文献速递肿瘤 PET/CT放射性药物与辐射防护

利用深度学习优化混合PET/MR中基于MR的衰减校正:通过平板插入件和一致患者体位进行验证

Optimizing MR-based attenuation correction in hybrid PET/MR using deep learning: validation with a flatbed insert and consistent patient positioning.
PubMed 文献 · Eur J Nucl Med Mol Imaging · 2025-02-06

本研究利用平板插入件和双臂下垂体位实现PET/CT与PET/MR的精确MR-CT匹配,在21例患者中验证四种基于深度学习的MR衰减校正方法。结果显示连续μ值法(MRAC-4)与CTAC一致性最佳,骨相关SUV干扰最小(r=0.01,p=0.8643),但在脊柱和肝脏等骨分割干扰易发区域再现性低于MRAC-2和MRAC-3。

要点

  • 平板插入件实现MR-CT精确匹配,可高精度验证MRAC方法。
  • 连续μ值MRAC-4与CTAC一致性最佳,骨相关SUV干扰最小(r=0.01,p=0.8643)。
  • MRAC-4在脊柱和肝脏等骨分割干扰易发区域再现性低于MRAC-2和MRAC-3。

核心结论

  • 使用平板插入件可高精度验证MRAC,连续μ值法MRAC-4准确性更优且骨相关SUV误差最小,但在骨丰富区域再现性面临挑战。

研究设计与样本

  • 验证数据集共21例患者,先接受全身[18F]FDG PET/CT,后接受[18F]FDG PET/MR。
  • 平板插入件确保体位一致,直接比较四种MRAC方法与CTAC:四组织与五组织模型分别搭配离散与连续μ图。
  • 深度学习框架在300例患者数据集上训练,从MR图像生成合成CT,作为所有MRAC方法的基础。

主要发现

  • 脊柱和股骨区域观察到MRAC方法间的明显差异。
  • 联合直方图分析显示MRAC-4(连续μ图)与CTAC高度一致。
  • 病灶距离分析显示MRAC-4使骨诱导的SUV干扰最小(r=0.01,p=0.8643)。
  • 脊柱和肝脏等骨分割干扰易发组织的SUV变异性更大、再现性更低,MRAC-4低于MRAC-2(2D骨分割,离散μ图)和MRAC-3(3D骨分割,离散μ图)。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

PURPOSE: To address the challenges of verifying MR-based attenuation correction (MRAC) in PET/MR due to CT positional mismatches and alignment issues, this study utilized a flatbed insert and arms-down positioning during PET/CT scans to achieve precise MR-CT matching for accurate MRAC evaluation. METHODS: A validation dataset of 21 patients underwent whole-body [18F]FDG PET/CT followed by [18F]FDG PET/MR. A flatbed insert ensured consistent positioning, allowing direct comparison of four MRAC methods-four-tissue and five-tissue models with discrete and continuous μ-maps-against CT-based attenuation correction (CTAC). A deep learning-based framework, trained on a dataset of 300 patients, was used to generate synthesized-CTs from MR images, forming the basis for all MRAC methods. Quantitative analyses were conducted at the whole-body, region of interest, and lesion levels, with lesion-distance analysis evaluating the impact of bone proximity on standardized uptake value (SUV) quantification. RESULTS: Distinct differences were observed among MRAC methods in spine and femur regions. Joint histogram analysis showed MRAC-4 (continuous μ-map) closely aligned with CTAC. Lesion-distance analysis revealed MRAC-4 minimized bone-induced SUV interference (r = 0.01, p = 0.8643). However, tissues prone to bone segmentation interference, such as the spine and liver, exhibited greater SUV variability and lower reproducibility in MRAC-4 compared to MRAC-2 (2D bone segmentation, discrete μ-map) and MRAC-3 (3D bone segmentation, discrete μ-map). CONCLUSION: Using a flatbed insert, this study validated MRAC with high precision. Continuous μ-value MRAC method (MRAC-4) demonstrated superior accuracy and minimized bone-related SUV errors but faced challenges in reproducibility, particularly in bone-rich regions.

来源:Eur J Nucl Med Mol Imaging,查看原文。版权归原作者所有。

引用

  1. Wang H, Wang Y, Xue Q, et al. Optimizing MR-based attenuation correction in hybrid PET/MR using deep learning: validation with a flatbed insert and consistent patient positioning. Eur J Nucl Med Mol Imaging. 2025. PMID: 39912939. doi:10.1007/s00259-025-07086-5

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。