长轴视野PET系统上16门控较8门控提高82Rb PET心功能参数准确性:与超声心动图验证
该研究回顾性纳入124例在长轴视野PET/CT系统上接受静息-负荷82Rb PET/CT的连续患者,静息图像分别用8门控和16门控重建,并与中位26天内的经胸超声心动图(TTE)对比。结果显示16门控PET的LVEF与TTE无显著差异(p=0.24),而8门控PET显著低估LVEF(平均偏差-5.49%,p<0.001)。16门控方案在低82Rb活度下即可改善左心室功能评估准确性。
要点
- 16门控82Rb PET的LVEF与TTE无显著差异,8门控显著低估LVEF。
- 16门控改善ESV准确性,8门控显著高估ESV。
- 低82Rb活度下即可实现16门控方案,不增加辐射负担。
核心结论
- 在长轴视野PET/CT上,将ECG门控从8帧增至16帧可显著提高左心室功能评估准确性,尤其ESV和LVEF,与TTE高度一致。
研究设计与样本
- 回顾性纳入124例连续患者,在长轴视野PET/CT系统上行临床指征的静息-负荷82Rb PET/CT,静息图像分别用8门控和16门控重建,与中位26天内的TTE对比。
- 所有患者静息和负荷采集均接受370 MBq 82Rb。
- 采用Wilcoxon检验、Pearson相关和Bland-Altman分析评估一致性。
主要发现
- 静息LVEF均值:8门控PET为52.3±15.5%,16门控PET为58.1±13.5%,TTE为57.7±9.7%。
- 16门控PET的LVEF与TTE无显著差异(p=0.24),相关性一般(r=0.58),偏差极小(+1.13%)。
- 8门控PET显著低估LVEF(平均偏差-5.49%,p<0.001;r=0.38),尤其在低LVEF值时。
- EDV在8门控与16门控PET间相当,但两者较TTE均轻度高估。
- 8门控PET显著高估ESV(p=0.002),16门控PET与TTE无统计学显著差异(p=0.085)。
临床怎么用
- 在长轴视野PET/CT系统上,推荐采用16门控协议进行82Rb PET心功能评估,以提高LVEF和ESV准确性且不增加辐射负担。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
INTRODUCTION: Left ventricular (LV) ejection fraction (EF) and volumes derived from ECG-gated 82Rb PET are clinically relevant, but their assessment can be impaired due to limited temporal sampling and low count statistics. Long axial field-of-view (LAFOV) PET/CT systems provide substantially higher temporal resolution compared to standard, analog PET systems, which allows to yield higher sensitivity without the need for increasing tracer activity. We aimed to evaluate, whether increasing ECG gating from 8 to 16 frames on 82Rb LAFOV PET improves the accuracy of LV functional parameters. To that end, we used transthoracic echocardiography (TTE) as reference of standard. MATERIALS AND METHODS: We retrospectively included 124 consecutive patients undergoing clinically indicated rest–stress 82Rb PET/CT on a LAFOV system. All patients received 370 MBq 82Rb for both rest and stress acquisitions. Rest images were reconstructed using both 8- and 16-gated frames. PET-derived LVEF, end-diastolic volume (EDV), and end-systolic volume (ESV) were compared with TTE values obtained within a median of 26 days. Agreement was assessed using Wilcoxon test, Pearson correlation, and Bland–Altman analysis. RESULTS: Mean rest LVEF was 52.3 ± 15.5% with 8-gated PET, 58.1 ± 13.5% with 16-gated PET, and 57.7 ± 9.7% on TTE. LVEF derived from 16-gated PET showed no significant difference compared with TTE (p = 0.24), with fair correlation (r = 0.58) and minimal bias (+ 1.13%). In contrast, 8-gated PET significantly underestimated LVEF compared with TTE (mean bias − 5.49%, p < 0.001; r = 0.38), particularly at lower LVEF values. EDV was comparable between 8- and 16-gated PET but showed modest overestimation versus TTE for both reconstructions. For ESV, 8-gated PET significantly overestimated values compared with TTE (p = 0.002), whereas 16-gated PET showed improved agreement with no statistically significant difference (p = 0.085). CONCLUSION: On LAFOV PET/CT, increasing temporal resolution from 8 to 16 ECG-gated frames significantly improves the accuracy of LV functional assessment, particularly for ESV and LVEF, yielding close agreement with TTE. These improvements were achieved using low 82Rb activities, supporting the routine clinical implementation of 16-gated protocols on LAFOV systems to enhance functional cardiac assessment without increasing radiation burden.