呼吸门控PET采集中的定量:数据驱动方法能否替代基于设备的系统——一项比较性回顾性研究
该研究回顾性分析本中心196例患者的[18F]FDG PET/CT扫描,共测量536个病灶,比较数据驱动呼吸门控(OncoFreeze AI™)与基于设备的呼吸门控在SUVmax和SUVpeak上的定量差异。全队列平均绝对偏差分别为3.8%(CI 3.4-4.2)和2.1%(CI 1.9-2.4),仅2%患者SUVmax偏差显著,0.5%患者SUVpeak偏差显著。结论认为数据驱动呼吸门控是常规肿瘤[18F]FDG-PET/CT评估中基于设备门控方法的可靠且准确的替代方案。
要点
- 全队列SUVmax平均绝对偏差3.8%(CI 3.4-4.2),SUVpeak为2.1%(CI 1.9-2.4)。
- 患者水平21%至少一个病灶SUVmax偏差超过10%,SUVpeak该比例为5%。
- 按PERCIST标准,仅2%患者SUVmax偏差显著,0.5%患者SUVpeak偏差显著。
核心结论
- 数据驱动与基于设备的呼吸门控PET扫描定量比较显示差异可忽略,证明数据驱动呼吸门控是常规[18F]FDG-PET/CT肿瘤评估中基于设备门控方法的可靠且准确的替代方案。
研究设计与样本
- 回顾性分析本中心连续筛查的692例患者中196例的[18F]FDG PET/CT扫描,共测量536个病灶。
- 所有PET/CT采集均在Siemens Biograph 64 Vision 600系统上进行,同时使用呼吸带门控和添加OncoFreeze AI™的数据驱动门控。
- 记录每例患者最多5个病灶的SUVmax和SUVpeak,计算两种门控方法在队列水平和亚组水平的平均绝对偏差及95%置信区间。
主要发现
- 全队列病灶SUVmax和SUVpeak的平均绝对偏差分别为3.8%(CI 3.4-4.2)和2.1%(CI 1.9-2.4)。
- 患者水平,21%的患者至少有一个病灶SUVmax偏差超过10%,而SUVpeak该比例为5%。
- 按PERCIST标准亚组分析,仅2%患者SUVmax偏差显著,0.5%患者SUVpeak偏差显著。
- 病灶大小或解剖部位对两种呼吸门控方法间SUV测量无临床显著影响。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
BACKGROUND: Device-based respiratory gating improves diagnostic and quantification accuracy in positron emission tomography (PET), but requires additional time to setup the device and the failure rate can be significant. Our aim was to internally validate the quantification performance of data-driven respiratory-gated PET imaging against the gold standard, the device-based method, in clinical oncological practice. We retrospectively analysed [18F]FDG PET/CT scans of patients from our centre with at least one measurable [18F]FDG-avid malignant lesion. All PET/CT acquisitions were performed on a Siemens Biograph 64 Vision 600 system with respiratory gating by belt and also by adding the data-driven gating with OncoFreeze AI™. We recorded the SUVmax and SUVpeak for up to a maximum of 5 lesions per patient. We computed the mean absolute bias between the two gating methods and the 95% confidence intervals (CI) at the cohort level and in subgroups. RESULTS: Of the 692 consecutive patients screened for inclusion, 196 patients were analysed, from whom 536 lesions were measured. The mean absolute biases in the SUVmax and SUVpeak of lesions in the whole cohort were 3.8% (CI 3.4-4.2) and 2.1% (CI 1.9-2.4), respectively. At patient-level, 21% of them had at least one lesion with a SUVmax bias above 10%, while for SUVpeak this proportion was 5%. In the subgroup analysis by PERCIST criteria, only 2% of patients had significant bias in the SUVmax, and 0.5% in SUVpeak. There was no clinically significant effect of lesion size or anatomical site on SUV measurements between the two respiratory gating methods. CONCLUSION: Quantitative comparison of data-driven and device-based respiratory-gated PET scans revealed negligible differences, proving that data-driven respiratory gating is a reliable and accurate alternative to the device-based gating method in routine [18F]FDG-PET/CT oncological evaluation.
引用
- Agrigoroaie L, Anizan N, Garcia C, et al. Quantification in respiratory-gated PET acquisition: can data-driven methods replace device-based systems?-a comparative and retrospective study. EJNMMI Res. 2025. PMID: 39953296. doi:10.1186/s13550-025-01195-w
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。