文献速递肿瘤 PET/CT放射性药物与辐射防护

基于患者个体特征的18F-FDG PET显像给药活度与采集时间优化

Patient-specific optimisation of administered activity and acquisition times for 18F-FDG PET imaging.
PubMed 文献 · EJNMMI Res · 2017-01-13

本研究纳入111例FDG扫描患者,通过生物阻抗测量与图像质量主客观评估,确定以每米噪声等效计数(NEC)为目标指标,建立按性别、身高、体重推算给药活度与采集时间的表达式;应用该表达式可使该组患者平均给药活度较现行方案降低66 MBq。

要点

  • 每米NEC与主观图像质量评估相关性最佳(Spearman秩相关系数0.77)。
  • 每米NEC鉴别「肯定合格」图像曲线下面积0.94,目标值37 Mcount/m。
  • 按性别、身高、体重推算活度与采集时间,可使平均给药活度较现行方案降低66 MBq。

核心结论

  • 已建立可获得一致图像质量的FDG显像给药活度与采集时间表达式,可降低患者与工作人员剂量。
  • 这些表达式可能需要针对其他系统和重建方案进行调整。

研究设计与样本

  • 纳入转诊行FDG扫描的患者,进行生物阻抗测量。
  • 注射3 MBq/kg的18F,最高370 MBq,在Siemens Biograph mCT上按每床3或4分钟采集。
  • 数据重分组模拟2分钟和1分钟采集,由经验丰富的医师主观评估图像质量,并与基于信噪比和噪声等效计数(NEC)的客观测量比较。
  • 共纳入111例患者,比较321幅全时间与缩短时间扫描的主客观图像质量评估。
  • 采用多元回归分析,以易测量的患者特征表达所需活度和采集时间。

主要发现

  • 每米NEC是与主观评估相关性最佳的客观指标(Spearman秩相关系数0.77)。
  • 每米NEC是鉴别主观评估为「肯定合格」图像的最佳判别指标(ROC曲线下面积0.94)。
  • 确定目标值为37 Mcount/m。
  • 以患者性别、身高和体重为变量建立了达到该目标所需活度和采集时间的表达式。
  • 将这些表达式用于该患者组,可使平均给药活度较现行方案降低66 MBq。

关键数值与推荐

  • 目标客观图像质量指标为每米NEC 37 Mcount/m。
  • 每米NEC与主观图像质量评估的Spearman秩相关系数为0.77。
  • 每米NEC鉴别「肯定合格」图像的ROC曲线下面积为0.94。
  • 在表达式中纳入身体成分测量没有用处。

临床怎么用

  • 原文未据此给出推荐。

局限性

  • 这些表达式可能需要针对其他系统和重建方案进行调整。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

BACKGROUND: The purpose of this study is to identify a method for optimising the administered activity and acquisition time for 18F-FDG PET imaging, yielding images of consistent quality for patients with varying body sizes and compositions, while limiting radiation doses to patients and staff. Patients referred for FDG scans had bioimpedance measurements. They were injected with 3 MBq/kg of 18F up to 370 MBq and scanned on a Siemens Biograph mCT at 3 or 4 min per bed position. Data were rebinned to simulate 2- and 1-min acquisitions. Subjective assessments of image quality made by an experienced physician were compared with objective measurements based on signal-to-noise ratio and noise equivalent counts (NEC). A target objective measure of image quality was identified. The activity and acquisition time required to achieve this were calculated for each subject. Multiple regression analysis was used to identify expressions for the activity and acquisition time required in terms of easily measurable patient characteristics. RESULTS: One hundred and eleven patients were recruited, and subjective and objective assessments of image quality were compared for 321 full and reduced time scans. NEC-per-metre was identified as the objective measure which best correlated with the subjective assessment (Spearman rank correlation coefficient 0.77) and the best discriminator for images with a subjective assessment of "definitely adequate" (area under the ROC curve 0.94). A target of 37 Mcount/m was identified. Expressions were identified in terms of patient sex, height and weight for the activity and acquisition time required to achieve this target. Including measurements of body composition in these expressions was not useful. Using these expressions would reduce the mean activity administered to this patient group by 66 MBq compared to the current protocol. CONCLUSIONS: Expressions have been identified for the activity and acquisition times required to achieve consistent image quality in FDG imaging with reduced patient and staff doses. These expressions might need to be adapted for other systems and reconstruction protocols.

来源:EJNMMI Res,查看原文。版权归原作者所有。

引用

  1. Wickham F, McMeekin H, Burniston M, et al. Patient-specific optimisation of administered activity and acquisition times for 18F-FDG PET imaging. EJNMMI Res. 2017. PMID: 28091978. doi:10.1186/s13550-016-0250-3

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。