超高灵敏度与连续床移动对Biograph Vision Quadra PET/CT性能特征的影响
本研究通过体模测量和7例肿瘤患者[18F]FDG PET/CT,评估超高灵敏度(UHS)与连续床移动(CBM)对恢复系数(RC)和空间分辨率(SR)的影响。UHS和CBM的平均RC分别低于HS和静态采集,CBM下SR差异可忽略,健康组织与病灶SUV相关性优良。作者认为采用UHS和CBM的精细方案可迈向快速全身PET检查,仅噪声轻度增加。
要点
- UHS模式平均RC低于HS模式(<4.2%),CBM模式平均RC低于静态采集(<5.9%)。
- CBM下空间分辨率差异可忽略(半高全宽差异<0.13 mm)。
- 健康组织与病灶SUV相关性优良(r²>0.94和r²>0.99),仅肝脏变异系数有显著差异。
核心结论
- 采用UHS和CBM的精细方案可迈向快速全身PET检查,仅噪声轻度增加。
- CBM允许检查更大身体范围,并有助于减轻沿轴视野的灵敏度差异,但代价是RC略低。
研究设计与样本
- 体模测量在静态床和CBM采集(床速3.5、7、14、30 mm/s)下进行,并分别以HS和UHS模式重建。
- 另纳入7例临床转诊肿瘤患者,行CBM联合UHS的[18F]FDG PET/CT扫描,比较不同采集模式的病灶和健康组织指标。
主要发现
- UHS模式平均RC低于HS模式(<4.2%),CBM模式平均RC低于静态采集(<5.9%)。
- UHS模式空间分辨率在轴视野中心平面轻度下降(半高全宽差异范围0.06-0.21 mm),向边缘则与HS模式相当(半高全宽差异范围0.03-0.07 mm)。
- CBM下空间分辨率差异可忽略(半高全宽差异<0.13 mm)。
- 健康组织SUV(r²>0.94)和病灶SUV(r²>0.99)相关性优良。
- 肝脏的变异系数有显著差异。
临床怎么用
- 原文未据此给出推荐。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
Continuous bed motion (CBM) for long-axial-field-of-view PET/CT enables swift total-body examination of patients. However, the sensitivity profile along the axial field of view (AFOV) varies significantly, and its effect on image quality when combined with CBM remains unexplored. This study assesses the effects of ultrahigh sensitivity (UHS) and CBM on recovery coefficients (RCs) and spatial resolution (SR). Methods: Phantom measurements, performed in static bed and CBM acquisition with different bed speeds (3.5, 7, 14, and 30 mm/s), were reconstructed for both high-sensitivity (HS) and UHS modes, and their quality was assessed using the RC and full width at half maximum of the point-spread function. In addition, 7 clinically referred oncologic patients underwent a combined CBM and UHS [18F]FDG PET/CT scan. Metrics derived from lesions and healthy tissues were compared across acquisition modes. Results: Mean RC was lower in both UHS (<4.2%) and CBM (<5.9%) than with HS and static acquisitions, respectively. SR in UHS was slightly deteriorated in the central plane of the AFOV (full width at half maximum difference range, 0.06-0.21 mm) but remained comparable toward the edge (full width at half maximum difference range, 0.03-0.07 mm) to SR in HS. In addition, for CBM, SR differences were negligible (full width at half maximum difference, <0.13 mm). SUVs in healthy tissues (r 2 > 0.94) and lesions (r 2 > 0.99) showed excellent correlation. The coefficient of variation was significantly different for the liver. Conclusion: Elaborate protocols using UHS and CBM could be a step toward swift total-body PET care with only a marginal increase in noise. CBM allows a larger part of the body to be examined and helps to mitigate substantial sensitivity differences along the AFOV, albeit at the cost of slightly lower RC.