构建用于胃肠道肿瘤检测的高灵敏度切伦科夫发光内镜系统
本研究构建了高灵敏度切伦科夫发光内镜(CLE)系统,在体内300秒内可分辨最低放射性活度达0.020微居里,灵敏度较先前系统提升近22倍。在荷瘤裸鼠中,CLE与组织病理学和PET/CT一致率为100%,肿瘤CL信号显著强于周围正常组织(P<0.05)。60秒成像质量与300秒相当(信噪比2.70±0.48 vs 2.98±0.69,P=0.56)。
要点
- CLE系统在体内300秒内最低可分辨放射性活度为0.020微居里,首次达到纳居里级灵敏度。
- CLE检测灵敏度较先前系统提高近22倍。
- 荷瘤裸鼠中CLE识别所有肿瘤,与组织病理学和PET/CT一致率100%。
核心结论
- 构建的高灵敏度CLE系统可检测纳居里放射性活度的放射性核素,通过快速分子成像准确检测癌症,并具有在临床实践中早期诊断胃肠道肿瘤的潜力。
研究设计与样本
- 研究构建了用于胃肠道肿瘤分子成像的高灵敏度切伦科夫发光内镜系统,并在荷瘤裸鼠中进行体内评价。
主要发现
- CLE在体内300秒内最低可分辨放射性活度达0.020微居里。
- CLE检测灵敏度较先前系统提高近22倍。
- 荷瘤裸鼠中CLE有效识别全部肿瘤,与组织病理学和PET/CT一致率为100%。
- 肿瘤CL信号显著强于周围正常组织(P<0.05)。
- CLE成像60秒与300秒的信噪比分别为2.70±0.48与2.98±0.69(P=0.56)。
临床怎么用
- 原文结论指出CLE具有在临床实践中早期诊断胃肠道肿瘤的潜力。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
BACKGROUND: The diagnostic yield of conventional gastrointestinal (GI) endoscopy for early cancers is low because it is mainly based on morphological changes of tumors. Molecular functional changes in tumors precede morphological changes. Cherenkov luminescence endoscopy (CLE) system can perform molecular imaging of GI cancers, achieving early diagnosis of cancers. However, previous CLE systems had only been able to detect Cherenkov luminescence (CL) from about one μCi nuclide at a minimum (in vivo), but the nuclide probe absorbed by the tumor of a patient was often much less than one μCi at a routinely administered dose. This study aims to construct a clinically usable high-sensitivity CLE for molecular imaging of GI cancers. RESULTS: The minimum resolvable radioactivity of the CLE reached 0.020 μCi within 300 s (in vivo), with a sensitivity at the nanocurie for the first time. The detection sensitivity of the CLE increased by up to nearly twenty-two times over the previous system. In tumor-bearing nude mice, CLE could effectively identify all tumors with 100% concordance with both histopathology and PET/CT, and the CL signals of tumors were much stronger than those of the surrounding normal tissues (P < 0.05). The quality of CLE imaging at 60 s was comparable to that at 300 s (signal-to-background ratio, 2.70 ± 0.48 versus 2.98 ± 0.69, P = 0.56). CONCLUSIONS: We constructed a high-sensitivity CLE that could detect radionuclides at the nanocurie radioactivity. The CLE could detect cancers accurately through rapid molecular imaging and had the potential for early diagnosis of GI cancers in clinical practice.
引用
- Yang Z, Pang TT, Wu ZJ, et al. Construction of a high-sensitivity Cherenkov luminescence endoscopy system for the detection of gastrointestinal cancers. EJNMMI Res. 2025. PMID: 40178783. doi:10.1186/s13550-025-01223-9
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。