文献速递肿瘤 PET/CT放射性药物与辐射防护

使用增强现实系统定位前哨淋巴结:一项尸体可行性研究

Localization of sentinel lymph nodes using augmented-reality system: a cadaveric feasibility study.
PubMed 文献 · Eur J Nucl Med Mol Imaging · 2025-03-24

该尸体可行性研究将氟-18标记的分子筛植入尸体头颈部模拟前哨淋巴结,用PET/MRI替代SPECT/CT生成虚拟影像并经HoloLens 2投影,5具尸体分别接受标准、增强现实及增强现实加头部运动补偿手术;增强现实表面定位平均误差2.5±2.0 mm、深度误差2.3±1.7 mm,均优于手动表面标记的平均误差18.6±13.0 mm(p<.001)。

要点

  • 增强现实表面定位平均误差2.5±2.0 mm、深度误差2.3±1.7 mm,均在PET体素分辨率内。
  • 与手动表面标记平均误差18.6±13.0 mm相比,增强现实在头直立和旋转位置均显著降低误差(p<.001)。
  • 增强现实将任务完成时间从手动方法的47.4 s缩短至12.3 s,减少74%。

核心结论

  • 增强现实系统在前哨淋巴结定位中表现出高准确性和效率,整合头部运动补偿与三维可视化以提高精度并减少手术室时间。

研究设计与样本

  • 尸体可行性研究:将8 mm分子筛在氟-18中孵育以模拟淋巴结,并植入尸体头颈部区域。
  • 因机构对尸体的限制,用PET/MRI替代SPECT/CT,虚拟PET/MRI渲染图经HoloLens 2和定制软件投影。
  • 5具尸体接受标准、增强现实及增强现实加头部运动补偿方法的手术。

主要发现

  • 增强现实表面定位平均误差2.5±2.0 mm(范围0-8 mm),深度误差2.3±1.7 mm(范围1-7 mm),均在PET体素分辨率内。
  • 对于更具挑战性的V区淋巴结,平均表面误差略增至2.9 mm。
  • 与手动表面标记平均误差18.6±13.0 mm(范围6-62 mm)相比,增强现实系统在头直立和旋转位置均显著降低误差(p<.001)。
  • 增强现实系统将任务完成时间减少74%(35.1/47.4 s),平均时间12.3 s,手动方法为47.4 s。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

PURPOSE: Sentinel lymph node biopsy (SLNB) helps stage melanoma. Pre-surgical single-photon emission computed tomography/computed tomography (SPECT/CT) visualizes draining lymph nodes, but intraoperative gamma probe detection only estimates SLN location. This study evaluates augmented reality (AR) for projecting pre-surgical SLN imaging onto patients to aid precise localization and extraction. METHODS: Molecular sieves (8 mm) incubated in fluorine-18 simulated lymph nodes and were implanted in the head and neck region of cadavers. Positron emission tomography/magnetic resonance imaging (PET/MRI) replaced SPECT/CT due to institutional restriction on cadavers. Virtual PET/MRI renderings were projected using the HoloLens 2 and custom software. Five cadavers underwent surgeries with standard, AR, and AR with head movement compensation methods. RESULTS: AR achieved a mean surface localization error of 2.5±2.0 mm (range, 0-8 mm) and a depth error of 2.3±1.7 mm (range, 1-7 mm), both within PET voxel resolution. For more challenging level V nodes, the mean surface error slightly increased to 2.9 mm. Compared to manual surface marking, which had an average error of 18.6±13.0 mm (range, 6-62 mm), the AR system significantly reduced errors both in the head-straight and rotated positions (p <.001). Additionally, the AR system reduced the task completion time by 74% (35.1/47.4 s), with an average time of 12.3 s compared to 47.4 s for manual methods. CONCLUSION: The AR system demonstrated high accuracy and efficiency in SLN localization, integrating head-movement compensation and 3D visualization to improve precision and reduce operating room time. CLINICAL TRIAL NUMBER: Not applicable.

来源:Eur J Nucl Med Mol Imaging,查看原文。版权归原作者所有。

引用

  1. Duan H, Yang Y, Niu WL, et al. Localization of sentinel lymph nodes using augmented-reality system: a cadaveric feasibility study. Eur J Nucl Med Mol Imaging. 2025. PMID: 40123008. doi:10.1007/s00259-025-07216-z

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。