically indicated whole-body 18F-FDG PET/MR scan was acquired. Baseline non-TOF and TOF PET images were reconstructed. Reconstruction was repeated after the introduction of artificial signal voids in the AC map to simulate metallic implants in standard anatomic areas. All reconstructed images were qualitatively and quantitatively assessed and compared with the baseline images. Results: In total, 51 lesions were assessed. In 40 and 50 of these cases (non-TOF and TOF, respectively), the detectability of the lesions did not change; in 9 and 1 cases, the detectability changed; and in 2 non-TOF cases, the lesions were no longer visible after the introduction of metallic artifacts. The inclusion of TOF information significantly reduced artifacts due to simulated implants in the femoral head, sternum, and spine (P = 0.01, 0.01, and 0.03, respectively). It also improved image quality in these locations (P = 0.02, 0.01, and 0.01, respectively). The mean percentage error was -3.5% for TOF and -4.8% for non-TOF reconstructions, meaning that the inclusion of TOF information reduced the percentage error in SUVmax by 28.5% (P < 0.01). Conclusion: Qualitatively, there was a significant reduction of artifacts in the femoral head, sternum, and spine. There was also a significant qualitative improvement in image quality in these locations. Furthermore, our study indicated that simulated susceptibility artifacts related to metallic implants have a significant effect on small, moderately 18F-FDG-avid lesions near the implant site that possibly may go unnoticed without TOF information. On larger, highly 18F-FDG-avid lesions, the metallic implants had only a limited effect. The largest significant quantitative difference was found in artifacts of the sternum. There was only a weak inverse correlation between lesions affected by artifacts and distance from the implant.