长轴视野PET/CT系统超低活度18F-FDG PET/CT显像的可行性
本研究为高-低活度重复扫描试验,11例肿瘤患者先后接受标准活度(3.0 MBq/kg)与超低活度(0.3 MBq/kg)18F-FDG PET/CT显像,并将列表模式数据重采样模拟0.3与0.03 MBq/kg。结果显示SUVmean在各活度与重建方案下保持稳定,SUVpeak除模拟标准活度1%时外保持稳定,SUVmax在低活度下显著升高;SUVmean与SUVpeak在0.03 MBq/kg时变异性仍<15%,而SUVmax常超出该限。0.3 MBq/kg病灶检出可靠,0.03 MBq/kg时部分病灶无法勾画。
要点
- 超低活度0.3 MBq/kg在长轴视野PET/CT上行18F-FDG显像可行,病灶检出可靠。
- SUVmean在各活度与重建方案下稳定,SUVpeak除模拟标准活度1%时外稳定。
- SUVmax在低活度下显著升高;SUVmean与SUVpeak在0.03 MBq/kg时变异性仍<15%。
核心结论
- 超低活度18F-FDG(0.3 MBq/kg)在长轴视野PET/CT系统上行PET/CT显像可行,得到半定量测量、重复性分析、病灶可检出性与半自动勾画方法的支持。
- 这些发现支持将辐射暴露保持在合理或实际可达到的最低水平的显像方案,对辐射敏感患者(如儿童、孕妇)尤其相关。
研究设计与样本
- 高-低活度重复扫描试验,11例肿瘤患者先后接受2次18F-FDG PET/CT显像。
- 第一次使用标准活度3.0 MBq/kg,7天内进行1次超低活度0.3 MBq/kg显像。
- 两次扫描的列表模式数据均重采样以模拟0.3与0.03 MBq/kg活度。
- 比较不同活度与重建方案下健康器官与病灶的SUVmean、SUVpeak、SUVmax及其重复性。
主要发现
- SUVmean在各活度降低与重建方案下保持稳定。
- SUVpeak保持稳定,但模拟标准18F-FDG活度1%时除外。
- SUVmax在较低18F-FDG活度下显著升高,尤其在临床偏好的扫描中。
- 重复性分析显示SUVmean与SUVpeak即使在0.03 MBq/kg时变异性仍保持在可接受范围内(<15%),而SUVmax变异性常超出该限。
- 0.3 MBq/kg时病灶检出可靠,但0.03 MBq/kg时若干病灶无法勾画,提示图像质量受损。
关键数值与推荐
- 标准活度为3.0 MBq/kg,超低活度为0.3 MBq/kg。
- 模拟活度为0.3与0.03 MBq/kg,后者相当于标准活度的1%。
- SUVmean与SUVpeak在0.03 MBq/kg时变异性<15%。
临床怎么用
- 作者结论认为超低活度18F-FDG(0.3 MBq/kg)在长轴视野PET/CT系统上行PET/CT显像可行。
- 作者结论认为这些发现支持将辐射暴露保持在合理或实际可达到的最低水平的显像方案,对辐射敏感患者(如儿童、孕妇)尤其相关。
局限性
- 0.03 MBq/kg时若干病灶无法勾画,提示图像质量受损。
- 模拟标准18F-FDG活度1%时SUVpeak未保持稳定。
- SUVmax在较低18F-FDG活度下显著升高,其变异性常超出可接受范围。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
18F-FDG PET/CT imaging is widely used in oncology. Long-axial-field-of-view (LAFOV) PET/CT systems exhibit ultrahigh sensitivity and signal-to-noise ratios, enabling significant reductions in tracer activity and shorter acquisition times. This study aimed to evaluate the feasibility of using ultralow activities for semiquantitative measurements in 18F-FDG PET/CT imaging performed on an LAFOV PET/CT system through a high-low activity test-retest study. Methods: Eleven oncology patients underwent 2 18F-FDG PET/CT scans, the first with standard activity (3.0 MBq/kg) and, within 7 d, 1 with ultralow activity (0.3 MBq/kg). List-mode data of both scans were resampled to simulate activities of 0.3 and 0.03 MBq/kg. Semiquantitative measurements (SUVmean, SUVpeak, and SUVmax) and their repeatability in healthy organs and lesions were compared across different activities and reconstruction protocols. Results: SUVmean remained stable across activity reductions and reconstruction protocols, whereas SUVpeak showed stability except when simulating 1% of the standard 18F-FDG activity. SUVmax significantly increased at lower 18F-FDG activities, particularly with clinically preferred scans. Repeatability analysis showed that SUVmean and SUVpeak maintained variability within acceptable limits (<15%) even at 0.03 MBq/kg, whereas SUVmax variability often exceeded these limits. Lesion detection was reliable at 0.3 MBq/kg, but several lesions could not be delineated at 0.03 MBq/kg, indicating compromised image quality. Conclusion: The use of ultralow 18F-FDG activity (0.3 MBq/kg) is feasible for PET/CT imaging on an LAFOV PET/CT system, as shown by semiquantitative measurements, repeatability analyses, lesion detectability, and semiautomated delineation methods. These findings support imaging protocols that keep radiation exposure levels as low as reasonably or practically achievable, which are particularly relevant for radiation-sensitive patients (e.g., children, pregnant women).